Catalog excerpts
HITECH INSTRUMENTS A member of the MTL Instruments Group technical datasheet Z1030 Oxygen Analysers Zirconia sensor analysers for oxygen • Proven technology from gas analyser experts • Remote sensor – great flexibility • Fast response sensor • ppm to % O2 levels • Non-depleting, long life sensor • Bi-directional RS232 Applications • Nitrogen generators • Inert gas systems • Air separation • Soldering systems • Reflow ovens • Packaging The Z series of oxygen analysers features sensors based on zirconium oxide technology. They have the advantages of fast response times, excellent accuracy at ppm levels and a non-depleting sensor for long life. A range of different configurations and internal sample systems are available – so we can find the product that is right for you. The competitively priced Z1030 is the workhorse panel mount analyser. The sensor is supplied in a remote housing complete with flowmeter and valve giving more versatility to the user. Sensor and analyser can be mounted separately - where you want them. The analyser features bi-directional RS232 allowing remote calibration and datalogging. There is a choice of outputs and both outputs and alarms are user programmable. Not all zirconia sensors are the same. Hitech Instruments manufacture zirconia sensors in-house. Careful choice of raw materials and Hitech’s manufacturing ‘know how’ result in a fast responding, robust sensor with excellent specification and a very fast warm-up time. Different sample systems are available for applications involving long sample paths and high pressure. Bespoke designs are also available according to the application. Oxygen is measured in a wide variety of processes to confirm the presence or the absence of it. The most common application is to measure oxygen as the impurity in other gases, e.g. nitrogen generators. Oxygen is required in aerobic processes and situations that require user intervention, e.g. to ensure that it is safe to enter a chamber for maintenance work. Some processes require the absence of oxygen to be efficient, e.g. furnace atmospheres, others just require a low level of oxygen for safety reasons, e.g. inert gas blanketing. Hitech offer a number of different solutions for oxygen measurement dependent on the application. Our technical sales engineers will be pleased to advise the system that is right for you. The Z series from Hitech Instruments – the last word in fast, ppm oxygen measurement.
Open the catalog to page 1SPECIFICATION Display Multi-digit LCD - characters 12.7mm Display range 0.01ppm to 100%, auto ranging Display Resolution From 10.0% to 99.9% From 1.00% to 9.99% From 0.100 to 0.999% From 100ppm to 999ppm From 10.0ppm to 99.9ppm From 0.00ppm to 9.99ppm ORDERING INFORMATION Part no. Panel mount oxygen analyser with remote sensor assembly, 4-20mA or 0-5V output Double ferrule sample connectors in brass or stainless steel Terminal interface software to provide datalogging Contact Hitech for further details Electronics enclosure - showing dimensions Panel cutout 92 x 92 +1 mm Stability Better...
Open the catalog to page 2All Hitech Instruments catalogs and technical brochures
-
Z130 oxygen analysers
2 Pages
-
Z230 oxygen analysers
2 Pages
-
Biogas Analyser Instruments
8 Pages