1. Catalogs
  2. Keithley Instruments
  3. Digital Multimeters Tektronix DMM4020 Datasheet
video corpo

Digital Multimeters Tektronix DMM4020 Datasheet

Digital Multimeters Tektronix DMM4020 Datasheet
1 / 12 PagesView full catalog

Digital Multimeters Tektronix DMM4020 Datasheet

Product catalog summary
Features & Benefits
The Tektronix DMM4020 is a high-precision digital multimeter with a 5.5 digit resolution. It offers a basic VDC accuracy of up to 0.015% across a voltage range of 200 mV to 1000 V, with a resolution as fine as 1 μV. It measures current from 200 μA to 10 A with a resolution up to 1 nA, and resistance from 200 Ω to 100 MΩ with a resolution up to 1 mΩ. The device is rated for CAT I 1000 V and CAT II 600 V.
Available Functions and Features
The DMM4020 supports various measurements including volts, ohms, amps, true RMS (AC, AC + DC), diode and continuity testing, frequency measurements, and a 2×4 ohms 4-wire measurement technique. It features dedicated DC leakage current measurements, six setup buttons for quick access, and a limit compare mode for pass/fail testing.
Connectivity
The multimeter includes front 2×4 measurement inputs and an RS-232 port for PC connectivity, with a USB to RS-232 interface adapter cable. It is compatible with National Instrument’s LabVIEW SignalExpress™ TE Limited Edition and comes with a 3-year warranty.
Ease of Use
Designed for intuitive operation, the DMM4020 has dedicated front-panel buttons for fast function access, reducing setup time. It features a dual display for simultaneous measurement of two parameters and a limit compare mode for quick pass/fail results.
Measurement Capabilities
The multimeter can accurately measure nanoampere signals using a current-to-voltage op amp technique, ensuring minimal circuit loading.
Specifications
  • General Specifications: Voltage settings range from 100V to 240V, frequency from 47 Hz to 440 Hz, with a power consumption of 15 VA peak (10 W average). Dimensions are 88 mm x 217 mm x 297 mm, and it weighs 2.1 kg.
  • Environmental Specifications: Operating temperature is 0 °C to 50 °C, storage temperature is –40 °C to 70 °C, and it complies with various safety and EMC standards.
  • Electrical Specifications: DC voltage accuracy is specified for various ranges with input protection up to 1000 V. AC voltage specifications include true RMS measurement with a bandwidth of 20 Hz to 100 kHz.
  • Resistance and Current Measurements: Supports 4-wire resistance measurements with detailed accuracy specifications for DC and AC current measurements.
  • Frequency Measurement: Frequency measurements are AC-coupled with a gate time of 131 ms, with specific considerations for accuracy.
Specifications:
1. Continuity Test: Threshold of 20 Ω, Test Current of 1 mA, Response Time of 100 S/s with audible tone, Fast Rate, Maximum Reading of 199.99 Ω, Resolution of 0.01 Ω.
2. Diode Test: Response Time of 100 S/s with audible tone, Fast Rate, Maximum Reading of 1.9999 V, Resolution of 0.1 mV.
Product Description:
The Tektronix DMM4020 includes a meter, TL710 Test Leads, Line Cord, Statement of Calibration Practices, Safety and Installation Instructions, CD-ROM with User Manual in multiple languages, RS-232 to USB Adapter Cable, and National Instruments LabVIEW SignalExpress™ Tektronix Edition Software.
Ordering Information:
Power plug options include North America, Universal Euro, United Kingdom, Australia, Switzerland, Japan, China, India, Brazil, and Euro and UK power cords. Service options include single calibration events, traceable calibration certificates, and a 5-year repair service.
Recommended Accessories:
Includes Calibration Manual, Premium Test Leads, Precision Test Lead, SMD Test Tweezers, Soft and Hard Carrying Cases, Rackmount Shelf Kit, Calibration Fixture, and Full Version of SignalExpress Software.
Compliance and Standards:
Tektronix is registered to ISO 9001 and ISO 14001. The product complies with IEEE Standard 488.1-1987 and RS-232C.
Contact Information:
Contact details for various regions including ASEAN, Australasia, Europe, Americas, and others are provided for customer support.
Additional Information:
For further resources, visit the Tektronix website. The document notes that specifications and prices are subject to change.
See more

