Product catalog summary
Technical Specifications
The UV Detector UVD 2.1L operates within a temperature range of 4 - 40 °C and a humidity level below 90% non-condensing. It features a variable wavelength UV detector with fiber optic connectors, utilizing a deuterium lamp as its light source. The wavelength range is 190 - 750 nm, with a spectral bandwidth of 11 nm. The detector ensures a wavelength accuracy of ± 2.5 nm and precision of 0.3 nm. Noise levels are ± 2.0 × 10-5 AU at 254 nm, and drift is 4.0 × 10-4 AU/h at the same wavelength.
Communication and Control
The device supports digital control via LAN-DHCP and RS-232, with an optional manual control unit. It offers a maximum data rate of 50 Hz for LAN and analog, and 10 Hz for RS-232. Programming capabilities include timed wavelength adjustments and event management.
Flow Cells
Flow cells are available in analytical and preparative types, with a wavelength range of 190 – 1020 nm. Analytical flow cells have path lengths of 10 mm and 3 mm, with maximum flow rates of 20 ml/min and 50 ml/min respectively. Preparative flow cells offer path lengths of 0.5/1.25/2 mm, with maximum flow rates up to 10,000 ml/min.
Replacing Components
When replacing the flow cell or fiber optics, it is crucial to avoid touching the ends with fingers to prevent measurement errors. The document provides detailed procedures for removing, storing, and installing flow cells, emphasizing the use of isopropanol and protective caps to maintain cleanliness and functionality.
Safety and Maintenance
Warnings are issued regarding UV light exposure, advising users to turn off the device and unplug it during maintenance. Regular checks and replacements of fiber optics are recommended after 6000 operating hours to ensure optimal performance.
Additional Technical Specifications
  • Path Length: Options include 0.5/1.25/2 mm, 0.5 mm, 10 mm, and 7 mm.
  • Capillary Connection: Various sizes including 1/8”, 1/16”, 3/8” with TRI-Clamp, and 1/2” with TRI-Clamp.
  • Inner Diameter: Options include 2.3 mm and 0.8 mm.
  • Volume: Options include 25 μl and 3 μl.
  • Material: PEEK, Stainless Steel, and Fused Silica.
  • Max. Flow Rate: 1000 ml/min and 250 ml/min.
  • Max. Pressure: Ranges from 10 bar to 200 bar.
Replacing the Flow Cell
  • Ensure the flow cell is rinsed and the device is switched off before removal.
  • Use isopropanol and cap fittings for handling.
  • Follow safety precautions to avoid UV light exposure.
  • Steps include unscrewing fittings, sealing connectors, and storing the flow cell securely.
Storing the Flow Cell
  • Fill the flow cell with isopropanol and seal with cap fittings to prevent contamination.
Installing the Flow Cell
  • Ensure the device is off and follow safety precautions.
  • Remove caps, screw fittings, and mount the flow cell appropriately.
Inspecting the Flow Cell
  • Conduct a function test with an integration time not exceeding 100 ms.
Replacing the Fiber Optics
  • Fiber optics should be replaced after approximately 6000 operating hours due to solarization.
  • Handle with care to avoid damage and ensure accurate measurements.
General Information
  • Document version: 1.2, last updated on 2013-06-05.
  • Produced by Wissenschaftliche Gerätebau Dr. Ing. Herbert Knauer GmbH.
  • For more information, visit www.knauer.net.
See more

Catalog excerpts

UVD 2.1L-1

Beiblatt Lichtwellenleiter Supplement Fiber Optics Version

 Open the catalog to page 1
UVD 2.1L-2

Inhaltsverzeichnis Hinweis Lesen Sie unbedingt zu Ihrer eigenen Sicherheit das Handbuch und beachten Sie immer die auf dem Gerät und im Handbuch angegebenen Warn- und Sicherheitshinweise! Technische Daten . . . . . . . . . . . . . . . . . . . . . . . . . . Umgebungsbedingungen . . . . . . . . . . . . . . . . . . Detektor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Durchflusszelle . . . . . . . . . . . . . . . . . . . . . . . . . . Analytische Durchflusszellen . . . . . . . . . . . . . Präparative Durchflusszellen . . . . . . . . . . . . . Austausch der Durchflusszelle . . ....

 Open the catalog to page 2
UVD 2.1L-3

Technische Daten Technische Daten Umgebungsbedingungen Temperatur- bereich Luftfeuchtigkeit unter 90 % nicht kondensierend Detektor Detektion Detektortyp UV-Detektor mit variabler Wellenlänge mit Lichtwellenleiter Deuteriumlampe (D2) mit integriertem GLP Chip Spektrale Bandbreite Rauschen Drift DE KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2

 Open the catalog to page 3
UVD 2.1L-4

Technische Daten Kommunikation Maximale Datenrate Error (IN), Start (IN), Autozero, Analoger Ausgang 1 x 0 - 5 V skalierbar, 16 bit, Offset einstellbar Steuerung Digital: LAN-DHCP, RS-232, Federleiste Analog: Steuerspannung zur Wellenlängenselektion Manuell: Control Unit (Optional) Zeitgesteuert: Wellenlängen, Events, Fraktionierventil, Links, Wake-Up (Programm, Link); 19 Programme, 99 Programmzeilen KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2 DE

 Open the catalog to page 4
UVD 2.1L-5

Technische Daten Detaillierter Report mit automatischer Lampenerkennung, Betriebsstunden, Lampenbetriebsstunden, Anzahl der Lampenzündungen Control Unit (Optional Allgemein Stromversorgung Durchflusszelle Der Wellenlängenbereich liegt für alle Durchflusszellen zwischen 190 – 1020 nm. Analytische Durchflusszellen Technische Daten DE KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2

