Product catalog summary
Document Overview
This document is a comprehensive guide for the installation, connection, and commissioning of the Ijinus overflow detector, including maintenance instructions. It provides essential safety information, product specifications, and operational guidelines.
Safety Instructions
Operations must be conducted by qualified personnel. Only specified batteries should be used to prevent risks, and the device should be handled carefully to avoid electrostatic discharge damage.
Product Compliance
The detector is CE marked, indicating compliance with European directives, and should not be disposed of with household waste. It complies with European Directive 2014/53/EU and FCC regulations for Class B digital devices.
Technical Specifications
  • LSCV4: Features a detection threshold of 90% for wastewater, operates between -20°C to 50°C, and has a memory capacity of 200 events.
  • CSCV4: Similar specifications to LSCV4, compatible with specific loggers, and supports various power supply configurations.
Installation and Connection
Connections should be made with the power off by authorized personnel. The detector connects to a PLC via Modbus and requires a secure voltage source between 9 and 24V DC. Installation should avoid areas with constant splashes or metal facing the electrode.
Power Supply
The internal battery lasts over 5 years under default settings. An external power supply can be used to extend battery life, with the internal battery as a backup.
Commissioning and Configuration
Before operation, check the detector's mechanical condition. Use the OVERFLOW mobile app for configuration via Bluetooth, allowing status viewing, measurement history management, and detection threshold settings.
Software and Equipment
Avelour software version 7.1 or later, a Wiji radio antenna, and specific logger types are required. The software can be downloaded from the Ijinus website and installed using command line parameters.
Logger Connection
Connect the Wiji radio access point to a computer's USB port. Use Avelour software to connect to the logger by selecting its serial number. Ensure the antenna is positioned at least one meter from the logger.
Modbus Configuration
Default communication parameters are 9600 baud, 8 data bits, 1 stop bit, and no parity. The Modbus ID is 1 by default, with configuration tables for speed, address, and capacitive saturation thresholds.
Maintenance
Avoid placing the product near magnets. For issues, contact customer service. If the internal battery is discharged, refer to the OverFlow app status display.
See more

Catalog excerpts

CSCV4 - LSCV4-1

Overflow detector User guide IJINUS - 25 ZA de Kervidanou 3, 29300 MELLAC - France – SAV / After-sales service : +33 2 98 09 03 32 / [email protected]

 Open the catalog to page 1
CSCV4 - LSCV4-2

User guide: Version 04 Publication date 29/09/2025

 Open the catalog to page 2
CSCV4 - LSCV4-3

User guide Bluetooth (CSC & LSC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Modbus (CSC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital output (CSC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....

 Open the catalog to page 3
CSCV4 - LSCV4-4

User guide 10.9.1. Change login password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Modify the detection threshold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Overflow detector

 Open the catalog to page 4
CSCV4 - LSCV4-5

User guide Chapitre 1. Document information 1.1. Background This user guide contains all the information required to install, connect and commission the unit, as well as important notes concerning maintenance. It is therefore essential to read it before commissioning any Ijinus equipment. 1.2. Symbols used This symbol indicates a situation or use that may result in damage, fault or equipment malfunction. This symbol indicates additional information useful for the understanding and correct use of the equipment. This symbol indicates a prerequisite for performing a task. Overflow detector

 Open the catalog to page 5
CSCV4 - LSCV4-6

User guide This document presents a number of operations and programming to be performed on a data logger, a sensor or an accessory supplied by Ijinus. These operations must only be performed by personnel qualified to use Ijinus products. The information provided in this user guide only ensures operational safety if the equipment is used correctly. Performing any work on the device requires the use of appropriate personal protective equipment. Below we have provided a non-exhaustive list of recommendations to apply to ensure the safety of Ijinus data logger users: • Only use batteries specified...