Catalog excerpts

Digital Multimeters Tektronix DMM4020 Datasheet-1

Digital Multimeters Features & Benefits ■ 200 mV to 1000 V Voltage Range, with up to 1 uV Resolution ■ 200 uA to 10 A Current Range, with up to 1 nA Resolution ■ 200 Q to 100 MQ Range, with up to 1 mQ Resolution Available Functions and Features ■ Volts, Ohms, and Amps Measurements ■ Diode and Continuity Testing ■ 2x4 Ohms 4-wire Measurement Technique ■ Dedicated DC Leakage Current Measurements ■ Six Dedicated Buttons for Fast Access to Instrument Setups ■ Limit Compare Mode for Pass/Fail Testing ■ Front 2x4 Measurement Inputs ■ RS-232 on Rear Panel for Quick PC Connectivity ■ Includes USB to RS-232 Interface Adapter Cable ■ Includes National Instrument's LabVIEW SignalExpress™ TE Limited Edition for Connecting Your Bench

 Open the catalog to page 1
Digital Multimeters Tektronix DMM4020 Datasheet-2

Measurements with the Push of a Button As the circuits in embedded system designs become more sophisticated, you must measure a multitude of different parameters to validate your design. The Tektronix DMM4020 5.5 digit bench multimeter offers a broad range of functions in one easy-to-use instrument. Typical multimeter measurements – volts, ohms, and amps – are made with a basic VDC accuracy of up to 0.015%, ensuring you have the performance you need for your design. You can also use the DMM4020 to measure frequency, and to perform continuity and diode tests. This allows you to replace your counter,...

 Open the catalog to page 2
Digital Multimeters Tektronix DMM4020 Datasheet-3

Digital Multimeters — Tektronix DMM4020 SignalExpress acquiring data from Tektronix DMM4020 and DPO3052. Connect Your Bench for Intelligent Debug Easily capture, save, and analyze measurement results from your multimeter with the special Tektronix Edition of National Instrument's LabVIEW SignalExpress™ software. Every DMM4020 multimeter ships with a free copy of the Limited Edition version of SignalExpress for basic instrument control, data logging, and analysis. The optional Professional Edition offers over 200 built-in functions that provide additional signal processing, advanced analysis,...

 Open the catalog to page 3
Digital Multimeters Tektronix DMM4020 Datasheet-4

Characteristics General Specications Triggering Voltage Characteristic 100V Setting 120V Setting 220V Setting 240V Setting Frequency Power Consumption Trigger Delay External Trigger Delay External Trigger Jitter Trigger Input Trigger Output Math Functions Min/Max, relative, hold, compare, and dB functions Dimensions Dimension Height Width Depth Multiple Measurement Combinations Secondary Function DC V AC V DC I Vacuum Fluorescent Display, segment Environment Characteristic Temperature Operating 0 °C to 50 °C Storage –40 °C to 70 °C Warm Up ½ hour to full uncertainty specications Relative Humidity...

 Open the catalog to page 4
Digital Multimeters Tektronix DMM4020 Datasheet-5

Digital Multimeters — Tektronix DMM4020 Electrical Specifications Specifications are valid for 5% digit mode and after at least a half-hour warm up. Characteristic Description Maximum Input 1000 V on any range Common Mode 120 dB at 50 or 60 Hz 0.1% (1 kQ unbalance) Normal Mode Rejection 80 dB at Slow Rate Input Bias Current <30 pA at 25 °C_ Settling Considerations Measurement settling times are affected by source impedance, cable dielectric characteristics, and input _signal changes_ Input Characteristics Range Full Scale Resolution Input Impedance (5y2 Digits) S|ow Medium Fast *3 Uncertainty...

