N9030A PXA Signal Analyzer
32Pages

{{requestButtons}}

Catalog excerpts

N9030A PXA Signal Analyzer - 1

Agilent PXA Signal Analyzer N9030A Data Sheet class C certified Available frequency ranges N9030A-503 3 Hz to 3.6 GHz N9030A-508 3 Hz to 8.4 GHz N9030A-513 3 Hz to 13.6 GHz N9030A-526 3 Hz to 26.5 GHz

Open the catalog to page 1
N9030A PXA Signal Analyzer - 3

3 This PXA signal analyzer data sheet is a summary of the complete specifications and conditions. The complete PXA Signal Analyzer Specification Guide can be obtained from the web at: www.agilent.com/find/pxa_specifications Defi nitions and Conditions Specifications describe the performance of parameters covered by the product warranty and apply to temperature ranges 0 to 55 °C for the solid-state drive (Option SSD), or 5 to 50 ºC for the hard disk drive (standard), unless otherwise noted. 95th percentile values indicate the breadth of the population (approx. 2 ó) of performance tolerances...

Open the catalog to page 3
N9030A PXA Signal Analyzer - 4

4 Frequency and Time Specifi cations Frequency range DC coupled AC coupled Option 503 3 Hz to 3.6 GHz 10 MHz to 3.6 GHz Option 508 3 Hz to 8.4 GHz 10 MHz to 8.4 GHz Option 513 3 Hz to 13.6 GHz 10 MHz to 13.6 GHz Option 526 3 Hz to 26.5 GHz 10 MHz to 26.5 GHz Band LO Multiple (N) DC coupled 0 1 3 Hz to 3.6 GHz 1 1 3.5 to 8.4 GHz 2 2 8.3 to 13.6 GHz 3 2 13.5 to 17.1 GHz 4 4 17 to 26.5 GHz Precision frequency reference Accuracy ± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy] Aging rate ± 1 x 10–7/year ± 1.5 x 10–7/2 years Temperature stability 20 to...

Open the catalog to page 4
N9030A PXA Signal Analyzer - 5

5 Frequency and Time Specifi cations (continued) Frequency span (FFT and swept mode) Range 0 Hz (zero span), 10 Hz to maximum frequency of instrument Resolution 2 Hz Accuracy Swept FFT ± (0.1% x span + horizontal resolution) ± (0.1% x span + horizontal resolution) Sweep time and triggering Range Span = 0 Hz Span 10 Hz 1 ìs to 6000 s 1 ms to 4000 s Accuracy Span 10 Hz, swept Span 10 Hz, FFT Span = 0 Hz ± 0.01% nominal ± 40% nominal ± 0.01% nominal Sweep trigger Free run, line, video, external 1, external 2, RF burst, periodic timer Trigger delay Span = 0 Hz or FFT Span 10 Hz, swept...

Open the catalog to page 5
N9030A PXA Signal Analyzer - 6

6 Frequency and Time Specifi cations (continued) Analysis bandwidth1 Maximum bandwidth Standard Option B25 Option B40 Option B1X 10 MHz 25 MHz 40 MHz 140 MHz Video bandwidth (VBW) Range 1 Hz to 3 MHz (10% steps), 4, 5, 6, 8 MHz and wide open (labeled 50 MHz) Accuracy ± 6% nominal (in swept mode and zero span) Measurement speed 2 Local measurement and display update rate 10 ms (100/s) nominal Remote measurement and LAN transfer rate 10 ms (100/s) nominal Marker peak search 2.5 ms nominal Center frequency tune and transfer (RF) 43 ms nominal Center frequency tune and transfer (ìW) 69 ms...

Open the catalog to page 6
N9030A PXA Signal Analyzer - 7

7 Amplitude Accuracy and Range Specifi cations Amplitude range Measurement range Displayed average noise level (DANL) to maximum safe input level Input attenuator range (3 Hz to 26.5 GHz) 0 to 70 dB in 2 dB steps Electronic attenuator (Option EA3) Frequency range 3 Hz to 3.6 GHz Attenuation range Electronic attenuator range Full attenuation range (mechanical + electronic) 0 to 24 dB, 1 dB steps 0 to 94 dB, 1 dB steps Maximum safe input level Average total power +30 dBm (1 W) With or without preamp Peak pulse power 10 ìs pulse width, 1% duty cycle and input attenuation 30 dB +50 dBm (100 W)...

