Power Frequency CT PT Analyzer Fast delivery Product Introduction HZCT-IV Power Frequency CT PT Analyzer is a new generation of innovative CT, PT test instrument which is launched market after a lot of market research and through the identification of strict professional testing and certification authorities. Device adopts high performance processor and microcomputer control architecture, the advanced manufacturing technology ,which ensure the stable and reliable product performance, complete functions, high degree of automation, high testing efficiency. CT Volt-Ampere Characteristic Test - Secondary Voltage / Current: 0-2500 V, 0-20 AResolution: 1 mV, 1mA CT Ratio Error & Phase Angle Error Test - Primary Side Current: 0-1000 A Resolution: 1 mA CT Ratio Error & Phase Angle Error Test - Secondary Side Current: 0-5 A Resolution: 1 mA CT Secondary Loop Load Test - Secondary Side Current: 0-5 A Resolution: 1 mA CT Secondary Circuit Impedance: 0-10 Q Resolution: 0.001 Q CT Primary Current Test - Primary Side Current: 0-1000 A Resolution: 1 mA PT Test Functions PT Volt-Ampere Characteristic Test - Output Voltage: 0-200 V Resolution: 1 mV PT Volt-Ampere Characteristic Test - Output Current: 0-10 A Resolution: 1 mA PT Ratio & Phase Angle Error Test - Primary Side Voltage: 0-2500 V Resolution: 1 mV PT Ratio & Phase Angle Error Test - Secondary Rated Voltage: 100 V, 57.7 V, 100/3 V