E1333A 3-Channel Universal Counter
4Pages

{{requestButtons}}

Catalog excerpts

E1333A 3-Channel Universal Counter - 1

Data Sheet 3-Channel Universal Counter Agilent E1333A • 1-Slot, B-size, register based • Six counter functions • 1 GHz frequency measurement • 1 ns time interval/pulse width resolution (avg mode) • Compatible with all Agilent Foundations • Measure time interval between transitions Agilent E1333A Description The Agilent Technologies E1333A 3-Channel Universal Counter is a B-size, 1-slot, register-based VXI module. It provides the capabilities needed for electronic test applications. You can connect three signals to one counter card and multiplex between them to measure frequency. Counter functions include frequency, period average, pulse width and pulse width average, time interval and time interval average, totalize, and ratio. You can select the resolution directly in seconds, the number of periods to be averaged, the number of intervals (2n to be averaged), or the number of transitions (2n on one channel). Additionally, you can measure the time interval between transitions from one channel to another channel, or measure the frequency ratio between channels 1 and 2 or channels 2 and 1. To complete the E1333A module’s functionality, you may also select the rising or falling edge via software commands, and count the number of transitions on channels 1 and 2. Input Signal Conditioning commands control all channels simultaneously. Trigger level/ sensitivity commands are available for each channel. Refer to the Agilent Technologies Website for instrument driver availability and downloading instructions, as well as for recent product updates, if applicable.

Open the catalog to page 1
E1333A 3-Channel Universal Counter - 2

2 • Minimum interval: 200 ns • Maximum interval: (6871/2N) s • Range of N: 0 to 7 • Resolution: (100/2N) ns Totalizing You can count the number of transitions on channels 1 and 2. • Minimum pulse width: 5 ns • Range: 1 to (236-1) Frequency Ratio Measurement You can measure the frequency ratio between channels 1 and 2 or channels 2 and 1. You select the resolution directly or the number of transitions, 2N on one channel. • Minimum pulse width: 5 ns • Range of N: 6 to 36 • Resolution: 12 N Input Signal Conditioning commands control all channels simultaneously. Trigger level/sensitivity...

Open the catalog to page 2
E1333A 3-Channel Universal Counter - 3

3 Sensitivity Low input range (dc to 100 MHz): 25 mV High input range (dc to 100 MHz): 250 mV Input range: (select the input attenuator) Low range: ± 5 V High range: ± 42 V Trigger level: Low input range: –2.56 V to 2.54 V in 20 mV step High input range: –25.6 V to 25.4 V in 0.2 V step Input attenuator: Programmable x1 or x10 attenuator Filter: Programmable Low-Pass Filter, 3 dB point at 100 kHz Input impedance (typical): Programmable 1 M shunted by 50 pF or 50 Coupling: Programmable ac or dc coupled Frequency range: ac coupled: 100 Hz to 100 MHz dc coupled: dc to 100 MHz General...

Open the catalog to page 3
E1333A 3-Channel Universal Counter - 4

Agilent Technologies’ Test and Measurement Support, Services, and Assistance Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the production life of the product. Two...

Open the catalog to page 4

All KEYSIGHT TECHNOLOGIES catalogs and technical brochures

  1. RF Products

    20 Pages

  2. KeysightCare

    8 Pages

  3. I/O Hardware

    18 Pages

  4. X-Series

    23 Pages

  5. CXA X-Series

    8 Pages

  6. Oscilloscopes

    16 Pages

Archived catalogs