Product catalog summary
Overview: This document provides technical specifications and guidelines for Calibration Reference Cylinders (CRC) used with thermal needle probes TP02/TP08, TP07, and TP09. These cylinders are essential for calibrating Non-Steady-State Probes using traceable thermal conductivity reference materials.
Specifications:
  • Test Methods: ASTM D 5334-00, D 5930-97, IEEE Std 442-1981
  • Temperature Range: 0 to 60 degrees Celsius
  • Thermal Conductivity of Reference: 0.2 W/mK (nominal value)
  • Traceability: NPL National Physical Laboratory, United Kingdom
CRC Cylinder Types and Applications:
  • CRC01: Dimensions - H = 260 mm, D = 100 mm, IL = 185 - 30 mm, ND = 1.6 mm; Suitable for TP02, TP08
  • CRC04: Dimensions - H = 206 mm, D = 100 mm, IL = 137 - 6 mm, ND = 3.6 mm; Suitable for TP07
  • CRC05: Dimensions - H = 260 mm, D = 100 mm, IL = 190 - 12 mm, ND = 6.6 mm; Suitable for TP09
Design and Materials:
  • The core of the CRC cylinders is made from a specially manufactured amorphous plastic reference material that lacks directionally dependent thermal properties.
  • A hole is drilled in the plastic material to match the needle's outer dimensions (ND and IL).
  • The plastic material is encased in a metal protection casing filled with glycerol to ensure perfect contact between the needle and reference material.
  • An additional hole is provided for material stability checks with thin needle probes.
Maintenance and Recommendations:
  • Users are advised to send the CRC to Hukseflux for material stability checks every five years.
  • Hukseflux offers a low-cost alternative reference material using glycerol filled with plastic fibre material, included as standard with most systems.
Note: Hukseflux reserves the right to alter specifications without prior notice.
See more

Catalog excerpts

CRC series-1

CRC SERIES CALIBRATION REFERENCE CYLINDERS FOR THERMAL NEEDLE PROBES TP02/TP08, TP07 AND TP09 1 D 2 3 4 H IL ND 5 6 7 Figure 1: generalised picture of CRC Calibration Reference Cylinder. Explanation can be found in the text. Type CRC01 Dimensions Suitability for needle types TP02, TP08 H = 260 D = 100 IL = 185 - 30 ND = 1.6 TP07 CRC04 H = 206 D = 100 IL = 137 - 6 ND = 3.6 TP09 CRC05 H = 260 D = 100 IL = 190 - 12 ND = 6.6 Table 1: CRC cylinder types and their application CRC’s contain traceable thermal conductivity reference materials for calibration of Non-SteadyState Probes. There are several different CRC types. The geometry of the different types is adapted to fit needles with a certain diameter. The hart of the CRC cylinders is a specially manufactured amorphous plastic reference material (5). Contrary to other plastics, this material does not have any directionally dependent thermal properties. In the plastic material a hole is drilled (4), with the approximate outer dimensions (ND Needle Diameter and IL Insertion Length) of the needle that needs to be calibrated. The plastic material is incorporated in a metal protection casing (2). A plug (7) is provided for sealing purposes. In order to guarantee perfect contact between needle (1) and reference material (5), the metal casing is filled with a fluid, glycerol (3). An additional hole (6) is provided in order to be able to perform checks of material stability with thin needle probes. The user is recommended to send the CRC to Hukseflux for a check of material stability every 5 years. Hukseflux is a leading supplier of Non-SteadyState Probes, and is the inventor of the CRC calibration method. NOTE: As a low-cost (lower accuracy) alternative reference material Hukseflux promotes use of glycerol filled with plastic fibre material. This is delivered as a standard with most Hukseflux systems. CRC SPECIFICATIONS Test method: ASTM D 5334-00 and D 5930-97 IEEE Std 442-1981 Temperature range: 0 to 60 degrees C Thermal conductivity of reference: 0.2 W/mK (nominal value) Traceability: to NPL National Physical Laboratory, United Kingdom Version 0616 (page 1 of 1) Hukseflux reserve the right to alter specifications without prior notice HUKSEFLUX™ THERMAL SENSORS . WWW.HUKSEFLUX.COM . [email protected]

 Open the catalog to page 1

All Hukx catalogs and technical brochures

  1. SRA30-M2-D1

    3  Pages

  2. SR30-M2-D1

    7  Pages

Archived catalogs

  1. TRSYS20

    4  Pages

  2. SR25

    4  Pages

  3. SR15 series

    4  Pages

  4. TP08

    2  Pages

  5. TP02

    2  Pages

  6. TP01

    1  Page

  7. TRSYS01

    1  Page

  8. TNS01

    1  Page

  9. THISYS

    3  Pages

  10. WS01

    1  Page

  11. GT series

    2  Pages

  12. TPSYS02

    1  Page

  13. THASYS

    2  Pages

  14. IR02-TR

    1  Page

  15. DR01

    2  Pages

  16. ALUSYS

    1  Page

  17. SR11-TR

    1  Page

  18. NF02

    1  Page

  19. HFP01

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.