Product catalog summary
Overview: The Smartline UV Detector 2520 is a versatile and compact HPLC spectrophotometer designed for routine HPLC applications, including fast LC methods. It offers excellent technical specifications and a flexible design, making it suitable for a wide range of applications.
Specifications: The detector features a wavelength range of 190 – 750 nm, with an optional range of 400 – 900 nm. It uses a deuterium light source, with a tungsten-halogen option, and provides a maximum data rate of 80 Hz via LAN or 10 Hz via RS-232. The optical bandwidth is 8 nm, with a wavelength accuracy of ± 2.5 nm and precision of 0.3 nm RSD.
Design and Features: The Smartline UV Detector 2520 is designed with a side-swinging front panel for easy access to the flow cell, which is protected from external influences. It is compact, making it one of the smallest detectors available. The detector supports a wide range of flow cells, including micro flow cells and preparative HPLC flow cells, and offers fiber optics for remote operation.
Control and Programming: The detector can be controlled via KNAUER ChromGate® or ClarityChrom® software, a front panel keypad, LAN, RS-232, or analog input/output. It supports programming for wavelength, events, valve switching, and includes a wake-up program for the lamp, with up to 19 programs and 99 lines.
Performance: The detector provides sensitive detection with automatic integration time selection and high-quality optical components. It features low noise and drift levels, with noise at ± 1.5 x 10-5 AU at 254 nm and drift at 3 x 10-4 AU/h at 254 nm.
Ordering Information: The Smartline UV Detector 2520 is available with a deuterium lamp, excluding the flow cell. Various analytical and preparative flow cells are available, with specifications including path length, volume, capillary connection, wetted materials, maximum flow rate, and pressure. Fiber optics versions are available for all flow cells except micro flow cells.
Contact Information: For more details, visit www.knauer.net or contact Wissenschaftliche Gerätebau Dr. Ing. Herbert Knauer GmbH in Berlin, Germany.
See more

Catalog excerpts

UV Detector 2520-1

Variable wavelength detector The Smartline UV Detector 2520 is a competitively priced HPLC spectrophotometer for routine HPLC applications including fast LC methods. Besides offering excellent technical specifications, this robust UV detector features a highly flexible and compact design. Sensitive detection – Automatic selection of optimal integration time as well as high quality optical components and electronics designed and manufactured in Germany guarantee optimal performance and high sensitivity. Practical features – The detector offers a wavelength range of 190 – 750 nm (400 – 900 nm optional). Built-in automatic lamp recognition improves performance and extends GLP data. Clever design – Positioned behind the side-swinging front panel, the flow cell is protected from external influences yet remains easily accessible. Its compact footprint makes the Smartline UV Detector one of the smallest detectors on the market. Wide applicability – A wide range of flow cells is available for practically any LC application, including micro flow cells for flow rates up to 10 µl/min and flow cells for preparative HPLC up to 10 l/min. Also available with fiber optics for remote flow cell operation. Flexible control – The Smartline UV Detector 2520 can be controlled with KNAUER ChromGate® or ClarityChrom® software, as well as from the front panel keypad (stand-alone operation), via LAN, via RS-232, or through analog input/output.

 Open the catalog to page 1
UV Detector 2520-2

Technical data Detector type Light source Number of signals Maximum data rate Wavelength range Optical bandwidth Wavelength accuracy Wavelength precision Noise* Drift* Linearity* Integration time Analog output Level/Event outputs Control Programming GLP function Display Power supply Dimensions / Weight monochromator spectrophotometer deuterium, tungsten-halogen optional 1 80 Hz (LAN) or 10 Hz (RS-232) 190 – 750 nm (deuterium) or 400 – 900 nm (tungsten-halogen) 8 nm ± 2.5 nm, auto-calibration with deuterium lines 0.3 nm RSD Standard version Fiber optics version ± 2.0 x 10-5 AU at 254 nm ± 1.5...