 Open the catalog to page 5
UVD 2.1L-6

Technische Daten Technische Daten KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2 DE

 Open the catalog to page 6
UVD 2.1L-7

Technische Daten Präparative Durchflusszellen Technische Daten DE KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2

 Open the catalog to page 7
UVD 2.1L-8

Technische Daten Technische Daten Clamp Material KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2 DE

 Open the catalog to page 8
UVD 2.1L-9

Technische Daten Technische Daten Clamp Material Clamp Material Clamp Material DE KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2

 Open the catalog to page 9
UVD 2.1L-10

Technische Daten Technische Daten Clamp Material Clamp Material KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2 DE

 Open the catalog to page 10
UVD 2.1L-11

Austausch der Durchflusszelle Beim Benutzen von Lichtwellenleitern folgende Punkte beachten: Die Enden des Lichtwellenleiters nicht mit den Fingern berühren, da dies zu fehlerhaften Messungen führen könnte. Stöße und hartes Aufschlagen des Lichtwellenleiters vermeiden. Lichtwellenleiter vorsichtig ohne Druck und ohne Verbiegen bewegen. Isopropanol Verschlusskappen für die Durchflusszelle und Blindverschraubungen für die Lichtwellenleiter WARNUNG! Reizung der menschlichen Netzhaut durch UV-Licht. Gebündeltes UV-Licht kann an der Durchflusszelle oder am Lichtwellenleiter austreten. Gerät ausschalten...

 Open the catalog to page 11
UVD 2.1L-12

Durchflusszelle ausbauen Voraussetzungen Die Durchflusszelle wurde gespült. Das Gerät wurde ausgeschaltet. Isopropanol Verschlusskappen für die Durchflusszelle und Blindverschraubungen für die Lichtwellenleiter WARNUNG! Reizung der menschlichen Netzhaut durch UV-Licht. Gebündeltes UV-Licht kann an der Durchflusszelle oder am Lichtwellenleiter austreten. Gerät ausschalten und Netzstecker ziehen! Die Enden des Lichtwellenleiters nicht mit den Fingern berühren, da dies zu fehlerhaften Messungen führen könnte. KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2 DE

 Open the catalog to page 12
UVD 2.1L-13

Ablauf und Abbildung 1. Die Verschraubungen 1 der Lichtwellenleiter an der Durchflusszelle von Hand lösen. Abb. 1 Verschraubungen der Lichtwellenleiter an der Durchflusszelle 2. Die Blindverschraubungen 2 auf die Enden des Lichtwellenleiter aufsetzen. 3. Die Verschraubungen 3 der Lichtwellenleiter am Detektor von Hand lösen. Abb. 3 Verschraubungen der Lichtwellenleiter am Detektor DE KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2

 Open the catalog to page 13
UVD 2.1L-14

Ablauf und Abbildung 4. Die Verschlusskappen 4 auf die Durchflusszelle aufsetzen und die Durchflusszelle sicher aufbewahren. Nächste Schritte Lagern Sie die Durchflusszelle an einem geschützten Ort. Wenn die Durchflusszelle beschädigt wird, schicken Sie diese an den Hersteller zurück oder entsorgen Sie diese. Durchflusszelle lagern Voraussetzungen Die Durchflusszelle wurde gespült. Hilfsmittel Isopropanol Verschlusskappen für die Durchflusszelle Die Durchflusszelle zum Lagern mit Isopropanol füllen und mit Verschlusskappen KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2 DE

 Open the catalog to page 14
UVD 2.1L-15

verschließen. Dies verhindert, dass die Durchflusszelle verkeimt. Ablauf und Abbildung 1. Die Durchflusszelle mit Isopropanol füllen. Die Durchflusszelle mit Verschlusskappen 1 verschließen. Durchflusszelle einbauen Voraussetzungen Das Gerät wurde ausgeschaltet. Hilfsmittel Isopropanol Verschlusskappen für die Durchflusszelle WARNUNG! Reizung der menschlichen Netzhaut durch UV-Licht. Gebündeltes UV-Licht kann an der Durchflusszelle oder am Lichtwellenleiter austreten. Gerät ausschalten und Netzstecker ziehen! DE KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2

 Open the catalog to page 15
UVD 2.1L-16

Die Enden des Lichtwellenleiters nicht mit den Fingern berühren, da dies zu fehlerhaften Messungen führen könnte. Ablauf und Abbildung 1. Die Blindverschraubung 2 des Lichtwellenleiters und die Verschlusskappen 1 der Durchflusszelle abziehen. Die Verschlusskappen und die Blindverschraubung aufbewahren. KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2 DE

 Open the catalog to page 16
UVD 2.1L-17

Ablauf und Abbildung 2. Die Verschraubungen 3 des Lichtwellenleiters von Hand an der Durchflusszelle festziehen. Abb. 8 Verschraubungen der Lichtwellenleiter an der Durchflusszelle 3. Die Verschraubungen 4 des Lichtwellenleiters von Hand am Detektor festziehen. 4. Die Durchflusszelle an geeigneter Position platzieren. Abb. 9 Verschraubungen der Lichtwellenleiter am Detektor Nächste Schritte Prüfen Sie die Funktionsfähigkeit der Durchflusszelle (siehe ’Durchflusszelle prüfen’ auf Seite 19.). DE KNAUER Detektor UVD 2.1L Beiblatt V6829, Version 1.2

 Open the catalog to page 17
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.