1. Catalogs
  2. Labthink Instruments Co., Ltd.
  3. i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test

i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test

i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test

i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test

Product catalog summary
Scope
The i-Sepamate 7800 Separation Membrane Permeability Analyzer is designed to measure the gas transmission volume, permeability rate, and selectivity of single or mixed gases through separation membranes under specified conditions.
Test Procedures
  • Specimen Preparation: Prepare test specimens ensuring uniform thickness and absence of defects. Identify the direction of permeation and avoid contamination during cutting.
  • Gas Setup: Open the main valve for test and carrier gases (helium), adjust output pressure, power on the instrument, and check the gas outlet with a flow meter.
  • Parameter Setup: Configure the gas chromatograph parameters.
  • Blank Test: Clean the test cell, tighten chambers without a specimen, perform a blank test, and save the report.
  • Specimen Placement: Apply vacuum grease, place filter paper and specimen on the lower test chamber ensuring tight contact.
  • Specimen Test: Set parameters and start the test, which runs automatically.
  • Data Collection: Analyze collected gases with GC, save chromatogram, and calculate results. View and print the complete test report.
  • Post-Test Procedure: Cut off gas supply if no further tests are planned shortly to prevent accidents.
See more

Catalog excerpts

i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test-1

i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test Scope i-Sepamate 7800 Separation Membrane Permeability Analyzer is designed to determine the gas transmission volume, permeability rate and selectivity of single gas or mixed gases for separation membrane under specified conditions. Test Procedures A: Prepare test specimens according to specific standards. The test specimen should be of uniform thickness and should be free of wrinkles, creases, pinholes, or grease. The direction of permeation should be identified. Extreme care should be taken when cutting the specimen to avoid any contamination.

 Open the catalog to page 1
i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test-2

B: Open the main valve for the test gas and carrier gas (helium) and adjust the output pressure as required. Power the instrument on. Turn on the vacuum pump and power the gas chromatograph on and check the gas outlet on the top of the gas chromatograph with the flow meter provided.

 Open the catalog to page 2
i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test-3

D: Wipe off the test cell and tighten the upper test chamber and lower test chamber with no test specimen placed in. Then turn on the vacuum pump and set up test parameters for blank test. Then start the blank test and save test report after the test is finished. Establish method for blank test. E: Apply a certain amount of vacuum grease on the outer surface 2mm around the hollow groove of lower test chamber and place a piece of filter paper (diameter is 20mm) on the hollow groove. Then place the prepared specimen on the lower test chamber and make sure the test specimen contacts the lower test...

 Open the catalog to page 3
i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test-4

200 River's Edge Drive, Medford, MA 02155 Test Name: Test Date: Test Method: Specimen Material: |pvc Remarks: (remark Test Standard: Specimen Name: Specimen Provider: Specimen Hanufacturer: Specimen Thickness(urn): Inspection Organization: |labthink Reference Value: r-Iest Parameters- Test Temperature: -U_Chamber Inflation Mode (§) Retain Pressure! Purge Time(s): -Test Mode- Vacuumization lime Time(s): ^ft^H System Purging Click "Calculation" so that the results analyzed by GC will be added to the result list. Calculation & Results Result List rInput Measured Value- Calculate Result

 Open the catalog to page 4
i-Sepamate 7800 Separation Membrane Permeability Analyzer ---Film Test-5

200 River's Edge Drive, Medford, MA 02155 Test parameters Test Mode: Test Date: Specimen Manufacturer: labthink Inspection Organization:labthink Test results Test Name: test U_Chamber Inflation Mode:Retain Pressure Test Method: method Specimen Provider: labthink Remarks: remark

 Open the catalog to page 5

All Labthink Instruments Co., Ltd. catalogs and technical brochures

  1. C610M

    3  Pages

  2. C690H

    4  Pages

  3. C130H

    4  Pages

  4. COF-P01

    2  Pages

  5. VAC-V2

    3  Pages

  6. C230H (2025)

    4  Pages

  7. C610H

    3  Pages

  8. C206H

    4  Pages

  9. C406H

    5  Pages

  10. C403H

    5  Pages

  11. C 681M

    1  Page

  12. C660B (2025)

    3  Pages

  13. C632M

    3  Pages

  14. C630H

    3  Pages

  15. C620H

    3  Pages

  16. C 640

    4  Pages

  17. C 660M

    3  Pages

  18. C306H (2025)

    4  Pages

  19. C203H

    4  Pages

  20. BTY-B3P

    2  Pages

  21. C360M

    4  Pages

  22. C632B

    3  Pages

  23. C190H

    3  Pages

  24. C390H

    4  Pages

  25. C360H

    4  Pages

  26. C660M

    3  Pages

  27. C303H

    4  Pages

  28. C106H

    5  Pages

  29. C640

    4  Pages

  30. MXD-02

    2  Pages

  31. CHY-C2A

    2  Pages

  32. C611B

    2  Pages

  33. CH130

    4  Pages

  34. C 631 H

    2  Pages

  35. C660 B

    3  Pages

  36. VAC-VBS

    3  Pages

  37. XYD-15K

    2  Pages

  38. C360

    4  Pages

  39. C631 H

    2  Pages

  40. FST-02

    2  Pages

  41. C681 M

    2  Pages

  42. C230

    4  Pages

  43. C631H

    2  Pages

  44. C612M

    3  Pages

  45. SLY-SI

    2  Pages

  46. C360 M

    4  Pages

  47. C611M

    2  Pages

  48. C330H

    4  Pages

  49. Leak Tester

    3  Pages

  50. Tensile Testers

    19  Pages

  51. FDI-01

    5  Pages

  52. DIN 75201

    2  Pages

  53. Abrasion Test

    2  Pages

  54. Torque Meter

    2  Pages

  55. Leak Detector

    2  Pages

  56. Torque tester

    2  Pages

  57. Impact tester

    2  Pages

  58. Gas Analyzer

    2  Pages

  59. TOQUE TESTER

    1  Page

Archived catalogs

  1. C611M

    2  Pages

  2. C306H

    3  Pages

  3. C230H

    4  Pages

  4. Leak Tester

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.