Panel Mount Instrumentation
12Pages

{{requestButtons}}

Catalog excerpts

Panel Mount Instrumentation - 1

Leaders in temperature measurement The Labfacility L200 Pt100 and thermocouple thermometer can be used in conjunction with a PC to provide accurate, versatile 8 channel thermocoupl e and Pt100 temperature measurement, scanning and logging of measured values. It can also be used as a "stand alone" indicator/logger and incorporates a digital display of measured temperature. The in-built, self-calibration facility for the thermocouple version is a rapid and convenient method for on-site calibration and does not require any additional equipment other than a special, external link. Self-calibration of Pt100 ranges is equally simple and uses plug-in precision resistors. The L200 is designed to provide exceptional stability with high measurement resolution and represents an ideal crossover between plant practicality and laboratory performance at a very competitive price. The PC software supplied with the instrument allows control, measure and calibration functions. The bench top case can be converted to a panel mounting configuration using the optional mounting brackets. www.labfacility.com | sales@labfacility.com sales tel: +44 (0)1243 871280

Open the catalog to page 1
Panel Mount Instrumentation - 2

● Low cost – high performance ● Bench top or panel mounting ● USB PC interface ● 8 thermocouple (type J,K,T,E,N,R,S & B) or Pt100 (3 wire) inputs ● Built-in display for selected channel or all channels auto-scrolling ● Resolution 0.1°C on L200 display, 0.01°C in software ● Self-calibration feature ● Select °C / °F ● PC software included for remote control and measure, logging and calibration ● Simple operation ● CE marked ● RoHS compliant ● 2 year warranty Applications ✦ Food preparation ✦ Storage facilities ✦ Free standing or in-panel ✦ Technical educational establishments ✦ Environmental...

Open the catalog to page 2
Panel Mount Instrumentation - 3

Specification at an ambient temperature of 20°C Input / Ranges

Open the catalog to page 3
Panel Mount Instrumentation - 4

* The integral, self-calibration facility for the thermocouple version is a rapid and convenient method for on-site calibration and does not require any additional equipment other than the special, external link (optional). Self-calibration of Pt100 ranges is quickly and conveniently performed using plug-in precision resistors (optional). Traceable calibration can be achieved by the user conveniently and without recourse to a accredited Laboratory if there is access to a certified DVM; this can be used to measure the L200 internally generated calibrated Voltage via the "cal port" presented...

Open the catalog to page 4
Panel Mount Instrumentation - 5

L200 Panel Mounting – Schematics Rear Panel Mount Front Panel Mount

Open the catalog to page 5
Panel Mount Instrumentation - 7

LABFACILITY L300 8-ZONE TEMPERATURE ALARM / ON-OFF CONTROLLER WITH 10A SWITCHING The Labfacility L300 Pt100 and Thermocouple temperature alarm / on-off controller can be used in conjunction with a PC to provide accurate monitoring and alarm or on-off control of up to 8- zones simultaneously. It can also be used as a stand-alone instrument without the need for a PC. The PC software supplied with the instrument allows control, configuration, measurement, logging, charting, alarm & relay configuration and calibration functions via a PC. The in-built, self-calibration facility for the...

Open the catalog to page 7
Panel Mount Instrumentation - 8

● Low cost – high performance ● Bench top or panel mounting ● USB PC interface ● 8 thermocouple (type J,K,T,E,N,R,S& B) or Pt100 (3 wire) inputs ● Built-in display for selected channel or all channels auto-scrolling ● Resolution 0.1°C on display, 0.01°C in software ● Self-calibration feature ● Select °C / °F ● PC software included for remote control and measure, logging, configuration and calibration ● 8 x configurable change-over relays 10A/250V ● Simple operation ● CE marked ● RoHS compliant ● 2 year warranty Applications ✦ Food preparation ✦ Storage facilities ✦ Free standing or in-panel...

Open the catalog to page 8
Panel Mount Instrumentation - 9

Specification at an ambient temperature of 20°C Measurement Input / Ranges Thermocouple to IEC 584 Type J -200°C to 750°C Type K -200°C to1200°C Type T -200°C to 350°C Type E -200°C to 900°C Type N 0°C to 1300°C Type R 0°C to 1760°C Type S 0°C to 1760°C Type B 300°C to 1800°C Pt100 to IEC751,3 wire -200°C to 850°C Note: all inputs are non-isolated and sensors must be of insulated construction. Accuracy Thermocouples J K T E & N Thermocouples R S & B Linearisation Pt100 range Zero drift Span drift Display Display resolution Indication Reference junction compensation for thermocouples better...

Open the catalog to page 9
Panel Mount Instrumentation - 10

* The integral, self-calibration facility for the thermocouple version is a rapid and convenient method for on-site calibration and does not require any additional equipment other than the special, external link (optional). Self-calibration of Pt100 ranges is quickly and conveniently performed using plug-in precision resistors (optional). Traceable calibration can be achieved by the user conveniently and without recourse to a accredited Laboratory if there is access to a certified DVM; this can be used to measure the L200 internally generated calibrated Voltage via the "cal port" presented...

Open the catalog to page 10
Panel Mount Instrumentation - 11

L300 Panel Mounting – Schematics Rear Panel Mount Front Panel Mount

Open the catalog to page 11

All Labfacility Limited catalogs and technical brochures

  1. Food Safety

    2 Pages

  2. IM-N-F-HTC

    1 Pages

  3. IM-R/S-F-HTC

    1 Pages

  4. IS-K-BTS

    1 Pages

  5. IS-J-BTS

    1 Pages

  6. IS-T-BTS

    1 Pages

  7. IS-R/S-BTS

    1 Pages

  8. IS-CU-BTS

    1 Pages

  9. IS-N-M

    1 Pages

  10. Wall Chart

    2 Pages