Overview: The document describes the Paste Cell, an accessory designed for measuring particle size in samples that cannot be diluted or are dispersed in viscous media. It ensures accurate measurement without altering dispersion conditions, making it suitable for precious materials.
Specifications: The Paste Cell can handle sample amounts ranging from 10 mg to 5 g, depending on particle size, distribution, and density. It features an in-line ultrasonic probe with 30 W, 20 kHz, and adjustable levels, along with a fully automated circulation pump with 15 adjustable speeds and 4 selectable fill levels.
Compressed Air Requirements: The system requires a compressed air supply with an origin pressure of 0.4 - 0.8 MPa and a controlling range of 0.01 - 0.4 MPa.
Applications: Particle size analysis is crucial for evaluating new formulations, characterizing raw materials, and ensuring product quality. It is particularly important in industries such as battery electrode materials and inkjet printing.
Technology: The LA-960 system offers state-of-the-art nanoparticle measurement with a wide range of applications. It includes NIST-traceable size standards and a user-friendly optical bench. The Method Expert software facilitates the creation of robust methods for research and quality control.
Global Presence: HORIBA has a global presence with offices in various countries, including Japan, China, India, Korea, the USA, Brazil, Germany, France, Singapore, Vietnam, Indonesia, Thailand, Italy, and the UK.
Certifications: The HORIBA Group integrates several management systems, including ISO9001, ISO14001, OHSAS18001, and ISO22301, to ensure stable product and service delivery even in emergencies.
See more

Catalog excerpts

LA-960-2

HORIBA proudly introduces the newest breakthrough in particle size technology. ‘ . . ' - ' . J tr W This latest evolution in the LA series advances scientific knowledge for tomorrow's world through intuitive software, unique accessories, and high performance. The LA-960 continues HORIBA's proud tradition of leading the industry with innovative design.

 Open the catalog to page 2
LA-960-4

HORIBA’s Original Optical DesignHORIBA's ground breaking optical design perfects the static light scattering particle sizing technique. Advanced Detector Design The number of detectors, angular range, and layout each contribute to overall system performance. The LA-960 uses 87 logarithmically spaced silicon photodiode detectors covering a range of 0.006 -165.7 degrees to measure complete particle size distributions. Superior Instrument-to-Instrument Precision The LA-960 is designed and built to provide the same experience regardless of manufacture date, operator skill, or geographic location....

 Open the catalog to page 4
LA-960-5

State of the Art Sampling Systems Auto-fill Circulation Flow cell The LA-960 Powderjet combines several unique and patented features to provide the most reproducible dry measurements. Use the Auto Measurement function to control vacuum, air pressure, powder flow, start/stop conditions, measurement duration, and data processing. Designed to handle every application including small sample amounts, friable powders, and highly agglomerated materials. Wet Measurement Sample-to-sample analysis in less than 6,0 seconds The HORIBA LA-960 wet circulation system is an easy, fast and very powerful dispersion...

 Open the catalog to page 5
LA-960-6

Powderjet Dry Feeder Accessory Laser scattering particle size distribution analyzer Model: LA-960 Measurement Principle ^Windows is a registered trademark of Microsoft Corporation in the United States and other countries. 5^ When ordering the Powderjet please specify the power requirements for the final destination. ^ Above specifications and functionality are valid only when Powderjet is installed on the LA-960 main unit and controlled using the LA-960 software. ^Manufacturers and models indicated for vacuum, air compressor, computer, monitor, and/or printer are subject to change. ©Air Compressor Inlet...

 Open the catalog to page 6
LA-960-7

Functional polymers The size of particles greatly influences several factors such as dissolution rate, dosing, bioavailability, and immunotoxic-ity, making it an essential parameter for pharmaceutical and biotechnology applications. From simple inhalers to advanced chemotherapies, particle size affects treatment efficacy. The performance of plastics, such as PET is determined by the molecular weight (i.e., size) of the polymers used to prepare the material. The same sort of analysis is also critical for sealants and adhesives. Particle size influences both capacity and coulombic efficiency of...

 Open the catalog to page 7
LA-960-8

State of the art nanoparticle measurement Measurement range 10 nm - 5,000 um The LA-960 features a wide measurement range to measure every application. The unique optical bench is user-friendly and standard in every LA-960 configuration. The advanced design of the LA-960 allows for easy measurement of nanoparticle applications. NIST-traceable size standards verify that the LA-960 accurately measures peaks as fine as 30 nanometers. 60 - second measurement cycle, even in wet mode This incredible speed is made possible by automatic laser alignment, fully automated liquid handling, and intuitive...

 Open the catalog to page 8
LA-960-9

HORIBA Vietnam ORIB OR tna na a Company Limited ompany Limited pan imited mt HORIBA Jobin Yvon GmbH mbH b Head Office HORIBA TAL Sr HO IBA ITALIA S l Mi n Milano Milano Office HORIBA In ia R India Privat td Private Ltd. v t Bangalore Office Off ce HORIBA (Thailand) Ltd. BA East Office s st HORIBA Korea Ltd. Seoul Branch HORIBA Instruments Inc. Edison office Guangzhou B a uangzh Branch Bran Head Office Beijing Branch HORIBA Instruments Inc. A Edison Ltda. Brasil, office The HORIBA Group adopts IMS (Integrated Management System) which integrates Quality Management System ISO9001, Environmental...

 Open the catalog to page 9
LA-960-10

HORIBA Scientific Automotive Test Systems I Process & Environmental I Medical I Semiconductor I Scientific

 Open the catalog to page 10

Archived catalogs

  1. B-700series

    5  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.