GC-4500 - 1

uiBomcD. me.GC-4500 Gas Chromatography System 5 kinds of sample injectors and all conventional detectors can be attachable depending on samples for analysis. In particular, for analysis of environmental samples (water quality, air, soil and wastes) satisfactory outcome can be achieved by connecting a sample preparation system such as the Purge & Trap. Features: • Automatic flow control and ignition of the FID- flame • Auto-start of chromatogram when sample is injected • Gas save function (Capillary Injector) • TCD maximum current control based on the type of gas • Optional valves (solenoid, 2-position) control function • Automatic notification of power-off Auto Sampler/Injector: • Dual inlet injection capability • Variable injection/syringe fill rates • Methods linking • Multiple solvenl/waste bottles • Syringe range 5uL-250uL • Air gapping capability • Multiple tray options: 120 vials, 1,8mL; 220 vials, 0.8mL; 60 vials, 10mL Gas Sampling Valve: • Two Valves can be controlled by Air Actuation • Full automation system is built along with the automatic Run/Stop function INLET SYSTEM - Excellent accuracy, reproducibility Packed Inlet: • 1/4", 1/8" Glass-lined Metal Liner are available • On-Column injection method (direct injection of sample into column) is available using 1/4" Glass Column • 0.53mm ID Capillary Column can be used Capillary Inlet: • High resolution, prompt analysis time and most typical inlet system • Split and Splitless mode can be selectable Labomed, Inc., Mailing Address: 2728 S La Cienega Blvd., Los Angeles, CA 90034 U.S.A. TEL (310) 202-0811 FAX (310) 202-7286 Email: spectro@labomed.com, labomedinc@gmail.com Capillary Inlet

Open the catalog to page 1
GC-4500 - 2

uiBomcD. me. GC-4500 Gas Chromatography System DETECTORS - High selectivity and sensitivity FID (Flame Ionization Detector): • Improved sensitivity into 19mCoulomb/g Carbon • Easy to disassemble and assemble detector • Minimization of sample degeneration as well as jet pollution by means of inert quartz jet • Auto ignition of hydrogen flame • Use as TID by replacing a collector and bead TCD (Thermal Conductivity Detector): • High detection reproducibility by excellent temperature stability • Supplying static current to 4 Rhenium-Tungsten filaments and selecting the optimal reactivity by...

Open the catalog to page 2
GC-4500 - 3

uiBomcD. me. GC-4500 Gas Chromatography System P STATUS INFORMATION STANDBYCONDITION Start the GC-4500 control program (17:13] Start for the function of flame auto ignition (17 13] [FID] Sequence 1/6 - START (17 13] (FID] Sequence 2/6 - FLOW READY (17 14) [FID] Sequence 3/6 - FLAME ON (17 14] (FID] Sequence 4/6 - SET RETURN FLOW (17 14] [FID] Sequence 5/6 - FLOW READY (17 14] (FID] Sequence 6/5 - STOP [17:14] Stop for the function of flame auto ignition Easy and Excellent Control: • Exact flow/pressure control by IPC (Intelligent Pneumatic Control) • Auto ignition of FID flame • Auto...

Open the catalog to page 3
GC-4500 - 4

innomcD. me.GC-4500 Gas Chromatography System Easy and Excellent Control: • Direct Syringe Injection • Variable Injection Rates • Variable Syringe Fill Rates • Dual Inlet Injection Capability • Syringe Range 5uL-250uL • Multiple Tray Option: 120 vials, 1,8mL; 220 vials, 0.8mL; 60 vials, 10mL • Uses Off-the Shelf Syringes The GC-4500 Liquid Auto Sampler is an entirely new sampling system designed and built to automate any GC. If you are looking for a sampling system to replace an older, less reliable auto sampler, the GC-4500 Liquid Auto Sampler is the logical choice. The flexibility of the...

Open the catalog to page 4
GC-4500 - 5

uiBomcD. me.GC-4500 Gas Chromatography System GC-4500 HEADSPACE AUTO SAMPLER Patented Dual Needle Dynamic Headspace The GC-4500 Headspace Auto Sampler utilizes a closed sample introduction system. Unlike syringe injection systems, the loss of headspace vapor during the injection process due to vapor pressure changes is never a concern. In addition to the sensitivity benefits of the Dynamic Headspace technology, the GC-4500 Headspace Auto Sampler delivers the added flexibility of performing traditional fixed loop injections and time-based injections incorporated within a single system....

Open the catalog to page 5
GC-4500 - 6

uiBomcD. me. GC-4500 Gas Chromatography System TECHNICAL SPECIFICATIONS COLUMN VIEW Inner volume Temperature Range Linear Programming Rate Heating Rate Cooling Rate Run Time Temp. Programming 16.25 liter (250mm W* 260mm D*250mm H) Ambient to 400°C Max: 120°C/min up to 400°C (0.113 increment) About 2.3 minutes 50°C to 300°C About 6 minutes 300°C to 50°C Automatically computed to 999.9 minutes by 0.1 minute increments 30 Step PNEUMATICS (IPC Module) IPC pressure range IPC flow range Pressure resolution Flow resolution Support IPC Channels Pressure/flow programming steps EPC control Split...

Open the catalog to page 6
GC-4500 - 7

uiBomcD. me. GC-4500 Gas Chromatography System TECHNICAL SPECIFICATIONS DATA ACQUISITION SYSTEM: Dynamic range Sampling Rate Analog signal voltage ranges Analog to digital conversion resolution External event I/O ports Remote Data Visualization Data Management Custom reports >2*106 User selectable 1 Q-60Hz per channel ±10,5,2.5,1.25V 21 bits at all sampling rates 8 Start, stop control independent Up to chromatograms in real time, zoom/unzoom at any time; Scrutinize focused peaks, display baseline; Chromatograms overlap, background subtraction. Data export results as Text files; Dynamic data...

Open the catalog to page 7
GC-4500 - 8

uiBomcD. me. GC-4500 Gas Chromatography System TECHNICAL SPECIFICATIONS DETECTOR Thermal Conductivity Detector (TCD) Cell Flow through cell design, 4 Rhenium/tungsten filaments (32/each) Temp. Range Ambient to 400°C Bridge Power Supply Constant current system (1 -400mA) Filament Protection circuit Automatically protect from over load and over current Micro-cell TCD Cell 4 coaxial, hot wire cell Column Ideal for use with capillary columns Flame Ionization Detector (FID) Electrode Voltage MDQ Linearity Sensitivity Jet Tip Temp. Range Variability -230V 0.700pgC/s for...