Z4010 portable oxygen analyser-1

Z4010 portable oxygen analyser Features N Measures from 100% to less than 1ppm N Fast response time of less than 5s for 90% change N Fully autoranging N Fully programmable analogue output Two alarms channels, user-config- analyser measures a wide range of urable for high, low or off states and oxygen concentrations from percentage hysteresis, each provide one set of volt- levels to less than ppm in clean non- free changeover contacts. They can be reactive gas mixtures. Its robust carry- ing case, fast warm up time and switch- span of the instrument. A choice of two able supply voltage provide an ideal analogue outputs is available, 0 to 5V solution for on-the-spot oxygen monitor- or 4 to 20mA, each with several pre-set ranges. Sample flow adjustment and monitoring are built-in, along with a A large multi-digit, autoranging sample pump for applications with a LCD is used to display measured con- centrations and also user-adjustable parameters for alarm and analogue out- The sampling system is a simple needle valve and flowmeter. Options analogue output provides an easy include: sample pump, fast flow loop means to log results where necessary. system, syringe injection port. Hitech also supply a range of sample conditioning accessories to cater for most sample sources. Applications N Air separation plants N Nitrogen purged soldering systems N Cylinder gas quality N Pharmaceutical and food packaging and storing N Purge gas monitoring N Glove boxes N Workplace air monitoring N Medical monitoring

 Open the catalog to page 1
Z4010 portable oxygen analyser-2

SPECIFICATION Display Multi-digit LCD – character height 12.7mm Ranges 0.1ppm to 100% oxygen, autoranging Display resolution From 10% to 100%: From 0.50% to 9.99%: From 500ppm to 4999ppm: From 50ppm to 499ppm: From 0.00ppm to 49.9ppm: Accuracy 10% to 100ppm: 99ppm to 10ppm: 0.1ppm to 9.9ppm: Stability Better than 2% of reading or 0.5ppm/month, whichever is greater Sample flow Between 100 and 250ml/min for optimum operation. Speed of response in clean inert atmospheres Percentage levels: less than 0.5s for 90% change Levels from 1000 to 10ppm: less than 5s for 90% change Levels less than 100ppm:...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.