Open the catalog to page 7
N9030A PXA Signal Analyzer - 8

8 Amplitude Accuracy and Range Specifi cations (continued) Frequency response (10 dB input attenuation, 20 to 30 °C, preselector centering applied at 3.6 GHz and above) 95th percentile (~2 ó) 3 Hz to 10 MHz ± 0.46 dB ± 0.19 dB 10 MHz to 3.6 GHz ± 0.35 dB ± 0.16 dB 3.5 to 8.4 GHz ± 1.5 dB ± 0.39 dB 8.3 to 13.6 GHz ± 2.0 dB ± 0.45 dB 13.5 to 22.0 GHz ± 2.0 dB ± 0.62 dB 22.0 to 26.5 GHz ± 2.5 dB ± 0.82 dB Preamp on (Option P03, P08, P13, P26) 9 to 100 kHz ± 0.36 dB attenuation 0 dB 100 kHz to 50 MHz ± 0.68 dB ± 0.26 dB 50 MHz to 3.6 GHz ± 0.55 dB ± 0.28 dB 3.5 to 8.4 GHz ± 2.0 dB ± 0.64 dB 8.3...

Open the catalog to page 8
N9030A PXA Signal Analyzer - 9

9 Amplitude Accuracy and Range Specifi cations (continued) Resolution bandwidth switching uncertainty (referenced to 30 kHz RBW) 1 Hz to 1.5 MHz RBW ± 0.03 dB 1.6 MHz to 2.7 MHz RBW ± 0.05 dB 3 MHz RBW ± 0.10 dB 4, 5, 6, 8 MHz RBW ± 0.30 dB Reference level Range Log scale Linear scale –170 to +30 dBm in 0.01 dB steps 707 pV to 7.07 V with 0.11% (0.01 dB) resolution Accuracy 0 dB Display scale switching uncertainty Switching between linear and log 0 dB Log scale/div switching 0 dB Display scale fidelity Between –10 dBm and –18 dBm input mixer level ± 0.10 dB ± 0.04 dB typical below –18 dBm...

Open the catalog to page 9
N9030A PXA Signal Analyzer - 10

10 Dynamic Range Specifi cations 1 dB gain compression (two-tone) At 1 kHz RBW with 100 kHz tone spacing, 20 to 30 °C Maximum power at input mixer 20 to 40 MHz –3 dBm 0 dBm typical 40 to 200 MHz +1 dBm +3 dBm typical 200 MHz to 3.6 GHz +3 dBm +5 dBm typical 3.6 to 16 GHz +1 dBm +4 dBm typical 16 to 26.5 GHz –1 dBm +2 dBm typical Preamp on (Option P03, P08, P13, P26) 10 MHz to 3.6 GHz –14 dBm nominal 3.6 to 26.5 GHz Tone spacing 100 kHz to 20 MHz – 28 dBm nominal Tone spacing > 70 MHz – 10 dBm nominal Displayed average noise level (DANL) (Input terminated, sample or average detector,...

Open the catalog to page 10
N9030A PXA Signal Analyzer - 11

11 Dynamic Range Specifi cations (continued) DANL with Noise Floor Extension (NFE) on 95th Percentile Improvement for noise-like signals Preamp Off Preamp On Band 0, f > 20 MHz Band 1 Band 2 Band 3 Band 4 8.5 dB 4 dB 7.5 dB 7 dB 6 dB 8.5 dB 7 dB 7 dB 7.5 dB 6 dB Examples of effective DANL Frequency 20 to 30 °C Mid-Band 0 (1.8 GHz) Mid-Band 1 (5.95 GHz) Mid-Band 2 (10.95 GHz) Mid-Band 3 (15.3 GHz) Mid-Band 4 (21.75 GHz) Preamp Off –163 dBm –158 dBm –157 dBm –153 dBm –145 dBm Preamp On –172 dBm –172 dBm –170 dBm –166 dBm –162 dBm Residues, images, and spurious responses Residual responses...

Open the catalog to page 11
N9030A PXA Signal Analyzer - 12

12 Dynamic Range Specifi cations (continued) Second harmonic distortion (SHI) Source frequency 10 to 100 MHz 0.1 to 1.8 GHz 1.75 to 2.5 GHz 2.5 to 4 GHz 4 to 6.5 GHz 6.5 to 10 GHz 10 to 13.25 GHz Mixer level –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm Distortion* –57 dBc/NA –60 dBc/NA –77 dBc/–95 dBc –77 dBc/–101 dBc –77 dBc/–105 dBc –70 dBc/–105 dBc –62 dBc/–105 dBc SHI* +42 dBm/NA +45 dBm/NA +62 dBm/+80 dBm +62 dBm/+86 dBm +62 dBm/+90 dBm +55 dBm/+90 dBm +47 dBm/+90 dBm Preamp on (Option P03, P08, P13, P26) Preamp level Distortion SHI 10 MHz to 1.8 GHz 1.8 to 13.25 GHz –45 dBm...

Open the catalog to page 12