M9182A Digital Multimeter, 6½ digit, PXI
14Pages

{{requestButtons}}

Catalog excerpts

M9182A Digital Multimeter, 6½ digit, PXI - 1

Agilent PXI Digital Multimeters 6½ Digit, High Performance M9182A M9183A • 6½ digit resolution • High throughput: up to 20,000 rdgs/sec, 50 ìs single reading interval • DCV basic one year accuracy: 40 ppm • Nine standard measurements: DCV, ACV, DCI, ACI, 2– and 4-wire resistance, frequency/ period, capacitance, and temperature

Open the catalog to page 1
M9182A Digital Multimeter, 6½ digit, PXI - 2

OVERVIEW Product Description Agilent’s M9182A and M9183A 6½ digit high performance PXI digital multimeters offer fast development, high throughput and trustworthy results. The M9182A provides nine built-in measurement types with all the accuracy and stability you would expect from an Agilent 6½ digit DMM. The M9183A provides the same capabilities as the M9182A, with market-leading measurement speed of up to 20,000 readings per second, additional ranges, and a DC source. Industries and Applications • Aerospace and defense • Automotive electronics test • Industrial electronics test • Medical...

Open the catalog to page 2
M9182A Digital Multimeter, 6½ digit, PXI - 3

Hardware Platform Compliance The M9182A and M9183A 6½ digit, digital multimeters (DMMs) are PXI compliant, using either a cPCI (J1), PXI- 1 (J1) or PXIe Hybrid slot. Designed to benefit from fast data interfaces, an M9182A or M9183A DMM can be integrated with other test and automation modules in a PXI, CompactPCI, or Hybrid chassis. The PXI format offers high performance in a small, rugged package. It is an ideal deployment platform for many automated test systems. A wide array of complementary PXI products are currently available. Products from Agilent include switches, multiplexers,...

Open the catalog to page 3
M9182A Digital Multimeter, 6½ digit, PXI - 6

Accuracy specifications continued ±(% of reading + % of range)1,2 Function Range or resolution Frequency or full scale reading 24 hour 23 °C ± 5 °C 90 day 23 °C ± 5 °C 1 year 23 °C ± 5 °C Duty cycle (typ) [M9183A only] 0.02 % 0.20 % 2.00 % 20.0 % 2 Hz - 100 Hz 100 Hz - 1 kHz 1 kHz - 10 kHz 10 kHz - 100 kHz 0.03 +0.03 0.03 +0.30 0.03 +3.00 0.03 +20.0 0.03 +0.03 0.03 +0.30 0.03 +3.00 0.03 +20.0 0.03 +0.03 0.03 +0.30 0.03 +3.00 0.03 +20.0 Pulse width (typ) [M9183A only] 2 ìs - 1 s .5 Hz - 100 kHz .01 ± 4 ìs 01 ± 4 ìs 01 ± 4 ìs Capacitance11 [M9183A and M9182A] 1000.0 pF 10.000 nF 100.00 nF...

Open the catalog to page 6
M9182A Digital Multimeter, 6½ digit, PXI - 7

TECHNICAL SPECIFICATI ONS AND CHARACTERISTICS, CONTINUED M9182A & M9183A Temperature accuracy (spec)1 Temperature Function Type R0 (Ù) Sensitivity Range/max temperature 1 Year 23 °C ± 5 °C RTD Temperature Measurement2,3 pt385 100 Ù, 200 Ù 0.01 °C -150 to 650 °C ± 0.06 °C 500 Ù, 1 kÙ 0.01 °C -150 to 650 °C ± 0.03 °C Less than 12 Ù 0.01 °C -100 to 200 °C ± 0.18 °C at 20 °C ± 0.05 °C otherwise Higher than 90 Ù 0.01 °C -100 to 200 °C ± 0.10 °C at 20 °C ± 0.05 °C otherwise Thermocouple Temperature Measurement4,5 B NA 0.01 °C 2200 °C ± 0.38 °C E NA 0.01 °C 1200 °C ± 0.035 °C J NA 0.01 °C 2000 °C...

