Model HGD 113

Model HGD 113

Model HGD 113

Product catalog summary
Model Overview:
The HGD 113 is an autoclave with a total chamber volume of 132.9 liters and an effective volume of 113 liters. It features chamber dimensions of 500 mm diameter by 576 mm height and external dimensions of 686 mm width, 930 mm depth, and 870 mm height. The working temperature ranges from 105°C to 135°C, with a maximum chamber pressure of 2.5 bar. It operates on a power supply of 400 Volt/50Hz/3 phases, with a power consumption of 6.8 KW and a net weight of 184 kg. The autoclave can accommodate two baskets.
Technical Features:
The base model includes a separate steam generator and a vacuum function for pre- and drying vacuum. It is fully automatic with microprocessor control and offers 12 installed programs, with options for special program adjustments. A timer-controlled night program allows for overnight sterilization without heating the laboratory. Additional programs include AGAR production, drying with vacuum and chamber warming, liquid sterilization with removal temperatures between 60-95°C, solid sterilization for glasses and instruments, and a steam pot program. The autoclave is equipped with a USB connection, water cooling system, automatic water supply, and a lid that opens and closes with a button press. It is mobile, allowing for cleaning underneath.
Safety Features:
Sterilization parameters are locked once a program starts. Safety features include a TÜV-proofed safety valve, automatic door locking with sensor control, a temperature limiter, and a large illuminated display for course announcements. The autoclave has a collective connection for resignation administrations, a bottom plate for grounding baskets and vessels, and a safety sensor for heating elements. It is approved by TÜV Bayern e.V. and CE-marked.
Construction and Maintenance:
The chamber and lid are made from stainless steel (SUS 304) with a polished finish for easy cleaning. The housing is powder-coated. The autoclave features heat protection under the lid handle and a built-in main switch. It has recessed handles for easy manipulation. The feed-water level is monitored, and the autoclave cannot start without sufficient water. An optional feed water connection automatically controls the water level. The exhaust steam condenser prevents fogging and odors. The lid remains closed until the chamber pressure is safe, and a heat-resistant plastic cover protects against hot parts.
Programs and Operation:
The autoclave offers 12 sterilization programs, including 3 for liquid sterilization, 3 for solid sterilization, 3 for textile sterilization with vacuum function, and 3 AGAR programs. Sterilization temperature and time can be adjusted before starting. The steam exhaust cooling time is adjustable for sensitive media. A timer-controlled start allows for overnight sterilization, providing ready-to-use AGAR by morning. The process is fully automatic, with temperature control inside liquids or the chamber, and pressure control for efficient power consumption and fast processing.
See more

Catalog excerpts

Model HGD 113-1

Model HGD 113 Autoclave, Chamber volume total/effective: 132,9/113 litre Chamber dimensions (mm): 500 Ǿ x 576 H External dimensions (mm): 686B x 930T x 870H Working temperature: 105• C – 135• C Max. Chamber pressure: 2,5 bar Power: 6,8 KW Power supply: 400 Volt/50Hz/3phases Net weight: 184kg Capacity baskets 2 Description and technical information: Contained in the base price: Separate steam generator Vacuum function: Pre- and Drying vacuum Fully automatic microprocessor controlled 12 installed programs, special programs adjustable Timer-controlled night program - overnight sterilization does not heat up the laboratory AGAR program with thermostatic heating for AGAR Production Drying program with vacuum and chamber warming Liquid program with removal temperature of 60- 95•C Solid Program for glasses and instruments Steam pot program USB Connection, Water cooling system Automatically water supply Lid-opening and lid-closing by pressing a bottom On wheels - cleaning is possible even under the autoclave Safety: Sterilisation parameters can’t be changed after starting the program run Safety valve TUV proofed Automatically door locking, sensor controlled and spring tension limited Limiter for max. chamber temperature Course announcement, big illuminated display Collective connection for resignation administrations Bottom plate to ground baskets and vessels Safety sensor for heating elements Approval by T„V Bayern e.V. CE-mark Chamber and basics: Lid with fast locking device and pressure are made from stainless steel (SUS 304). For easy cleaning the chamber is polished. The housing is powder coated and also easy to clean. additionally heat protection under the lid handle. The main switch is built-in and is protected by unintentional using. Two recessed handles facilitate easy manipulation. V:\HMC-Europe\Beschreibungstexte_Flyer_ Datenblätter\HG\Datenblätter\Datenblätter englisch\HGD\Angebotstext HGD 113_engl.doc

 Open the catalog to page 1
Model HGD 113-2

The feed-water-level is controlled and low water level is indicated. For protection of heating elements the autoclave cannot be started without enough water. An optional feed water connection automatically controls the water level. The exhaust steam condenser cools steam and condensate. This prevents fogging of instruments and unpleasant odours. The central quick lock can be actuated with one finger touch. An optional thermo lock with sterilization time control prevent hazards to the laboratory staff by delayed boiling. The lid remains closed until overpressure inside of sterilization chamber....

 Open the catalog to page 2