1. Catalogs
  2. HIOKI EUROPE GmbH
  3. POWER QUALITY ANALYZER PQ3198
video corpo

POWER QUALITY ANALYZER PQ3198

POWER QUALITY ANALYZER PQ3198
1 / 24 PagesView full catalog

POWER QUALITY ANALYZER PQ3198

Product catalog summary
Overview
The document provides a comprehensive overview of the Power Quality Analyzers PQ3198 and PQ3100, focusing on their capabilities in measuring and analyzing power characteristics to address power-related issues. These devices are designed to assist field personnel in efficiently maintaining and managing power supplies.
Specifications and Features
  • PQ3198: A high-end model with advanced troubleshooting capabilities, capable of measuring transient voltages up to 6000 V and high-order harmonic components. It complies with IEC 61000-4-30 Ed. 2 Class A standards for high precision and reliability.
  • PQ3100: A mid-range model for investigating power supply conditions and preventing problems, featuring simple setup with QUICK SET, 8 hours of battery operation, and the ability to record event waveforms for up to 11 seconds.
Measurement Capabilities
  • Both models measure AC/DC power, record fluctuations in voltage, current, power, harmonics, and flicker, and analyze equipment power problems.
  • The PQ3198 offers two-circuit measurement, high-voltage, wideband performance, and GPS time synchronization.
  • The PQ3100 provides features like demand recording and display of event statistics.
Applications
  • Suitable for verifying power quality from solar power systems and EV rapid chargers, performing load rejection testing, and capturing power anomalies.
  • The PQ3198 is particularly useful for high-precision inverter measurements and simultaneous measurement of multiple circuits.
Data Recording and Analysis
  • Both models can simultaneously record event waveforms and trend graphs, ensuring comprehensive data capture.
  • They offer extensive event parameters and simple, one-touch setup for capturing power supply anomalies reliably.
Conclusion
The PQ3198 and PQ3100 Power Quality Analyzers provide robust solutions for analyzing and managing power quality issues, with features tailored to meet the needs of both high-end and mid-range applications. Their compliance with international standards and ease of use make them valuable tools for engineers and field personnel.
Measurement Specifications
  • Active, reactive, and apparent power demand values have a measurement accuracy of ±10 digits. Cumulative time accuracy is ±10 ppm ±1 sec. at 23°C. Power factor demand value is calculated from active and reactive power demand values with an accuracy of ±1 digit.
  • Voltage and current unbalance factors are calculated for 3-phase circuits using fundamental components. Voltage unbalance accuracy is ±0.15%.
  • Harmonic voltage and current measurements have varying accuracies depending on the order, with specific accuracies for 0th to 50th orders.
  • Inter-harmonic voltage and current components are measured with defined accuracies for inputs above 1% of nominal voltage.
  • THD is measured relative to wave and fundamental wave with an accuracy of 0.5%.
  • High-order Harmonic Components are calculated using the true RMS method, with specific measurement bands and accuracies.
  • Instantaneous flicker values are measured as per IEC 61000-4-15, with specific accuracies for Pst and ΔV10 flicker values.
  • RMS value frequency characteristics are defined for various frequency ranges with specific accuracies.
Recording Settings
Various recording intervals and settings for event detection and waveform recording are specified.
Interfaces and General Specifications
Includes details on SD card compatibility, LAN and USB interfaces, operating conditions, and standard compliance.
Power Supply and Memory
Details on power supply options, battery pack specifications, and internal memory capacity.
Display and Dimensions
Features a 6.5-inch TFT color LCD with multiple language options, and provides external dimensions and weight.
Current Sensors
  • Models include AC/DC Auto-Zero Current Sensors (CT7731, CT7736, CT7742), AC Flexible Current Sensors (CT7044, CT7045, CT7046), and AC Leak Current Sensor (CT7116). Each model has specific rated measured current, measurable wire diameter, and accuracy specifications.
Accuracy and Range
Current sensors have specified accuracy ranges for different current levels, with accuracy guaranteed up to 120% of the range.
Dimensions and Weight
Each sensor model has specific dimensions and weight, with cord lengths typically at 2.5 meters.
Accessories and Options
  • Various accessories are available, including voltage cords, adapters, battery packs, and carrying cases.
  • Options for connecting legacy current sensor models and extension cables are also provided.
Usage Recommendations
The document emphasizes using genuine HIOKI SD memory cards for data integrity and provides options for voltage measurement and current sensor connections.
Additional Information
Includes contact information for HIOKI headquarters and notes on the availability of options and accessories.
See more

Catalog excerpts

POWER QUALITY ANALYZER PQ3198-1

POWER QUALITY ANALYZER PQ3198, PQ3100 [new] Investigate power characteristics and analyze the causes of problems Exceptional ease of use and international standard-compliant reliability Analytical Software

 Open the catalog to page 1
POWER QUALITY ANALYZER PQ3198-2

Maintain and manage power supplies and analyze problems more easily and reliably than ever before POWER QUALITY ANALYZER PQ3198 and PQ3100 The critical importance of electrical power in today's society necessitates daily maintenance and management to ensure that problems don’t occur. When they do, for example due to an equipment failure or abrupt surge in demand, engineers face the need to analyze the cause quickly. The POWER QUALITY ANALYZER PQ3198 and PQ3100 provide robust support for field personnel who need to analyze power characteristics in the form of measurement capabilities that reliably...

