ECO 2

ECO 2

Product catalog summary
Overview
ECO 2 is a compact and economical digital pressure measuring instrument designed for medium accuracy with high resolution and repeatability. It features a piezoresistive pressure transducer that has undergone extensive pressure and temperature testing. The device measures pressure twice per second, displaying the current pressure and the maximum or minimum pressure since the last reset.

Functions
  • RESET: Sets the Max. and Min. values to the current pressure.
  • ZERO: Allows setting any pressure value as a zero reference to compensate for barometric pressure variations.
  • CONT: Deactivates the automatic turn-off feature, which is set to 15 minutes after the last key function.
  • UNITS: Pressure can be displayed in various units including bar, mbar/hPa, kPa, MPa, PSI, and kp/cm².

Specifications
  • Pressure Ranges: -1 to 30 bar and 0 to 300 bar.
  • Resolution: 10 mbar for -1 to 30 bar range, 100 mbar for 0 to 300 bar range.
  • Overpressure: 60 bar for -1 to 30 bar range, 400 bar for 0 to 300 bar range.
  • Accuracy: Less than 0.5% FS at room temperature.
  • Total Error Band: Less than 1% FS including temperature error (0 to 50 °C).
  • Temperature Range: Storage: -20 to 70 °C, Operating: 0 to 50 °C.
  • Power Supply: 3V battery, type CR 24300, with a battery life of 1,000 hours of continuous operation.
  • Pressure Connection: 7/16”-20 UNF with an adapter G 1/4” included.
  • Materials: ECO 2 is made of AISI 316L stainless steel, with a brass nickel-plated swivel and galvanised steel adapter.
  • Protection: IP 65 rated.
  • Dimensions: 59 x 95 x 32 mm.
  • Weight: Instrument: approximately 100 g, Adapter: 40 g.

Optional Accessories
  • Protective rubber covering
  • Carrying bag

Intrinsically Safe Version (ECO 2 Ei)
The ECO 2 Ei is an intrinsically safe version compliant with 94/9/CE (ATEX 100a) standards, classified as Ex II 1 G EEx ia IIC T6 or T5. It includes an additional protection switch, with functions, ranges, and accuracy identical to the standard version. Certification File: LCIE 01 ATEX 6001 X.
See more

Catalog excerpts

ECO 2-1

Digital Manometer ECO 2 is a compact and economical digital pressure measuring instrument providing medium accuracy at high resolution and repeatability. The piezoresistive pressure transducer as the heart of the instrument has gone through extensive pressure- and temperature tests. The pressure is measured twice per second and displayed. The top display indicates the actual pressure, the bottom display shows the Max.- or Min.- pressure since the last RESET. ECO 2 has two operating keys. The left key turns on the instrument and is used to select the functions and the pressure units. The right key executes the selected function or unit. The right key is also used to switch between the Max.- and Min.-value. The instrument has the following functions: RESET With the RESET-function, the Max.- and Min.-value is set to the actual pressure value. The ZERO-function allows to set any pressure value as a zero reference. Barometric pressure variations can thus be compensated. The factory setting of the Zero for the range -1...30 bar is at vacuum (0 bar absolute). For relative pressure measurements, activate “ZERO SEt” at ambient pressure. Instruments with the range 0...300 bar are calibrated with the zero at atmospheric pressure. The instrument turns off 15 Min. after the last key function. Activating CONT (Continuous) deactivates this automatic turn-off. All standard instruments are calibrated in bar. The pressure can be indicated in the the following units: bar, mbar/hPa, kPa, MPa, PSI, kp/cm2. ECO 2 has 7/16” UNF male thread and is delivered complete with an O-ring seal swivel fitting which allows ECO 2 to be rotated 360°. Only minimal torque is needed to seal up to 300 bar. Process pressure connection is G 1/4” male. Adapter G 1/4” Optional accessories: - Protective rubber covering - Carrying bag Pressure Ranges Resolution Overpressure Accuracy @ Room Temperature * Total Error Band (0...50 °C) Storage-/Operating Temperature Supply Battery Life Pressure Connection Materials in Media Contact < 0,5 %FS < 1 %FS (accuracy incl. temperature error)) -20…70 °C / 0…50 °C 3 V battery, type CR 24300 1’000 hours continuous operation 7/16”-20 UNF (adapter G 1/4” included) ECO 2: AISI 316L (DIN 1.4404/1.4435) Swivel: Brass nickel-plated, Nitril O-Rings Adapter: Galvanised steel (9 SMnPb 28K), NBR profile seal IP 65 59 x 95 x 32 mm ≈ 100 g / 40 g Protection Diameter x Height x Depth Weight Instrument / Adapter Intrinsically safe version, 94/9/CE (ATEX 100a) Classification: Ex II 1 G EEx ia IICcT6 essor T5 ro X P Certification File: LCIE 01 ATEX 6001 in * Includes linearity, repeatability, hysteresis and resolution of the display. Does not include Zero stability. Subject to alterations rtof ECO 2 incorporates an The EEx ia version -Ce Ex additionalEprotection switch. Functions, ranges and accuracy are identical to

 Open the catalog to page 1
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.