1. Catalogs
  2. inelta Sensorsysteme
  3. Standard Force sensors Type: KMM65

Standard Force sensors Type: KMM65

Standard Force sensors Type: KMM65

Standard Force sensors Type: KMM65

Product catalog summary
Technical Specifications:
The force sensors from the KMM65 series have a measuring range from 1kN to 500kN, suitable for both tension and compression measurements. They are robustly constructed with small dimensions and have a protection degree of IP54.

Options:
Custom cable lengths, extended temperature range (-40°C to +120°C), individual calibration services, and internal or external signal conditioners are available upon request.

Applications:
These sensors are applicable in various fields including industry and laboratory, medical technology, weighing systems, testing facilities, engine and installation construction, sports equipment, winding machines, and industrial presses.

Technical Specifications:
- Measuring range: 1, 2, 5, 10, 20, 50, 100, 200, 500 kN
- Service load: 130% F.S.
- Limit load: 150% F.S.
- Breaking load: 300% F.S.
- Sensitivity: 1.5 ± 2% mV/V
- Zero signal tolerance: < 2% F.S.
- Linearity: < 0.2% F.S.
- Repeatability: < 0.1% F.S.
- Creep error (30 min): < 0.1% F.S.
- Hysteresis: < 0.2% F.S.
- Cut off frequency: 300 Hz
- Temperature coefficient sensitivity: < 0.01% F.S./°C
- Temperature coefficient zero point: < 0.01% F.S./°C

Electrical Specifications:
- Excitation voltage: 10 (max. 15) V
- Input impedance: 1-50 kN = 725 ± 20%, 100-500 kN = 1075 ± 20% Ω
- Output impedance: 1-50 kN = 700 ± 10%, 100-500 kN = 1050 ± 10% Ω
- Insulation resistance: > 5000 MΩ at 500 VDC

Mechanical Specifications:
- Housing material: Steel, nickel-plated
- Measuring method: DMS Wheatstone bridge
- Operating temperature: -10 to +70°C (compensated: 0 to +50°C)
- Storage temperature: -25 to +80°C
- Relative humidity: 95% r.H. (non-condensing)

Connection Details:
- Cable length: 2m
- Cable configuration: Brown (+ Excitation), White (- Excitation), Green (+ Signal), Yellow (- Signal), Shield (Housing)

Additional Information:
For application directions and user guides, visit www.inelta.de/service. Signal conditioners are available in various output signals including 0..5 V, 0..10 V, ±5 V, ±10 V, 0..20mA, and 4..20 mA.
See more

Catalog excerpts

Standard Force sensors Type: KMM65-1

DMS Messbrcke DMS Wheatstonebridge Gem줟 / According: VDI/VDE/DKD 2638 > Andere Kabellnge auf Anfrage (Standard = 2m) / Other cable lenght on request ( standard = 2m) Erweiterter Temperaturbereich (-40C .. +120䰰C) / Extended temperature range (-40C .. +120аC) Individuelle Kalibrierungen / Individual calibration service Sensorsignalverstrker (intern, extern) / Signal conditioner (internal, external) Weitere Optionen auf Anfrage / More options on request inelta Sensorsysteme GmbH & Co. Haidgraben 9a ䷷ D-85521 Ottobrunn/Mnchen, Germany Phone +49 (0)89/45 22 45 - 0 Fax +49 (0)89/45 22 45 췖 244 eMail: [email protected] ׷ www.inelta.de >

 Open the catalog to page 1
Standard Force sensors Type: KMM65-2

Kraftmessbereich Measuring range 125102050100200500[kN]Max.Gebrauchskraft / Service load 130[% F.S.]Grenzkraft / Limit load 150[% F.S.]Bruchlast / Breaking load 300[% F.S.]Empfindlichkeit / Sensitivity 1,5 2%[mV/V]Rel. Abweichung d. Nullsignals Zero signal tolerance < 2[% F.S.]Rel. LinearitѤtsabweichung / Linearity < 0,2[% F.S.]Reproduzierbarkeit / Repeatability < 0,1[% F.S.]Kriechen (30 min) / Creep error (30 min) < 0,1[% F.S.]Hysterese / Hysteresis < 0,2[% F.S.]Grenzfrequenz fg / Cut off frequency fg 300[Hz]Temp.einfluss auf Kennwert pro K Temp. coefficient sensitivitly < 0,01[% F.S./C]Temp.einfluss...

 Open the catalog to page 2
Standard Force sensors Type: KMM65-4

Serie Kraftmessbereich Series Measuring range KMM65-1kN Standard 1kN, 2kN, 5kN, 10kN, 20kN, 50kN, 100kN, 200kN, 500kN Standard Datenblatt www.inelta.de > Datasheet Zubehr auf Anfrage / Accessories on request www.inelta.de > Datasheet Serie / Series IMA2-DMSISM-DMSIVM2-DMS Ausgangssignale / Output signals 0..5 V, 0..10 V, 5 V, ѱ10 V, 0..20mA, 4..20 mA0..10 V, 4..20 mA0..10 V Datenblatt PDF Downloadwww.inelta.de/service > Die Broschre zum Produkt mit Applikationsbeispielen und Benutzerhinweisen finden Sie im Internet unter www.inelta.de/service Suitable for this product you can download the brochure...

 Open the catalog to page 4
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.