8 Series Sampling Oscilloscope
14Pages

{{requestButtons}}

Catalog excerpts

8 Series Sampling Oscilloscope - 1

8 Series Sampling Oscilloscope TSO820 and TSO8C17/18 Datasheet Key features • • The 8 Series Sampling Oscilloscope provides a comprehensive optical test solution for telecom and datacom applications, as well as general purpose optical component testing. The TSO8C17 and TSO8C18 optical modules provide > 30 GHz optical bandwidth, plus fully integrated Optical Reference Receivers (ORR)1 enabling both single mode and multi-mode conformance testing at 850 nm, 1310 nm, and 1550 nm bands. Optical bandwidth above 30 GHz Single mode and multi-mode support for short and long wavelength optical testing Optical Reference Receiver (ORR)1 support for standard compliance testing Design/verification of High-Speed Components and Systems Signal integrity analysis Compliance test for industry PAM4 standards: 50G, 100G, 200G, 400G IEEE 802.3TM standanrd such as 400GBASE-FR8, 400GBASEDR4, and similar optical direct detect standards. Accurate testing and characterization of short or long wave optical signals using the high sensitivity and low noise performance of the TSO8C17 or TSO8C18 modules Optical reference receivers (ORR)1 supports specified requirements for standards-mandated compliance testing Extinction ratio measurements with built-in variable ER correction to ensure accuracy and repeatability Analysis with TSOVu® • Optical modules • Disaggregated: Separation of the acquisition hardware and software analysis allows the data to be streamed over Ethernet to a connected PC. Users have the capability to scale their analysis platforms to their needs and can connect from anywhere on the network Configurable: The TSO820 mainframe supports userswappable current and future optical modules TSOVu: A new software platform that runs independent of the oscilloscope mainframe on users’ computers or server for both live and post-processing of acquired data TSOVu offers comprehensive analysis of PAM4 optical signals. Includes support for eye diagrams, optical measurements such as TDECQ, and other standard measurements Measurement plug-in designed to interface with the base software that can be dynamically installed as needed High test throughput • • Simultaneous capture of all channels at a high sample acquisition rate of 300 kS/s Sophisticated Programmatic Interface (PI) for automation environments enables the highest test throughput. Each command supports full data synchronization, eliminating the need for wait / sleep statements Optical Reference Receiver (ORR) is a 4th order Bessel-Thomson filter with a frequency response and tolerances as defined by the standards. Tektronix optimizes the response for best nominal fit and highest

Open the catalog to page 1
8 Series Sampling Oscilloscope - 2

TSO Series Sampling Oscilloscope Compliance testing Compliance test for 50G, 100G, 200G, and 400G IEEE 802.3TM standards such as 400GBASE-FR8, 400GBASE-DR4, and similar optical direct detect standards. The TSO820 Sampling Oscilloscope is also designed to support high speed NRZ standards such as 100 Gb Ethernet (100GBASE-LR4 or similar) that operates at a rate of 25.78125 Gb/s. Stay up-to-date on the latest standards changes by updating or adding new measurement plug-ins as they are released. The 8 Series Sampling Oscilloscope has been designed with modularity in mind. The TSO820 mainframe...

Open the catalog to page 2
8 Series Sampling Oscilloscope - 3

TSO Series Sampling Oscilloscope For more information regarding the TCR801, refer TCR801 Optical Clock Recovery Datasheet on www.tek.com/ Connection diagrams Figure 5: Connection diagram demonstrating the TSO820 mainframe with one TSO8C18 optical module, triggered by the TCR801 Optical Clock Recovery. User interface Figure 4: Connection diagram demonstrating the TSO820 mainframe with one TSO8C18 optical module, triggered directly from DUT or pattern generator The 8 Series Sampling Oscilloscope features a brand-new sampling oscilloscope software architecture called TSOVu. This new software...

Open the catalog to page 3
8 Series Sampling Oscilloscope - 4

TSO Series Sampling Oscilloscope Note: Wi-fi requires consistent and stable network connection for proper use. Measurement plug-in interface Using the flexible plug-in architecture in TSOVu, measurement plug-ins that can interface with TSOVu will display directly in the Add Measurement window. This includes Tektronix’ standard Pulse Measurement and PAM4 Optical Measurement plug-ins, and enables quick development of custom measurement libraries that are fully integrated in TSOVu to be called from the user interface or through PI commands. Figure 6: Example of the TSOVu session connected to a...

Open the catalog to page 4
8 Series Sampling Oscilloscope - 5

TSO Series Sampling Oscilloscope Supported measurements Pulse measurements: Amplitude High Low Amplitude Max Min Mid Mean Pk-Pk AC RMS Pulse measurements: Timing Period Frequency Rise Fall Positive Cross Negative Cross Positive Width Neagtive Width RMS Jitter Pk-Pk Jitter Delay RLM Level Level Deviation Level THickness OMAouter Extinction Ratio Effective Symbol Levels Eye Width Eye Height Transition Time Overshoot Undershoot TDECQ High-sensitivity operation accommodates lowamplitude signals The TSO8C17 and TSO8C18 optical modules feature high input sensitivity for measurement of low power...

Open the catalog to page 5
8 Series Sampling Oscilloscope - 6

TSO Series Sampling Oscilloscope Specifications TSO820 Mainframe Specifications All specifications are guaranteed unless noted otherwise. All specifications apply to all models unless noted otherwise. Vertical system Rise time / bandwidth Determined by the sampling modules used Vertical resolution (nominal) 15.6 bits over the sampling modules’ dynamic range Horizontal system Main time base / horizontal scale Record length Trigger system Trigger source Clock Prescale Input (front panel) Clock Prescale Input Clock input sensitivity Clock input range Pattern lengths supported (Pattern Sync)...

Open the catalog to page 6
8 Series Sampling Oscilloscope - 7

TSO Series Sampling Oscilloscope Input / output ports Front Panel Anti-static protection connector Clock Prescale Input 200 mVp-p to 1 Vp-p operational, AC coupled with maximum DC offset (-2.2 V to +2.2 V); 2 Vp-p absolute maximum Rear Panel Ethernet port RJ45 connector; supports IEEE 802.3TM Ethernet 100/1000BASE-T Control Control interface Ethernet port on page 7 Device information Instrument serial number, software version, other available using TSOVu Physical Height Weight (with blank module) Environmental Temperature Operating 5 to 45 °C, above 1500 m de-rate 1 °C per 300 m; automatic...

Open the catalog to page 7

All Keithley Instruments catalogs and technical brochures

  1. 6220-6221

    5 Pages

  2. 6482

    3 Pages

  3. 2520

    8 Pages

  4. 2606B

    13 Pages

  5. 2601B

    20 Pages

  6. AFG1000 Series

    13 Pages

  7. 6 Series B MSO

    69 Pages

  8. 3 Series MDO

    36 Pages

  9. 4 Series MSO

    40 Pages

  10. RTPA2A

    6 Pages

  11. 4200-SCS

    16 Pages

Archived catalogs