LEO 2

LEO 2

Product catalog summary
Overview: The document provides technical specifications and operational details for the LEO 2 digital manometer, a pressure measuring instrument by KELLER AG für Druckmesstechnik. It highlights the device's accuracy, functionality, and safety features.
Specifications:
  • Pressure Ranges: -1 to 3 bar, -1 to 30 bar, 0 to 300 bar, 0 to 700 bar with respective resolutions of 1 mbar, 10 mbar, 100 mbar, and 200 mbar.
  • Accuracy: Less than 0.1% FS at room temperature.
  • Total Error Band: Less than 0.2% FS within 0 to 50 °C.
  • Temperature Range: Operating: 0 to 50 °C, Storage: -20 to 70 °C.
  • Power Supply: 3V battery, type CR 2430, with a life of 1,000 hours of continuous operation.
  • Pressure Connection: 7/16”-20 UNF with an adapter for G 1/4”.
  • Protection: Rated IP 65.
  • Dimensions and Weight: 59 x 95 x 32 mm, approximately 125 g.
Functions:
  • RESET: Resets Max and Min pressure values to the current pressure.
  • ZERO: Allows setting a new zero reference to compensate for barometric pressure variations.
  • CONT: Disables automatic turn-off after 15 minutes of inactivity.
  • UNITS: Pressure can be displayed in various units including bar, mbar/hPa, kPa, MPa, PSI, and kp/cm².
Design Features:
  • Equipped with a piezoresistive pressure transducer and microprocessor for high accuracy.
  • Displays current pressure and records Max/Min pressure since last reset.
  • Includes a swivel fitting for 360° rotation and minimal tightness sealing up to 300 bar.
Optional Accessories: Protective rubber covering and carrying bag.
LEO 2 EI Version:
  • Intrinsically safe version compliant with ATEX 100a standards.
  • Classification: Ex II 1 G EEx ia IIC T6 or T5.
  • Certification File: LCIE 01 ATEX 6001 X.
  • Identical functions, ranges, and accuracy as the standard version with an additional protection switch.
See more

Catalog excerpts

LEO 2-1

Digital Manometer LEO 2 is a compact, micro-processor (µP) controlled, highly accurate and versatile pressure measuring instrument with digital indication. The piezoresistive pressure transducer as the heart of the instrument has gone through extensive pressure- and temperature tests. Its characteristics are stored in the instruments internal EEPROM. The µP of the LEO 2 reads the characteristic values and calculates therefrom the pressure to an accuracy of < 0,1 %FS at room temperature. The pressure is measured twice per second and displayed. The top display indicates the actual pressure, the bottom display shows the Max.- or Min.- pressure since the last RESET. LEO 2 has two operating keys. The left key is to turn the instrument on, to select the functions and the pressure units. The right key executes the selected function or unit. The right key is also used to switch between the Max.- and Min.-value. The instrument has the following functions: RESET With the RESET-function, the Max.- and Min.-value is set to the actual pressure value. The ZERO-function allows to set any value as a new Zero reference. Barometric pressure variations can thus be compensated. The factory setting of the Zero for the ranges -1…3 bar or -1…30 bar is at 0 bar absolute. For sealed gauge pressure measurements, activate “ZERO SEt” at ambient pressure. Instruments with ranges > 30 bar are calibrated in a sealed gauge mode at ambient pressure. The instrument turns off 15 Min. after the last key function. Activating cont (Continuous) deactivates this automatic turn-off. All standard instruments are calibrated in bar. The pressure can be indicated in the the following units: bar, mbar/hPa, kPa, MPa, PSI, kp/cm2. LEO 2 has 7/16” UNF male thread and is delivered complete with an O-ring seal swivel fitting which allows LEO 2 to be rotated through 360 °. Only minimal tightness is needed to seal up to 300 bar. Process pressure connection is G 1/4” male. Adapter G 1/4” Optional accessories: - Protective rubber covering - Carrying bag SpeCifiCationS Pressure Ranges, Resolution, Overpressure Accuracy RT (room temperature) * Total Error Band (0…50 °C) * Storage- / Operating Temperature Compensated Temperature Range Supply Battery Life Pressure Connection Protection, CEI 529 Diameter x Height x Depth Weight Resolution 1 mbar 10 mbar 100 mbar 200 mbar < 0,1 %FS < 0,2 %FS -20…70 °C / 0…50 °C 0…50 °C 3 V battery, type CR 2430 1’000 hours continuous operation 7/16”-20 UNF (adapter G 1/4” included) IP 65 59 x 95 x 32 mm ≈ 125 g Intrinsically safe version, 94/9/CE (ATEX 100a) Classification: Ex II 1 G EEx ia IIC T6 or T5 Certification File: LCIE 01 ATEX 6001 X The EEx ia version of LEO 2 incorporates an additional protection switch. Functions, ranges and accuracy are identical to the standard version. * Includes linearity, repeatability, hysteresis, temperature error and resolution of the display. Does not include Zero stability.

 Open the catalog to page 1
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.