 Open the catalog to page 5
Digital Multimeters Tektronix DMM4020 Datasheet-6

AC Voltage specifications are for AC sinewave signals >5% of range. For inputs from 1% to 5% of range and <50 kHz, add an additional error of 0.1% of range, and for Maximum Input 750 VRMS or 1000 V peak or 8 * 107 Volts-Hertz product Measurement Method AC-coupled true RMS. Measures the AC component of input with up to 1000 VDC bias on any range_ Common Mode Maximum Crest Factor 3:1 at Full Scale Additional Crest Factor Crest Factor 1-2, 0.05% of full scale Crest Factor 2-3, 0.2% of full scale Only applies for non-sinusoid signals Input Characteristics Range Full Scale Resolution Input Impedance...

 Open the catalog to page 6
Digital Multimeters Tektronix DMM4020 Datasheet-7

Digital Multimeters — Tektronix DMM4020 Resistance Specications are for 4-wire resistance function, or 2-wire resistance with REL. If REL is not used, add 0.2 Ω for 2-wire resistance plus lead resistance. Characteristic Measurement Method Max Lead Resistance (4-wire ohms) Input Protection Current source referenced to LO input 10% of range per lead for 200 Ω, 2 kΩ ranges. 1 kΩ per lead on all other ranges 1000 V on all ranges Input Characteristics Range Full Scale (5½ Digits) Current Source Accuracy Range Temperature Coefcient/°C Outside 18 - 28 °C *3 Uncertainty given as ±(% of reading + % of...

 Open the catalog to page 7
Digital Multimeters Tektronix DMM4020 Datasheet-8

Characteristic Description Input Protection Tool-accessible 11 A / 1000 V and 440 mA/ 1000 V Shunt Resistance 0.01 Q for 2 A and 10 A ranges _Burden voltage <5 mV for 200 uA and 2 mA range Input Characteristics Range Full Scale Resolution Burden Voltage (5y2 Digits) Slow Medium Fast *3 Uncertainty given as ±(% of reading + % of range).

 Open the catalog to page 8
Digital Multimeters Tektronix DMM4020 Datasheet-9

Digital Multimeters — Tektronix DMM4020 The following AC current specifications are for sinusoidal signals with amplitudes greater than 5% of range. For inputs from 1% to 5% of range, add an additional Characteristic Description Input Protection Tool-accessible 11 A / 1000 V and 440 mA/ 1000 V Measurement Method AC-coupled true RMS Shunt Resistance 0.01 Q for 2 A and 10 A ranges Maximum Crest Factor 3:1 at Full Scale Additional Crest Factor Crest Factor 1-2, 0.05% of full scale Errors (<100 Hz) Crest Factor 2-3, 0.2% of full scale _Only applies to non-sinusoid signals_ Input Characteristics Range...

 Open the catalog to page 9
Digital Multimeters Tektronix DMM4020 Datasheet-10

Frequency Characteristic Gate Time Measurement Method AC-coupled input using the AC voltage measurement function Settling Considerations When measuring frequency after a DC offset voltage change, errors may occur. For the most accurate measurement, wait up to 1 second to allow input-blocking RC time constant to settle Measurement Considerations To minimize measurement errors, shield inputs from external noise when measuring low-voltage, low-frequency signals Input Characteristics Range Temperature Coefcient/°C Outside 18 - 28 °C Diode Test Continuity Threshold Test Current Response Time Rate...

 Open the catalog to page 10

Archived catalogs

  1. 6 Series B MSO

    69  Pages

  2. 3 Series MDO

    36  Pages

  3. 4 Series MSO

    40  Pages

  4. TSG4100A Series

    24  Pages

  5. RTPA2A

    6  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.