AMS 3000/5000 - Goniophotometers for large lamps and modules
16Pages

{{requestButtons}}

Catalog excerpts

AMS 3000/5000 - Goniophotometers for large lamps and modules - 1

Goniophotometers for large lamps and modules Ill Instrument /Systems KONICA MINOLTA Group Optronik Line

Open the catalog to page 1
AMS 3000/5000 - Goniophotometers for large lamps and modules - 2

CIE Type A goniophotometer for large test samples Fast, robust and precise Automated set-up mode and sequential measuring procedures Data export in IES, KRS, CSV formats Conformity with GTB and SAE J1330 photometry precision requirements for type testing of vehicle lights 01 \\ Fast – robust – precise The AMS 3000 and the AMS 5000 of the Optronik line are Type A goniophotometers in accordance with CIE 121-1996 and IESLM-75-01 for the measurement of photometric and colorimetric data in the far field. Due to their highly dynamic servo drives and robust design, they are suitable for measuring...

Open the catalog to page 2
AMS 3000/5000 - Goniophotometers for large lamps and modules - 3

Optronik Line 02 \\ Equipment and function The AMS 3000 and AMS 5000 goniometers consist of a sample stage that can be moved in three linear axes in addition to the two rotational axes for the attachment of the test sample, as well as the AMS controller, the central control system of the goniophotometer. The AMS 5000 goniometer has a control unit with touchscreen on the sample stage. In addition, the RecoCAN hand-held remote control is optionally available for the two systems. The photometer head belonging to the system is beyond the photometric limit distance in a stray light tube adjusted...

Open the catalog to page 3
AMS 3000/5000 - Goniophotometers for large lamps and modules - 4

The goniometer systems have a sample stage with mounting grooves and locating pin sockets for holding the test samples. The electrical connection of the samples is realized via a lamp multiplexer mounted directly on the sample stage. This ensures that the cables move synchronously with the sample during operation. Sensor lines are used to measure the electrical operating data of the sample and stabilize it directly on the sample. The operator switches manually between the individual channels via the SNT 10 power supply units or remote controlled via the LightCon software program. In a...

Open the catalog to page 4
AMS 3000/5000 - Goniophotometers for large lamps and modules - 5

The sample connector of the AMS 5000 also offers the option of selecting a power supply and power supply channel directly by touchscreen and channel buttons on the lamp multiplexer and to set voltage and current. This is enabled by the CAN bus integration of all system components. On the lamp multiplexer of both goniometer systems a CAN bus can be used for integrating further functions, e.g. for the temperature measurement of an LED module. During the lamp burn-in, the temperature and light intensity profiles can be recorded simultaneously. An optional PWM generator can also be connected,...

Open the catalog to page 5
AMS 3000/5000 - Goniophotometers for large lamps and modules - 6

The AMS controller is the central control unit of the goniophotometer and provides detailed display options on the modern LCD panel. The controller comprises the following functions: Y Display of the goniometer angular positions This way, a remote display for retroreflectometer, tristimulus colorimetric device and SNT 10 power supply units, for example, can be stored in a rack. Y Display of the X-Y-Z linear axes position Y Manual input of an angular or linear axis position via the keypad Y Display of the photometric value of the connected DSP 200 photometer Y Switching between different...

Open the catalog to page 6
AMS 3000/5000 - Goniophotometers for large lamps and modules - 7

Optronik Line The AMS 3000 and AMS 5000 goniophotometers can be combined with different devices in order to cover all the necessary measurement tasks in a light laboratory. The accessory range of the Optronik product line is ideally matched to the requirements of the automotive industry in the light channel and fulfills all the relevant standards. Conventional configuration with the DSP 200 photometer Normally, the systems are operated in combination with the fast Optronik line of DSP 200 photometers which are particularly suitable for scanning “on-the-fly” measurements. Like other...

Open the catalog to page 7
AMS 3000/5000 - Goniophotometers for large lamps and modules - 8

Layout of a light laboratory 30.48 m Instrument Systems creates for customers an individual lab layout tailored to the structural requirements and desired application spectrum. The positions/ distances of goniometer, photometer, control cabinet, sensors and accessories (e.g. tube positioning unit, imaging colorimeter and retroreflectometer) are 12 m specified therein. Side view: Stand with DSP photometer & CM colorimeter for signal lamps / license plates for rapid scans of light distributions TPU positioning unit, DSP photometer & CM colorimeter Colorimeter with projector for color of...

Open the catalog to page 8
AMS 3000/5000 - Goniophotometers for large lamps and modules - 9

Stand with DSP photometer RMS retroreflectometer

Open the catalog to page 9
AMS 3000/5000 - Goniophotometers for large lamps and modules - 10

Photometric limiting distance According to the definition, the light intensity distribution in the far field must be measured at a distance at which the test sample can be seen as a point light source. The distance of the detector from the test sample from which this criterion is fulfilled is called the photometric limiting distance. It depends on the size of the light source to be measured and the light-sensitive surface of the detector, as well as the required measurement deviation. The minimum distance to be observed between test sample and detector is often stated as being a multiple of...

Open the catalog to page 10
AMS 3000/5000 - Goniophotometers for large lamps and modules - 11

Dimensions of AMS 3000 – front view. ~ Dimensions of AMS 3000 – side view.

Open the catalog to page 11
AMS 3000/5000 - Goniophotometers for large lamps and modules - 12

Dimensions of AMS 5000 – front view. ~ Dimensions of AMS 5000 – side view.

Open the catalog to page 12
AMS 3000/5000 - Goniophotometers for large lamps and modules - 14

Instrument Systems is constantly striving to develop and improve its products. Technical changes, errors or printing mistakes do not justify claims for damages. In all other cases, our terms and conditions shall apply.

Open the catalog to page 14
AMS 3000/5000 - Goniophotometers for large lamps and modules - 15

Optronik Line 11 \\ Ordering information Order number An extensive additional accessory range is available for the AMS goniophotometer.

Open the catalog to page 15
AMS 3000/5000 - Goniophotometers for large lamps and modules - 16