Catalog excerpts
Elemental Analysis Series Proximate Analysis Series ► The first automatic proximate analyzer in China was invented by Sundy in the year of 2000; ► 41 patents (including 9 invention patents) for proximate analysis; ► Three generations proximate analyzer have been developed; ► Unique characteristics: balance no drift, no waiting for batches of test. In the field of proximate analysis, Sundy is capable of supplying not only automatic analyzer for replacing traditional mode "muffle furnace+drying oven", but also moisture tester, muffle furnace and drying oven. ► Determination of moisture, ash and volatile matter as well as calculation of fixed carbon and calorific value in coal, coke and other solid biomass fuel; ► Analysis of carbon content in fly ash, clinker; ► Analysis of moisture, ash, volatile matter in calcined petrol coke; ► Ignition loss of cement / minerals. ► ASTM D 5142-09 - Standard Test Method for Proximate Analysis of the Analysis Sample of Coal and Coke by Instrumental Procedures. ASTM D 3174-04 - Standard Test Method for Ash in the Analysis Sample of Coal and Coke. ► ASTM D 3175-07 Standard Test Method for Volatile Matter in the Analysis Sample of Coal and Coke. ► ISO 11722 - Solid Mineral Fuels - Hard Coal - Determination of Moisture in the General Analysis Test ► Sample by Drying in Nitrogen. ► ISO 1171 - Solid mineral fuels - Determination of Ash. ► ISO 562 - Hard coal and coke - Determination of Volatile Matter.
Open the catalog to page 1l. Dual Furnaces, dual internal balances, stable performance, accurate test result. 2. The only proximate analyzer which can test batches of sample without waiting for cooling. Sample mass (0.5-1.1) g (Seif-defined) Furnace temperature_ Room temperature~950'C_ Sample Loader 18 samples Temperature control precision_ ± 2.5°C (Moisture), ± 5*C (Ash, Volatile matter)_ Test efficiency Full indexes for 3 batches of samples (54 samples) within 8h Gas requirement_ Compressed air / Oxygen & Nitrogen_
Open the catalog to page 2Proximate Analysis Series ► High precision & accuracy. Two internal balances away from high temperature and heat radiation, no drift occurs. With automatic calibration function, each weighing is Simulate test environment and condition of drying oven and muffle furnace, exact in accordance with The unique automatic cleaning function of low tem- perature furnace completely eliminates the influence of ► Human based design. Environment friendly and energy Sample tray locates inside, real time data transmis- sion. Weighing sample at room temperature, away from heat radiation. Special crucible...
Open the catalog to page 3Proximate Analysis Series Main body size: 845mm"563mm'596mm Main body weight: 146KG Volatile Matter Self-defined Method Sample Mass_ Temperature Control Precision Furnace Temperature Range Analytical Balance Precision Power Requirement ±2.5'C (moisture), ±5"C(ash, volatile matter) Simulate test environment and condition of drying oven and muffle furnace. Applicable for all type of coal analysis. Patents ensure the reliability and stability of me- chanic movement system. No crucible fall off or turn Assembly line operation mode. Sample weighing, introducing, test, taking, discarding and data...
Open the catalog to page 4Proximate Analysis Series Test index Continuous test in batches Cooling time required for batches Furnace Cooling (forced air/N2) Balance Resolution Sample introduction and taking Quantity of sample/ Analysis time Sample weight Furnace Temp Furnace Heating Temperature Calibration Temp, control precision Power requirement Sample weighing Test Temperature Operational Control Advance Calculation Analysis Report Statistics Report System Monitor Moisture, ash and volatile matter Moisture, ash and volatile matter (Moisture); 16 samples/ 230 min (America 10 samples / 7 min (ASTM Volatile matter 1)...
Open the catalog to page 5Proximate Analysis Series 20 Samples/batch, high efficiency. Main body size: 636mm*651mm*451mm Main body weight: 65KG Sample mass Sample particle Furnace temperature Test duration Power requirement 20 samples (Inherent moisture and total 0.9 to 1.1g for air dried basis moisture 10 to 12g for total moisture sample ( User -defined sample mass is also available) 15-40 min for analysis moisture test 20~60min for total moisture test One batch 20 samples, short waiting time between two batches.high test efficiency. When the power is off, test data can be stored in the computer without lost. The...
Open the catalog to page 6Proximate Analysis Series SDTGA200 Ash & Volatile Matter Tester Sundy ash & volatile matter tester can be used to determine ash and volatile content in coal, coke and other solid biomass fuels in power plant, coal industry, metallurgy industry, petrochemical industry, environmental protection, cement industry, paper plant, geological prospecting, research institutes , etc. Temperature Control Range Temperature Control Precision ±2.5X1 Temperature Resolution Furnace Dimension Power Requirement ► Multi test mode, reliable operation. ► Humanity Design & High Precision. 5 test mode supported....
Open the catalog to page 7Proximate Analysis Series SDDH315 Blast Drying Oven SDDH315 Drying Oven can be used to determine moisture content in coal, coke and solid biomass fuel and is applicable to drying, baking, wax melting and heat treatment in laboratory. Time of temperature rise can be preset. With both blast and nitrogen feeding function, nitro- all types of coal, low gas consumption, high tem- perature control precision. Temperature control range is 50% wider than normal drying oven. Advanced heat circulation and insulation system, imported low noise motor realize uniform tempera- ture in working chamber....
Open the catalog to page 8Proximate Analysis Series I ISDDH323/306 Blast Drying Oven Fast temperature rise, accurate temperature control and stable operation. The Drying Oven can be used to determine moisture content in coal, coke and solid biomass fuel and is applicable to drying, baking, wax melting and heat treatment in laboratory. Temperature Conlrol Range Temperature Control Precision Power Requirement Main body size: SDDH323:730mm"670mm"1220mm Main body weight: SDDH323:105KG The shell of the oven is made of superior cold-roll steel sheets with beautiful appearance and long life. Use air blower with low noise...
Open the catalog to page 9All Hunan Sundy Science and Technology Co. catalogs and technical brochures
-
SDAC6000 Bomb Calorimeter
61 Pages
-
SDAC6000 bomb calorimeter
2 Pages
-
SDTGA8000 proximate analyzer
2 Pages
-
SDVD3mm quick drying
1 Pages
-
SDVD25 Quick drying device
1 Pages
-
SDJC Jaw crusher
1 Pages
-
SDAF105b ash fusion tester
2 Pages
-
SDPP pulverizer
1 Pages
-
SDRC double roller crusher
1 Pages
-
SDRD divider
1 Pages
-
SDTGA6000
2 Pages
-
SDMD16 automatic divider
1 Pages
-
SDHC hammer crusher
1 Pages
-
SDVD3mm quick dryer
1 Pages
-
SDTGA520 moisture analyzer
1 Pages
-
SDS350 sulfur analyzer
2 Pages
-
SDC5015 calorimeter
1 Pages
-
SDAC6000 calorimeter
2 Pages
-
Sundy ISO 9001 Certification
1 Pages