 Open the catalog to page 2
POWER QUALITY ANALYZER PQ3198-3

Analyze equipment power problems Capture the full range of power supply anomalies, including momentary interruptions, voltage drops, and frequency fluctuations, while recording trends to help investigate the causes of unexpected equipment malfunctions and sudden stoppages. Record quality data for power systems Record fluctuations in voltage, current, power, harmonics, and flicker when connecting a highly variable system such as a renewable energy source or EV charging station to the grid. Easily analyze the data with the included PQ ONE software. Measure AC/DC power Use AC/DC auto-zero current...

 Open the catalog to page 3
POWER QUALITY ANALYZER PQ3198-4

High-end model Troubleshoot power supplies and verify power quality Class A compliance under international standards Basic voltage measurement accuracy of ±0.1% High-voltage, wideband performance Two-circuit measurement Simple inverter measurement 400 Hz line measurement GPS time synchronization Extensive array of event measurement parameters Investigate power supply anomalies Verify the quality of power from a solar power system Verify the quality of power supplied by an EV rapid charger Investigate the causes of equipment failures and malfunctions, including issues that are difficult to identify,...

 Open the catalog to page 4
POWER QUALITY ANALYZER PQ3198-5

High-precision, wideband, broad-dynamic-range measurement The PQ3198 delivers the high-end specifications and high reliability needed to capture the full range of power anomalies and analyze the underlying data with a high degree of precision. The PQ3198 complies with the IEC 61000-4-30 Ed. 2 Class A standard. As a result, it can perform standard-mandated measurement tasks such as gapless, continuous calculation; detection of events such as swells, dips, and interruptions; and time synchronization using GPS (optional). Basic measurement accuracy (50/60 Hz) Thanks to basic measurement accuracy...

 Open the catalog to page 5
POWER QUALITY ANALYZER PQ3198-6

Mid-range model Investigate power supply conditions and prevent problems Features Simple setup with QUICK SET Record event waveforms of up to 11 sec. in duration 8 hours of battery operation 200 ms and 600 ms data save capability CAT III (1000 V)/CAT IV (600 V) Display event statistics Demand recording Investigate power supply conditions Measure voltage fluctuations, equipment capacity, and harmonics before installing new electrical equipment. You can also check whether newly installed equipment is affecting other equipment by repeating those measurements after installation and comparing the...

 Open the catalog to page 6
POWER QUALITY ANALYZER PQ3198-7

Launch QUICK SET to navigate the connection and setup processes so you can get started recording quickly. (example: 2-meter power measurement of a 3-phase/3-wire circuit) Connect the voltage cables and current sensors to the circuit to be measured. Current sensor Current range The instrument will perform an automatic wiring check and display the results. 1 i || Voltage phase 1 Current input II XJ/Turrent phase 1 Phase difference Bilx jPr tDPF) You need only set the recording parameters and interval in order to start measurement. Recording parameters can be set simply by choosing a simple setup...

 Open the catalog to page 7
POWER QUALITY ANALYZER PQ3198-8

Features shared by both the PQ3198 and the PQ3100 Measurement functionality and data recording capabilities that ensure you’ll capture the full picture with a single measurement Capture power anomalies reliably with simple settings The PQ3198 and PQ3100 can measure all parameters at once, including power, harmonics, and anomaly waveforms. The instruments also provide simple setup functionality for automatically configuring recording parameters for popular applications. Capture power supply anomalies reliably | Transient voltages Capture phenomena characterized by precipitous voltage changes and...

 Open the catalog to page 8
POWER QUALITY ANALYZER PQ3198-9

Extensive range of recording pa, rameters Simultaneously record event waveforms and trend graphs Each time it makes a measurement, the PQ3198/PQ3100 records trend data for all parameters. When a power anomaly is detected, an event is recorded. Since the instrument records the maximum, minimum, and average values during the interval, you can rest assured that you won’t miss peak values. Simultaneous recording of waveforms and trend data | Event waveform When an event occurs, the instrument records the instantaneous waveform for 0.2 seconds. Triggers can be set for all event parameters in parallel,...

 Open the catalog to page 9
POWER QUALITY ANALYZER PQ3198-10

I Flexible sensors: Easy installation in confined locations Flexible current sensors provide a convenient way to measure double- and triple-wired power supplies and in confined locations, with capacities of up to 6000 A. I Auto-zero sensors: Stable measurement of DC power over extended periods of time Auto-zero current sensors allow measurement of DC power over extended periods of time, eliminating the need to concern yourself with zero-point drift. | No need for an external power supply Since sensor power is supplied by the instrument, there’s no need for an AC adapter when using AC/DC sensors...

 Open the catalog to page 10
POWER QUALITY ANALYZER PQ3198-11

| When it's hard to clip leads to terminals In locations where it’s hard to attach alligator clip-style leads to metal terminals, you can replace the tips of the voltage cords with magnetic adapters so that you can more easily detect the voltage. | Secure the PQA to the side of a distribution panel Use two heavy-duty magnetic straps to attach the instrument to the side or door of a distribution panel. Magnetic adapters are easy to affix to terminals in confined locations. Magnetic straps can also be used to help keep voltage cords from coming loose. Extensive range of interfaces | Remote control...

 Open the catalog to page 11
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.