Open the catalog to page 7
M9182A Digital Multimeter, 6½ digit, PXI - 8

1. Specifications are for one hour warm up, within one hour self-cal, slow AC filter 2. For temperatures outside the range of 23 °C ± 5 °C, but within 0 °C to 50 °C, add 0.1 × accuracy specification per °C 3. Repetitive reading at an aperture of 133 ms or higher M9183A Source DC Voltage, Measure DC Voltage (typ) Parameter Extended Resolution Basic Resolution DC voltage source (output) range -10.000 to +10.000 V DC current source/sink at 5 V output 5 mA DAC resolution 18 bits 12 bits DC voltage measurement accuracy 1 year, (23 °C ± 5 °C)11,2,3 0.015% ± 0.004% 1.0% ± 0.4% Typical settling...

Open the catalog to page 8
M9182A Digital Multimeter, 6½ digit, PXI - 9

TECHNICAL SPECIFICATI ONS AND CHARACTERISTICS, CONTINUED Trigger modes Analog threshold trigger (Pre-trigger or post-trigger) Trigger point Selectable threshold and positive or negative edge transition Buffer type Circular Captures 80 post-trigger readings Aperture range 2.5 ìs to 160 ms Read interval range 1/aperture to 65 ms Post-trigger readings Selectable from 0 to 80 Pre-trigger readings Selectable from 0 to 80 Trigger delay (Default values ensure 1st reading accuracy) Measurement delay 50 ìs to 1 s Resolution 1 ìs to 64 ms, 20 ìs to 1 s External hardware trigger Trigger input voltage...

Open the catalog to page 9
M9182A Digital Multimeter, 6½ digit, PXI - 10

Transaction Speed Transactional I/O speed is a single reading measurement. This is important when you are taking many single measurements with the DMM. The M9183A delivers the highest transactional measurement speed in its class. These fast readings, up to 20,000 readings per second with a read interval rate of 50 ìs, translate into higher test-system throughput and lower cost of test per unit tested. Resolution vs. Aperture and Reading Rate for DCV, DCI, Ù Measurement aperture Maximum readings per second Resolution 10 ms 98 6½ digits (22 bits) 625 ìs 1,200 5½ digits (18 bits) 130 ìs 4,500...

Open the catalog to page 10
M9182A Digital Multimeter, 6½ digit, PXI - 11

DC Voltage Measurement method Delta-Sigma A/D conversion Input resistance 200 mV, 2.0 V Ranges: >10 GÙ with typical leakage of < 50 pA; 20 V, 200 V, 300 V Ranges: 10.0 MÙ Input isolation 330 VDC, 250 VAC from Earth Ground Input overvoltage protection 330 VDC all ranges DCV noise rejection Normal mode rejection at 50, 60, or 400 Hz ± 0.5%; > 95 dB (apertures 0.160 s); CMRR (1 kÙ lead imbalance) 120 dB True RMS AC Voltage Measurement method AC coupled (10 Hz to 100 kHz) true RMS — measures the AC component only Analog RMS DC converter Crest factor Maximum crest factor of 4 at full scale, 7 at...

Open the catalog to page 11
M9182A Digital Multimeter, 6½ digit, PXI - 12

Environmental and Physical Characteristics Temperature range Operating -10° to 55 °C Non-operating -40 ° to +85 °C Relative humidity Operating to 80% at 40 °C Storage to 95% at 40 °C Connectors V HI, 2-wire Ù IN , DCV OUT Sheathed Banana Jack V LO, 2-wire Ù IN , DCV OUT Sheathed Banana Jack I HI, 4-wire Ù IN Sheathed Banana Jack I LO, 4-wire Ù IN Sheathed Banana Jack Sync OUT DIN 7, pin 2 External Trigger IN DIN 7, pin 7 Trigger & Sync common DIN 7, pin 4 Safety Complies with IEC 61010-1, Cat II 300 V, Pollution degree 2 EMC Complies with EN61326-1 Industrial Environment Warm-up time One...

Open the catalog to page 12

All KEYSIGHT TECHNOLOGIES catalogs and technical brochures

  1. RF Products

    20 Pages

  2. KeysightCare

    8 Pages

  3. I/O Hardware

    18 Pages

  4. X-Series

    23 Pages

  5. CXA X-Series

    8 Pages

  6. Oscilloscopes

    16 Pages