i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test - 1

i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test Scope i-Sepamate 7800 Separation Membrane Permeability Analyzer is designed to determine the gas transmission volume, permeability rate and selectivity of single gas or mixed gases for separation membrane under specified conditions. Test Procedures A: Prepare test specimens according to specific standards. The test specimen should be of uniform thickness and should be free of wrinkles, creases, pinholes, or grease. The direction of permeation should be identified. Extreme care should be taken when cutting the specimen to avoid any contamination.

Open the catalog to page 1
i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test - 2

B: Open the main valve for the test gas and carrier gas (helium) and adjust the output pressure as required. Power the instrument on. Turn on the vacuum pump and power the gas chromatograph on and check the gas outlet on the top of the gas chromatograph with the flow meter provided.

Open the catalog to page 2
i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test - 3

D: Wipe off the test cell and tighten the upper test chamber and lower test chamber with no test specimen placed in. Then turn on the vacuum pump and set up test parameters for blank test. Then start the blank test and save test report after the test is finished. Establish method for blank test. E: Apply a certain amount of vacuum grease on the outer surface 2mm around the hollow groove of lower test chamber and place a piece of filter paper (diameter is 20mm) on the hollow groove. Then place the prepared specimen on the lower test chamber and make sure the test specimen contacts the lower...

Open the catalog to page 3
i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test - 4

200 River's Edge Drive, Medford, MA 02155 Test Name: Test Date: Test Method: Specimen Material: |pvc Remarks: (remark Test Standard: Specimen Name: Specimen Provider: Specimen Hanufacturer: Specimen Thickness(urn): Inspection Organization: |labthink Reference Value: r-Iest Parameters- Test Temperature: -U_Chamber Inflation Mode (§) Retain Pressure! Purge Time(s): -Test Mode- Vacuumization lime Time(s): ^ft^H System Purging Click "Calculation" so that the results analyzed by GC will be added to the result list. Calculation & Results Result List rInput Measured Value- Calculate Result

Open the catalog to page 4
i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test - 5

200 River's Edge Drive, Medford, MA 02155 Test parameters Test Mode: Test Date: Specimen Manufacturer: labthink Inspection Organization:labthink Test results Test Name: test U_Chamber Inflation Mode:Retain Pressure Test Method: method Specimen Provider: labthink Remarks: remark