TOS9213AS

TOS9213AS

TOS9213AS

Product catalog summary
Overview: The TOS9213AS is a DC withstanding voltage and insulation resistance tester designed for testing photovoltaic (PV) modules. It complies with international standards IEC61215, IEC61646, and IEC61730-2, and supports insulation tests up to a maximum system voltage of 1500 V.
Key Features:
  • DC withstanding voltage test with a maximum output of 10 kV / 5 mA and 50 W power.
  • Insulation resistance test range from -25 V to -1500 V, with resistance measurement from 0.01 MΩ to 9.99 MΩ.
  • Low output ripple and rise time control for voltage build-up.
  • Judgment options based on resistance or current values.
  • Capability to apply high voltage and monitor current for Potential Induced Degradation (PID) testing.
IEC61730-2 Standard Test: The applied voltage varies by class. For Class A, voltage is raised at ≤500 V/s to [2000 V + 4 times the maximum system voltage] and maintained for 1 minute. Example: For a system voltage of 1500 V, the applied voltage is 8000 V.
Insulation Resistance and Wet Leakage Tests: Voltage is raised at ≤500 V/s to 500 V or the module's maximum system voltage, whichever is higher, and maintained.
PID Reproduction: The TOS9213AS can perform PID tests by using a negative polarity DC power source. The timer can be turned off for continuous testing.
Technical Specifications:
  • Output voltage range: 0.05V to 10.0kV with a resolution of 10V.
  • Accuracy: ±1.5% of setting + 20V.
  • Maximum rated load: 50W (10kV/5mA).
  • Short-circuit current: 40 mA typical.
  • Discharge function with adjustable time from 0.5 s to 300 s.
  • Voltmeter and ammeter with specified accuracy and response times.
General Specifications:
  • Power requirements: AC100V to 240V, selectable.
  • Power consumption: 100 VA (no load), 200 VA (rated load).
  • Operating temperature: 20°C to 40°C, humidity 20% to 80% RH.
  • Dimensions: 430(W) x 132(H) x 400(D) mm, weight approx. 13 kg.
Accessories: Includes AC power cord, high-voltage test leadwire, interlock jumper, high voltage danger sticker, fuse, and operation manual.
See more

Catalog excerpts

TOS9213AS-1

DC WITHSTANDING VOLTAGE/ INSULATION RESISTANCE TESTER TOS9213AS The maximum system voltage is the key for insulation test of PV (photovoltaic) modules. It can also be used to check for PID. The insulation tests in international standards IEC61215 Edition 2.0 2005-04, IEC61646 Edition 2.0 2008-05, and IEC61730-2 2004-10 are composed of 3 tests: the withstanding voltage test, insulation resistance test, and wet leakage test. The TOS9213AS is capable of insulation tests up to a maximum system voltage of 1500 V. Capable of tests up to a maximum system voltage of DC WITHSTANDING VOLTAGE/ INSULATION RESISTANCE TESTER IEC61730-2 standard test The applied voltage varies depending on the applicable class in IEC61730-2. For Class A, the voltage is raised at a speed of 500 V/s or less to [2000 V + 4 times of the maximum system voltage] and then maintained for 1 min. Example: With a mega-solar system that has a maximum system voltage of 1500 V, the applied voltage is [2000 + 4 x 1500] V = 8000 V. * (Excluding tester accuracy and similar factors) Image of insulation tests with the TOS9213AS PV module Frame Insulation resistance test and wet leakage test With each standard, the voltage is raised to a speed of 500 V/s or less to 500 V or the module maximum system voltage (whichever is higher) and maintained at that voltage. As the maximum system voltage in solar power generation projects continues to trend higher, the TOS9213AS can raise the DC voltage that is applied at the insulation resistance test up to 1500 V. Electrode Connect the HIGH terminal to the short-circuited output electrodes. Connect the LOW terminal to the frame. Reproduction of PID When performing tests for PID (Potential Induced Degradation), which is a problem with current PV modules, a negative polarity DC power source is required. By turning OFF the timer setting in the TOS9213AS insulation resistance test mode, it is possible to perform these tests with the TOS9213AS in the same way.(Maximum current capacity: 1 mA up to 1020 V, 0.1 mA at or above 1020 V) Features/ Functions 10 kV / 5 mA, maximum output power of 50 W in DC withstanding voltage test. -25 V to -1500 V / 0.01 M 9.99 M insulation resistance test. Conforms to international standards including IEC61215 Ed2.0, IEC61646 Ed2.0, and IEC61730-2. Low output ripple of (100 Vp-p at 10 kV) in consideration of capacitive load. TOS9213AS * The module connection method for PID reproduction is the same as for the insulation tests. By selecting the TOS9213AS insulation resistance test mode (IR), the applied voltage will have negative polarity. The rise time control function allows the voltage build-up rate to be set. Also includes a discharge function. Judgment of the insulation resistance test can be selected between resistance value and current value. Capable to apply high-voltage and monitor current for PID. (-1500 VDC / 100 µA)

 Open the catalog to page 1
TOS9213AS-2

Specifications ■ DC Withstanding Voltage Test Mode ■ Insulation Resistance Test Mode Output sectionDC Output section Output-voltage range Output-voltage range Maximum rated load ※1 Maximum rated load Maximum rated current Maximum rated current 1mA0.1 mA maximum when the test voltage setting exceeds -1020 V Maximum rated load 100Vp-p Typ. No load at 1 kV 2 Vp-p or less Maximum rated load 10 Vp-p or less Voltage regulation 1 % or less [maximum rated load → no load] Voltage regulation 1% or less [Maximum rated load → no load] Short-circuit current Short-circuit current Discharge function Forced...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.