 Open the catalog to page 6
CSCV4 - LSCV4-7

User guide Chapter 3. Product marking information The CE mark indicates that the product complies with current European directives. Consult the user manual before using the product. Do not dispose of this product in household waste. The product must be sent to a specific collection point, or collected by an organization that will ensure its further processing. Direct current Meaning of symbols Overflow detector

 Open the catalog to page 7
CSCV4 - LSCV4-8

User guide The detector is equipped with a Bluetooth LE radio transmitter. IJINUS declares that CSC and LSC radio equipment are compliant with European Directive 2014/53/EU. The CE declaration of conformity is available for download from the website www.ijinus.com. The OVERFLOW detector, models CSCV4 and LSCV4, comply with 47 CFR Part 15.107 & Part 15.109 regulations. This equipment must be installed by a professional. This equipment has been tested and found to comply with the limits imposed on Class B digital devices under Part 15 of the FCC rules. These limits are designed to provide sufficient...

 Open the catalog to page 8
CSCV4 - LSCV4-9

User guide Chapter 5. Description 5.1. Principle of operation The overflow detector is a binary sensor that emits a signal when it comes into contact with the effluent. The purpose of the overflow detector is to indicate the presence of water above a certain threshold by means of a digital “On/Off” state. The overflow detector decides on the ON/OFF status based on a capacitive measurement of its environment and modifiable thresholds. The overflow detector takes into account elements in physical contact with the housing and up to a few centimeters away from it. The detector is highly insensitive...

 Open the catalog to page 9
CSCV4 - LSCV4-10

User guide 5.1.3. Communication Bluetooth (CSC & LSC) The overflow detector is configured in Bluetooth Low Energy (BLE) using the Overflow mobile App (see Chapter 10, Configuration on the OVERFLOW application). The use of BLE allows permanent “listening” to the overflow detector. It is therefore not necessary to activate the detector to communicate with the OVERFLOW application. Modbus (CSC) In its bare wire version and connected to a PLC, the overflow detector can transmit a change of state as well as a fouling indicator (if the Modbus connection is configured). Refer to the paragraph Chapter...

 Open the catalog to page 10
CSCV4 - LSCV4-11

User guide Overflow detector

 Open the catalog to page 11
CSCV4 - LSCV4-12

User guide Overflow detector

 Open the catalog to page 12
CSCV4 - LSCV4-13

User guide Overflow detector

 Open the catalog to page 13
CSCV4 - LSCV4-14

User guide Chapter 6. Connection (CSC only) 6.1. Overflow detector wiring 6.1.1. Connections Making electrical connections is strictly reserved for authorized personnel. Electrical connections must always be made with the power off. Armoire électrique / Electrical cabinet Digital output ● Connect the pink wire (Open-Drain) and the brown wire (V-) to the Open-drain digital input. Overflow detecto

 Open the catalog to page 14
CSCV4 - LSCV4-15

User guide External power supply The overflow detector must be electrically powered using a voltage source between 9 and 24V DC. Power is supplied via the green wire (V+) and the brown wire (V-) Connection is made to a secure voltage source equipped with a 100 mA limitation. The detector must be powered by an external power supply as indicated in the paragraph the section called "External power supply". • Connect the yellow wire (Modbus high) and the grey wire (Modbus low) to the PLC Modbus board. Overflow detector

 Open the catalog to page 15
CSCV4 - LSCV4-16

User guide Overflow detector

 Open the catalog to page 16
CSCV4 - LSCV4-17

User guide Chapter 7. Power supply 7.1. Internal power supply Battery life is in excess of 5 years if the detector is set to its default configuration, i.e. a frequency of 1 measurement every 4 seconds. Any modification to the measurement frequency will impact the autonomy of the product. The battery charge level can be viewed via the OverFlow application (see Status display on the OverFlow app). ● On the wireless version, the internal battery is essential for detector operation. ● On the wired version, the use of an external power supply preserves the life of the internal battery. 7.2. External...

 Open the catalog to page 17