video corpo

C201B Oxygen Transmission Rate Test System
3Pages

{{requestButtons}}

Catalog excerpts

C201B Oxygen Transmission Rate Test System - 1

C201B Oxygen Transmission Rate Test System C201B Oxygen Transmission Rate Test System is designed and manufactured based on the coulometric sensor method and conforms to ASTM D3985. This instrument can be used to measure the oxygen transmission rate of barrier materials with high and medium barrier properties with high accuracy and high efficiency. C201B is applicable to the determination of oxygen permeability of plastic films, sheeting, paper, and other packaging materials used in food, pharmaceutical, medical apparatus, consumer products, photovoltaic and electronic industries, etc. Features note1 • Equipped with Labthink self-developed coulometric sensor conforming to ASTM D3985, intrinsic standard and no calibration is required • Support temperature control to satisfy tests under different conditions (optional) • Built-in high-quality stainless steel test cell with better sealing performance • Imported manual isolation valve and flow regulator are adopted for better sealing performance and lower failure rate • Industrial computer appearance design, small size and fast cooling • Reference film is provided for quick calibration • The system adopts single-chip microcomputer control and can run independently • The test report can be exported in common formats such as EXCEL and PDF • Support micro printer, automatically print test data (optional) • Equipped with RS232 data interface, which can be connected to computer software for curve analysis, data storage, report printing, etc. (optional) • The gas purifier independently developed by Labthink can remove trace oxygen in nitrogen and provide oxygen-free carrier gas (optional). Labthink Instruments Co., Ltd. 144 Wuyingshan Road, Jinan, P.R.China (250031) Phone: +86-531-85068566 FAX: +86-531-85062108 Labthink International, Inc. 200 River's Edge Drive, Medford, MA, 02155, U.S.A. Phone:+1-617-830-2190 FAX: +1-781-219-3638 www.labthink.com

Open the catalog to page 1
C201B Oxygen Transmission Rate Test System - 2

Test principle The pre-conditioned specimen is clamped in the test cell, oxygen or air flows on one side of the specimen while a stream of high purity nitrogen flows on the other side. Oxygen molecules permeate through the specimen into the nitrogen side and are carried to the coulometric sensor. The sensor analyzes the oxygen concentration and calculates the oxygen transmission rate. Standards ASTM D3985x ASTM F1307x GB/T 19789x GB/T 31354x DIN 53380-3x JIS K7126-2-Bx YBB 00082003-2015 Oxygen transmission rate test of various plastic films, paper-plastic composite films, coextruded films,...

Open the catalog to page 2
C201B Oxygen Transmission Rate Test System - 3

TC03 Temperature control device Optional Table 2: Technical specifications Test Cell Specimen Size Specimen Thickness Standard test area 99.999% High purity nitrogem 99.5%Oxygen (Outside supply scope) Port Size Instrument Dimension Table 3: Product Configuration Standard Configuration Instrument mainframe, sampler, vacuum grease Optional Parts Professional software, GP-01 Gas Purifier ,TC03 temperature control device, Computer System requirement, Micro-printer Note 1: The described product functions are subject to the specification in "Technical Parameters” Note 2: The parameters in the...

Open the catalog to page 3

All Labthink Instruments Co. catalogs and technical brochures

  1. FDI-01

    5 Pages

  2. DIN 75201

    2 Pages

  3. Torque Meter

    2 Pages

  4. Gas Analyzer

    2 Pages

  5. TOQUE TESTER

    1 Pages

Archived catalogs

  1. Leak Tester

    2 Pages