1. Catalogs
  2. HUBA CONTROL
  3. Flow sensor 210 0.5 ... 150 l/min

Flow sensor 210 0.5 ... 150 l/min

Flow sensor 210 0.5 ... 150 l/min
1 / 11 PagesView full catalog

Flow sensor 210 0.5 ... 150 l/min

Product catalog summary
Overview: The Huba Control Type 210 flow sensor utilizes the Kármán vortex trail principle, offering improved power supply and output signals over the OEM Type 200. It is designed for liquid media and can include temperature measurement capabilities.
Key Features:
  • Resistant to debris with no moving parts, ensuring high accuracy and minimal pressure loss.
  • Output options include voltage, current, pulse, or frequency.
  • Temperature measurement range from -40 to +125 ºC.
  • Approved for drinking water applications (KTW, W270, ACS, WRAS).
Specifications:
  • Flow range: 0.5 to 150 l/min.
  • Nominal diameters: DN 6, 8, 10, 15, 20, 25.
  • Accuracy: Less than 1% fs for under 50% fs and less than 2% measuring value for over 50% fs.
  • Response time: Immediate for spigot use, with a signal delay of less than 100 ms for unfiltered frequency output.
Operating Conditions:
  • Suitable for heating circuit water and other media upon request.
  • Maximum media temperature: +125 ºC; ambient temperature: -15 to +85 ºC.
  • Pressure ratings vary with temperature, up to 18 bar at +40 ºC.
Materials and Compatibility:
  • Sensor paddle made of ETFE; case with damming body made of PA6T/6I (40% GF).
  • Sealing materials include EPDM and FPM.
  • Electrical connection via M12x1 connector, IP 65 protection standard.
Electrical Specifications:
  • Power supply ranges from 4.75 to 33 VDC depending on output type.
  • Output options include analogue signals (0-10 V, 4-20 mA) and frequency outputs.
  • Includes short circuit, reverse voltage, and external voltage protection.
Accessories:
  • Connection kits available for various DN sizes with copper tubes or adapters.
  • Various O-rings and clips for different DN sizes.
Installation Guidelines:
  • Ensure correct sensor function by allowing only diameter changes from large to small.
  • Avoid repeated elbows at the entry side; recommended minimum distances provided.
Technical Specifications: Detailed specifications for connection tubes and accessories with diameters DN 8, 10, 15, and 20, including precise measurements for each diameter.
Electrical Connections: Descriptions of various electrical connections, including connectors with and without temperature measurement capabilities, current and voltage output configurations, and frequency and pulse outputs. Specific pin color codes and connection diagrams are provided.
Flow Measurement Formulas: Includes formulas for calculating flow rates based on frequency, impulse, and voltage outputs for different diameters (DN10, DN15, DN20, DN25), as well as formulas for current output in mA.
Viscosity Influence: Discusses the impact of media viscosity on measurement accuracy, particularly for media with higher viscosity than water, and provides corrections for kinematic viscosity of ethylene glycol-water and propylene glycol-water mixtures.
Graphical Data: Graphs illustrate the relationship between viscosity and temperature, minimal detectable flow, and the influence of viscosity on flow measurement accuracy.
Contact Information: Contact details for Huba Control AG's headquarters and various international branches, including phone numbers, fax numbers, and email addresses for customer support.
See more

Catalog excerpts

Flow sensor 210 0.5 ... 150 l/min-1

Vortex Flow sensor Flow sensor for liquid media Type 210 The flow sensor type 210 is based on the Kármán vortex trail principle. In comparison to the OEM flow sensor (type 200), the type 210 is available with enhanced power supply and output signals and is available with and without temperature measurement With no moving parts the flow sensor is not sensitive to debris, has marginal pressure loss and high accuracy. + Flow measuring with voltage, current, pulse or frequency output + Temperature non-sensitive measuring principle + Excellent media resistance (measuring element not in contact with the media) + Wide application temperature range + Marginal loss of pressure + Measuring element not sensitive to debris + Direct temperature measurement in the medium + Drinking water approval KTW, W270, ACS, WRAS Huba Control Type 210 | Technical data subject to change | Edition 09/2020

 Open the catalog to page 1
Flow sensor 210 0.5 ... 150 l/min-2

Technical overview Flow measurement Measuring principle Vortex Measuring range Nominal diameters Accuracy at < 50% fs (water) Accuracy at > 50% fs (water) Immediately. Therefore suitable for spigot use. Frequency output (unfiltered) Response time Frequency output (filtered) Analogue output Piezoelectric sensor element 0.5 ... 150 l/min DN 6 / 8 / 10 / 15 / 20 / 25 < 1% fs < 2% measuring value Signal delay Response time Signal delay Response time Operating conditions Suitable for heating circuit water with the usual additives Medium Other medium on request Drinking water Media Ambient Temperature...

