KT210 CT/PT Analyzer

KT210 CT/PT Analyzer

KT210 CT/PT Analyzer

Product catalog summary
Overview: The KT210 CT/PT Analyzer is a comprehensive testing device for current transformers (CTs) and potential transformers (PTs) used in electrical power systems. It is designed to test both protection and metering CTs according to IEEE and IEC standards.
Key Specifications:
  • Ratio measurement up to 10000:1.
  • Current ratio error and phase error measurement.
  • Winding resistance and excitation/saturation voltage current.
  • Secondary burden and saturated/unsaturated inductance.
  • Remanence flux factor, secondary time constant, and accuracy limiting factor.
  • Instrument security factor and dimensioning factor according to class PX, TPS.
  • Accuracy limiting voltage/current and turns ratio according to class PX.
  • Rated symmetrical short-circuit current factor and transient dimensioning factor.
  • Peak instantaneous error and maximum emf voltage.
  • Knee-point voltage/current and impedance/admittance of CT secondary load.
Protection and Metering CTs:
  • Protection CTs must be accurate during normal and fault conditions to prevent relay misoperation.
  • Metering CTs require high accuracy for correct billing and must saturate above nominal primary current to protect metering equipment.
Testing Capabilities:
  • CT-ratio and phase-angle accuracy with nominal and operational burden consideration.
  • CT winding resistance and excitation/saturation characteristics.
  • ALF and FS measurements, burden impedance, and CT residual magnetism.
  • PT ratio, polarity, and excitation curve testing.
Automated Testing Procedure:
  1. Measurement of CT parameters such as excitation curve and ratio.
  2. Modeling of CT elements with variable frequency.
  3. Automated assessment according to IEEE or IEC standards.
  4. Reporting via XML file and reporting tool.
Additional Features:
  • Auto demagnetization and PC control via RJ45 interface.
  • Data handling and reporting with customizable reports.
  • "Nameplate guesser" for unknown CT data.
  • Excellent noise immunity and small, lightweight design.
  • Automatic assessment for accuracy class > 0.1 and transient behavior measurement.
Technical Data:
  • Current ratio accuracy: 0.03% to 0.2% depending on ratio range.
  • Phase displacement accuracy: 1 min (typical) / 3 min (guaranteed).
  • Winding resistance range: 0.1 - 100 Ω with 0.05% accuracy.
  • Power supply: 176 Vac to 264 Vac, 50/60 Hz.
  • Output: 0 Vac to 120 Vac, 0 A to 5 A (15 A peak).
  • Physical dimensions: 360 x 140 x 325 mm, weight <8 kg.
  • Operating temperature: -10°C to +55°C, storage temperature: -25°C to +70°C.
Manufacturer: KINGSINE ELECTRIC AUTOMATION CO., LTD, a professional manufacturer of power testing products.
See more

Catalog excerpts

KT210 CT/PT Analyzer-1

KINGSINE ELECTRIC AUTOMATION CO., LTD A professional Manufacturer of Power Testing Products The KT210 CT/PT Analyzer is the first device with such capabilities.  Current ratio error and phase error for all measurement points defined in selected standard Winding resistance Excitation/saturation voltage current Secondary burden Remanence flux factor (Kr) Secondary time constant (Ts) Accuracy limiting factor (ALF / ALFi) Instrument security factor (FS / FSi) Dimensioning factor according to class PX, TPS (Kx) Accuracy limiting voltage/current according to class PX (Ek / Ie) Turns ratio according to class PX (N) Turns ratio and composite error (εt ,εc) Rated symmetrical short-circuit current factor (Kssc) Transient dimensioning factor (Ktd) Peak instantaneous error (ε^) Maximum emf voltage (Emax - calculated value) Accuracy limiting voltage/current (Val/Ial) The impedance / admittance of CT secondary Load, Like the burdens of various meters, relays, selector switches etc. are measured Current transformers are used for relaying and metering purposes in electrical power systems. They connect the high power primary side to the protection and metering equipment on the secondary side. Depending on the application they are used for, current transformers are designed differently. Protection current transformers As it is used to feed protective relays, the CT must be accurate during normal and fault It allows all types of single and multi-ratio current transformers to be tested on-site in conditions. Failures in transformation could lead to misoperation of the relay along with power system grids. Manufacturer of CTs, transformers or GIS use the KT210 in their production unwanted and costly outages. To test CTs according to the requirements of modern protection facilities and test / development labs. systems, it is compulsory to consider transient components and auto-reclosure systems. The KT210 offers a wide range of measurements, such as:  Metering current transformers CT-ratio and phase-angle accuracy with consideration of nominal and operational burden for various currents CTs for metering purposes must provide high accuracy up to class 0.1 to guarantee correct billing. It is therefore essential to test and calibrate the metering current transformer, as the entire metering chain is only as accurate as the instrument transformers feeding the meter. CT excitation / saturation (unsaturated and saturated) Burden impedance PT ratio, polarity and excitation curve In contrast to protection CTs, metering CTs must go into saturation directly above the nominal primary current level to protect the connected metering equipment. KT210 - a new way of testing CTs The KT210 is the most complete testing system for protection and metering CTs according Automated testing procedure The KT210 is designed to accurately measure all relevant CT parameters and compare them to the requirements of the defined IEEE or IEC standard. Due to this automated assessment, testing engineers receive the 'pass or fail' decision within seconds. Step 1: Measurement of parameters Measurement of CT parameters like Excitation curve Eddy current Ratio, etc Step 2: Modeling Definition of CT model elements with variable frequency and calculation of CT parameters through embedded mathematical functions. Step 3: Assessment according to IEEE or IEC standard Automated comparison of test results with the defined values according to the selected IEEE or IEC standard. Step 4: Reporting All data is delivered i

 Open the catalog to page 1
KT210 CT/PT Analyzer-2

IK I I ■ I I^U A fHroft^ion^bL tA(WAtf(MAv*rtr of Powtr Teytisu) Prod^oty Auto Demagnetizes V Software-based tool to determine the residual magnetism in current transformers V Analysis of the remanence condition before putting into operation the CT to assure proper function V S impl ifies power gr id fa ilure analysi s after unwanted operati on of protect ive relays V Demagnetizes the CT core after measurement PC Control Available V Full access to all functions of the KT210 via a PC using the RJ45 interface V Opti mizes the i ntegrat ion i nto automated testi ng procedures in product ion lines...