Overview: The CAS 140CT-HR is a high-resolution array spectroradiometer designed for measuring narrow band emitters such as laser diodes. It offers high spectral resolution and short testing times, making it suitable for both production and laboratory applications.
Key Features:
  • High resolution down to 0.2 nm FWHM with a 0.08 nm data point interval.
  • High-end CCD detector actively cooled to -10°C.
  • Integration time as short as 9 ms.
  • Integrated density filter wheel.
Spectral Resolution: The device achieves spectral resolutions down to 0.2 nm for an 80 nm measuring range. Other ranges include 120 nm and 160 nm with resolutions of 0.3 nm and 0.4 nm, respectively.
Wavelength Ranges: Offers gratings with 1200, 1500, and 1800 lines/mm, covering typical measuring ranges of 80, 120, and 160 nm within the 800 to 1000 nm spectral range. Additional ranges in the visible spectrum are available upon request.
CCD Sensor: Utilizes a back-thinned and back-illuminated CCD array sensor with 1024 x 128 pixels, providing high sensitivity and a dynamic range of 30000:1. Active cooling ensures stable operation.
Technical Specifications:
  • Spectral range: 800 – 1000 nm.
  • Gratings: 1200, 1500, 1800 lines/mm.
  • Wavelength accuracy: ±0.05 nm.
  • Integration time: 9 ms – 65 s (PCIe); 10 ms – 65 s (USB).
  • Sensor dynamic range: 30000:1.
  • Non-linearity: ±0.5%.
Additional Features:
  • Focal length: Approx. 120 mm, f/3.5.
  • Filter wheel with density filters: OD 0.5, 1, 1.5, 2, 2.5.
  • PC interface: USB 2.0 standard, optional PCIe.
  • Triggering: 1 TTL input, 2 software-controlled TTL outputs, 1 TTL output with flash pulse.
Physical and Environmental Specifications:
  • Dimensions: 192 mm x 330 mm x 348 mm.
  • Weight: Approx. 10 kg.
  • Power supply: 100 VAC to 240 VAC, 50/60 Hz.
  • Operating temperature: 15 – 35 °C, max. 70% relative humidity, non-condensing.
Compliance: Conforms to EN 61010-1:2002-08 safety standards for electrical equipment used in measurement, control, and laboratory settings.
Ordering Information: Various models are available with different spectral ranges, resolutions, and configurations. Instrument Systems continues to develop and improve its products, with technical changes possible.
See more

Catalog excerpts

CAS 140CT-HR - High resolution array spectroradiometer-1

Instrument Systems KONICA MINOLTA Group CAS 140CT-HR High Resolution Array Spectroradiometer Key features at a glance A High resolution down to 0.2 nm FWHM, 0.08 nm data point interval A High-end CCD detector actively cooled to -10 °C A Down to 9 ms integration time A Integrated density filter wheel The high resolution CAS 140CT-HR is based on Instrument Systems’ well-proven high-end array spectroradiometer CAS 140CT. Particularly designed for the measurement of narrow band emitters, e.g. laser diodes, the CAS 140CT-HR combines high spectral resolution and short testing times for sophisticated production and laboratory applications. \\ VERY HIGH SPECTRAL RESOLUTION \\ ACTIVELY COOLED CCD SENSOR The CAS 140CT-HR models achieve very high spectral resolutions down to 0.2 nm (0.08 nm data point interval) for a spectral measuring range of 80 nm (see figure below). Measuring ranges of 120 and 160 nm result in spectral resolutions of 0.3 and 0.4 nm, respectively. \\ CUSTOMIZED WAVELENGTH RANGES The CAS 140CT-HR offers a selection of different gratings with 1200, 1500 and 1800 lines/mm. Typical measuring ranges of 80, 120 and 160 nm are available in the spectral range from 800 to 1000 nm. Further spectral ranges in the VIS are available upon request. A back-thinned and back-illuminated CCD array sensor with 1024 x 128 pixels is used for detection in the CAS 140CT-HR. This sensor design in combination with hardware binning of the vertical pixels offers a high level of sensitivity and large dynamic range (30000 : 1). The active cooling to -10°C ensures highly stable operation and reproducibility. The CCD allows capturing the spectrum of an optical emitter in a single exposure. Additionally, short integration times from 9 ms make the CAS 140 CT-HR particularly suitable for the measurement of emitters with pulsed and continuous operation modes. \\ TECHNICAL SPECIFICATIONS

 Open the catalog to page 1
CAS 140CT-HR - High resolution array spectroradiometer-2

Instrument Systems KONICA MINOLTA Group \\ TECHNICAL SPECIFICATIONS CAS 140CT-HR High Resolution Array Spectroradiometer 1 Further spectral ranges upon request. 2) Applies to Penray lamp or laser. 3) For a single acquisition with 10 ms integration time. 4) For option [P] an additional PCIe interface card and a 3 m cable with order number CAS140CT-403 is required. 5) At shortest integration time, without averaging and at 30,000 counts signal level. When averaged, this value improves (e.g. averaged over 9 times equals a threefold noise reduction). Instrument Systems is continually working to develop...

 Open the catalog to page 2

All Instrument Systems catalogs and technical brochures

  1. VTC 2400

    2  Pages

  2. CAS 125-HR

    2  Pages

  3. Photometers

    24  Pages

  4. LEDGON 100

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.