 Open the catalog to page 2
Flow sensor 210 0.5 ... 150 l/min-3

Analogue output - Electrical overview Temperature measurement (> 8 DN) Measuring principle Resistance Measuring range @ T = 0 ºC PT1000 Accuracy class B DIN EN 60751 @ T ≠ 0 ºC Measuring range Accuracy 0 ... 10 V Calculation temperature Measuring range Accuracy 4 ... 20 mA Calculation temperature Electronic Voltage output Current output Dual power output Power supply 11.5 ... 33 VDC 8 ... 33 VDC 10 ... 33 VDC Output flow (Q) anaogue signal 0 ... 10 V 4 ... 20 mA 4 ... 20 mA 0 ... 10 V – 4 ... 20 mA Output temperature (T) signal Load agianst GND or IN < 6 mA / < 100 nF 1) < (UIN - 8 V) / 20 mA...

 Open the catalog to page 3
Flow sensor 210 0.5 ... 150 l/min-4

Frequency output (filtered) and pulse output - Electrical overview Temperature measurement (> 8 DN) Measuring principle Resistance Measuring range PT1000 @ T = 0 ºC Accuracy Class B DIN EN 60751 @ T ≠ 0 ºC Self-heating at temperature sensor Temperature influences Conduction resistance to connector Electronic Power Supply 4.75 ... 33 VDC Output flow (Q) Level height (open collector) < 0.5 ... > UIN - 0.5 V Output temperature (T) Resistant signal PT1000 class B DIN EN 60751 Load agianst GND or IN > 1 kΩ / < 10 kΩ < 3 mA Current consumption load free (IIN) Electrical reliability Short circuit, reverse...

 Open the catalog to page 4
Flow sensor 210 0.5 ... 150 l/min-5

Frequency output (unfiltered) - Electrical overview Temperature measurement (> 8 DN) Measuring principle Resistance Measuring range PT1000 @ T = 0 ºC Accuracy Class B DIN EN 60751 @ T ≠ 0 ºC Self-heating at temperature sensor Temperature influences Conduction resistance to connector Electronic Power Supply 4.75 ... 33 VDC Output flow (Q) Level height (push-pull) < 0.5 ... > UIN - 0.5 V Output temperature (T) Resistant signal PT1000 class B DIN EN 60751 Load agianst GND or IN < 1 mA / < 100 nF < 2 mA Current consumption load free (IIN) Electrical reliability Short circuit, reverse voltage and...

 Open the catalog to page 5
Flow sensor 210 0.5 ... 150 l/min-6

Accessories (supplied loose) Order number Connection kit DN 8, 10 with copper tube Connection kit 1) DN 8, 10 with adapter Rp ⅜ Stainless steel 1.4305/AISI 303 Connection kit 1) DN 15 with copper tube Connection kit 1) DN 15 with adapter Rp ½ Stainless steel 1.4305/AISI 303 Connection kit 1) DN 20 with copper tube Connection kit 1) DN 20 with adapter Rp ¾ Stainless steel 1.4305/AISI 303 Straight-wire box for connector M12x1 with cable 3-pole 200 cm Corner-wire box for connector M12x1 with cable 3-pole 200 cm Straight-wire box for connector M12x1 with cable 5-pole 200 cm (with temperature) Corner-wire...

 Open the catalog to page 6
Flow sensor 210 0.5 ... 150 l/min-7

Huba Control Type 210 | Technical data subject to change | Edition 09/202

 Open the catalog to page 7
Flow sensor 210 0.5 ... 150 l/min-8

Admissible locking torque inside thread Geometry of customers connection tube DN 8, 10, 15, 20 o-ring Huba Control Type 210 | Technic

 Open the catalog to page 8
Flow sensor 210 0.5 ... 150 l/min-9

Connector M12x1 without temperature measurement Frequency output filtered Pulse output Current output Ω Voltage output with temperature measurement PT1000 Current output with temperature measurement PT1000 Current output with temperature measurement 4 ... 20 mA Frequency output filtered Pulse output frequency output with temperature measurement PT1000 Voltage output Connector M12x1 with temperature measurement Frequency output unfiltered Voltage output with temperature measurement 0 ...10 V Pin Colour 1 brown 3 blue 1 4 black 1 brown 2 white 3 blue 2 4 black 5 gray «OUT T» is only in operation...

 Open the catalog to page 9
Flow sensor 210 0.5 ... 150 l/min-10

With the following definitions we are able to correct the influence of media with higher viscosity than water (= media viscosity > 1.8 cST) in order to reach 0 a measuring accuracy of 3% fs in the range of 1.8 - 4 cST and of 4% in the range of 1 - 14 cSt3(υ = 4 4 2 viscosity 6 cSt). 8 in 7 5 9 10 11 12 13 14 Viskosität in cSt Minimal detektierbarer Durchfluss Definition of viscosity of glycol-water-compound Kinematische Viskosität Propylenglykol‐Wasser Kinematic viscosity propylene glycol water Minimaler Durchfluss in l/min 25 Kinematische Viskosität Ethylenglykol‐Wasser Kinematic viscosity ethylene...

 Open the catalog to page 10
Flow sensor 210 0.5 ... 150 l/min-11

Huba Control AG Headquarters Schweiz Industriestrasse 17 CH-5436 Würenlos Telefon Niederlassung Deutschland Vestiging Nederland Succursale France Branch Office United Kingdom Unit 13 Berkshire House, County Park Technopôle Forbach-Sud Business Centre, Shrivenham Road

 Open the catalog to page 11

All HUBA CONTROL catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.