Instron Autoinjector Testing System - 1

AUTOINJECTOR TESTING SYSTEM Semi-Automated System for Testing Pen and Autoinjectors to ISO 11608

Open the catalog to page 1
Instron Autoinjector Testing System - 3

AUTOINJECTOR TESTING SYSTEM Semi-Automated System for Testing Pen and Autoinjectors Developed in close partnership with pharmaceutical device manufacturers, Instron’s latest generation Autoinjector Testing System can perform full functionality testing on a wide range of drug delivery devices – including needle shield and button-activated devices, as well as safety syringes. Instron’s Autoinjector Testing System measures a variety of essential performance requirements, including cap removal, dose accuracy, activation force, injection time, needle depth, and needle guard lockout – allowing...

Open the catalog to page 3
Instron Autoinjector Testing System - 4

AUTOINJECTOR TESTING SYSTEM Semi-Automated System for Testing Pen and Autoinjectors Due to the complexity of autoinjector devices, labs often require multiple pieces of equipment to perform the wide range of tests required. This approach is more time consuming, requires more specimens, and creates a need for data consolidation. Simplify the Testing Process Instron’s Autoinjector Testing System replaces traditional test procedures that require separate pieces of equipment, enabling users to run a complete sequence of tests on a SINGLE SYSTEM. This allows labs to accelerate time to market by:...

Open the catalog to page 4
Instron Autoinjector Testing System - 5

FULL FUNCTIONALITY TESTING WITH A SINGLE SYSTEM ACTIVATION FORCE CLICK DETECTION INJECTION TIME NEEDLE DEPTH DELIVERED VOLUME NEEDLE GUARD LOCKOUT Simplify the Testing Process | 05

Open the catalog to page 5
Instron Autoinjector Testing System - 6

BLUEHILL® UNIVERSAL Dedicated Test Methods Bluehill Universal uses simplified test types that empower users to develop and change method parameters with ease, while offering the flexibility to readily accommodate future devices without requiring Instron’s support. Device Properties Button and needle shield devices are options in the specimen properties tab. The device type is selected and dimensions are entered. These dimensions are used to properly position the crosshead throughout the test sequence, optimizing test time and ensuring tests are run consistently, even on different systems.

Open the catalog to page 6
Instron Autoinjector Testing System - 7

Simple Test Method Type Users simply select the required functional tests – cap removal, injection, and safety check of the needle shield – and enter the parameters. The system automatically runs the tests in the appropriate sequence. TestProfiler Method Type TestProfiler allows users to customize the standard test method parameters, providing greater flexibility when designing new devices. Pre-Loaded Templates Pre-loaded templates for needle shield, button activated, and safety syringe devices are included with the software, making it fast and easy for a user to create a new device method...

Open the catalog to page 7
Instron Autoinjector Testing System - 8

SYSTEM SUITABILITY TESTING Improved Compliance in Bluehill® Universal Bluehill Universal’s autoinjector testing software integrates system suitability testing into its workflow, prompting users to perform tests and automatically tracking results in the audit trail. Good Manufacturing Practice (GMP) labs require daily checks of transducers prior to testing to ensure they are functioning accurately. In-house development of the hardware and test methods for performing these checks are time consuming and difficult to validate. Often labs use paper records to track the completion and performance...

Open the catalog to page 8
Instron Autoinjector Testing System - 9

Enforced System Suitability Testing A system administrator can turn on the SST requirement and specify which transducers must be checked as part of the process. Frequency of SSTs are also set by the administrator based on a time period and/or upon the creation of a new sample. Bluehill Universal will prevent additional device testing until the SSTs have been successfully completed. Traceability A report of the system suitability results is stored as a PDF and the successful completion of these checks is recorded in the audit log of the Traceability module, offering an efficient way to...

Open the catalog to page 9
Instron Autoinjector Testing System - 10

MACHINE VISION CAMERA Needle Depth and Injection Time Instron’s system incorporates a machine vision camera to provide a measured injection time as well as indications of the exposed needle depth at both the start and end of the injection. Measurement of the injection time and needle depth during the injection of the fluid is a critical piece of data to ensure the device is working as therapeutically intended. The use of an optical measurement method to capture this data is more effective than traditional methods, such as gravimetric or laser-based systems, especially as drug formulations...

Open the catalog to page 10
Instron Autoinjector Testing System - 11

Needle Depth at Start of Injection A measurement of the needle depth when the initial fluid expulsion occurs provides the ability to identify whether the device is delivering its dose into the therapeutic range. This measurement can be brought into Bluehill Universal. A photo of this event is stored, making it available for analysis any time after the test has been completed. Measurement of Injection Time The machine vision camera provides high accuracy measurements of injection time, which supports a wide range of delivery profiles, including high viscosity formulations with non-continuous...

Open the catalog to page 11
Instron Autoinjector Testing System - 12

ROOT CAUSE ANALYSIS High-Resolution Video Camera Instron’s Autoinjector Testing System uses a high-resolution video camera, along with the machine vision camera, to provide critical visuals to aid in root cause analysis when a device failure occurs. Some device failures are clearly detectable based on the calculated results, data curves, or the appearance of the device itself. Other failures are not obvious until the entire sample batch has been tested and statistically analyzed (K-value). In either scenario it is important to determine whether a bad result was truly a device failure or...

Open the catalog to page 12
Instron Autoinjector Testing System - 13

Video Recording of Injection Needle Depth Image Storage A high-resolution camera monitors the injection and records a video. This video can be viewed and replayed through Bluehill’s TestCam module, allowing further analysis to be performed on a previously run test. When the video is replayed the closest corresponding data point on the test curve is highlighted. Conversely, a scanning cursor can be used to select a particular point of interest on a data curve which pulls up the associated video frame. Images of the needle at both the start and end of the injection are stored for each device....

Open the catalog to page 13