video corpo

K.130 A/C series
6Pages

{{requestButtons}}

Catalog excerpts

K.130 A/C series - 3

K.130 A - K.130 C series freezers at -125 °C STRUCTURE: The internal casing is in AISI 304 stainless steel (or AISI 316 upon request) with rounded angles for easy cleaning; the external cabinet is a plasticised, zinc-plated steel sheet; the insulation is in polyurethane resin, foamed on site, with a density of about 40 Kg/m3 and a noticeable average thickness: 190 mm. CFC-, HCFC-free; the door is placed on the upper part, mounted on self-balancing hinges; internal counter doors (in plastic-coated foamed polyurethane resin) sliding on a special track; it has a triple silicone rubber seal,...

Open the catalog to page 3
K.130 A/C series - 4

K.130 A - K.130 C series freezers at -125 °C Self-diagnosis test: upon boot up, the freezer’s LEDs and buzzer are activated for about 1”. User Menu access protected by password. Password-protected Setup access menu for service operations. Cooling Startup/Shutdown The startup and shutdown of the cooling cycle requires password-protected manoeuvres: The cycle starts from a standby status (lower display showing OFF). First the “START” key is pressed, then the display shows a password request; if correctly inserted, the cooling cycle will start; otherwise the display will show OFF again. The...

Open the catalog to page 4
K.130 A/C series - 5

PRESS ALARM DOOR OPEN POWER FAIL TEMP ALARM ACCESSORIES Upon request, each system can be furnished with: • Electronic video graphical recorder with RTD input Pt 100 Ω, V230/1/50 • Additional sensors for T, RTD Pt 100, to connect to an external system for the acquisition and recording of T values, such as KW SPY® or similar. • Additional RTD Pt 100 sensor complete with 4-20 mA converter mounted on a DIN bar to connect to an external recording system. • Internal-external ø 25 mm. connection holes with external rubber stopper. • Emergency LN2 system (complete with connection, flexible hose and...

Open the catalog to page 5
K.130 A/C series - 6

= min./max temperature alarm Contacts: Managing Director: management@kwkw.it Technical Production Manager: technician@kwkw.it Sales Manager: sales@kwkw.it Tender Commercial Office: commerciale@kwkw.it International Sales Manager: expdpt@kwkw.it Technical Support & Spare Parts: assistenza@kwkw.it Service line: service@kwkw.it Delivery Office: delivery@kwkw.it Administration: administration@kwkw.it = DATA LOGGER function = energy failure alarm = Alarm broken down probe = Open door alarm = Disaster Ricovery / Safety Control = Temperature recorder KW Apparecchi Scientici, taking into...

Open the catalog to page 6

All KW catalogs and technical brochures

  1. WPLR series

    6 Pages

  2. WRC series

    3 Pages

  3. SLC series

    6 Pages

  4. BloodLine

    8 Pages

  5. KFT series

    4 Pages

  6. Containers

    4 Pages

  7. KW.86 series

    4 Pages

  8. BlueLine

    10 Pages

  9. HS series

    18 Pages

  10. WS series

    4 Pages

  11. KFCE series

    6 Pages

  12. ServiceLine

    32 Pages

Archived catalogs

  1. GreenLine

    40 Pages