video corpo

MEMORY HiCORDER MR8740, MR8741
12Pages

{{requestButtons}}

Catalog excerpts

MEMORY HiCORDER MR8740, MR8741 - 1

Fully Integrate into High-Speed, Multi-channel Measurement Systems The MR8740 uses a two-block internal architecture, essentially giving it the capabilities of two MEMORY HiCORDERs. High-speed isolated measurement Simultaneous 20M sampling within the same block HIOKI company overview, new products, environmental considerations and other Information are available on our website. Digital Voltage Meter Measure minute changes in voltage at a high level of precision. Simultaneous measurement of all channels—rather than scanner-type measurement-dramatically reduces cycle times. Systems Integration

Open the catalog to page 1
MEMORY HiCORDER MR8740, MR8741 - 2

Are you having problems with multi-channel measurement or testing? “We're using multiple DMM units with a scanner to switch inputs. Measurement takes too long…” Reduced cycle times “We can't embed our oscilloscope, so we use it on a shelf. Our setup would be a lot sleeker if we could t it in.” “We need to perform many different types of measurements on a large number of channels.” Measure across multiple channels at the same time “Tall, large racks are dangerous in a production setting. I wonder if our setup can be made smaller…” “We're using multiple measuring instruments, and it's hard to...

Open the catalog to page 2
MEMORY HiCORDER MR8740, MR8741 - 3

Application Function testing of ECUs and EV inverter motors The MR8740/MR8741 can be used to perform a range of waveform measurements of inverter output. Torque sensor MEMORY HiCORDER ANALOG UNIT CURENT UNIT CLAMP SENSOR DVM UNIT The MR8740 simultaneously measures 34 analog voltage waveform channels, 3 current waveform channels, and 16 DC voltage channels. DC voltage is measured at a high level of precision with a DVM unit. (The instrument provides functionality for time-difference and statistical calculations for voltage waveforms.) Torque sensor • Torque • Rotation Testing of EV batteries...

Open the catalog to page 3
MEMORY HiCORDER MR8740, MR8741 - 4

The MR8740 is a rack-mountable instrument that can measure up to 54 channels. It uses a two-block architecture, essentially giving it the capabilities of two MEMORY HiCORDERs. Accommodates up to 27 measurement units. Two-block architecture (Block I: 16 units; block II: 11 units) Standard support for 16 logic channels Support for multi-channel measurement of up to 54 channels Independent block operation Switchable inter-block trigger synchronization Support for applications using different functions LOGIC Input terminal Block I : Analog 32ch, Logic 8ch Block II : Analog 22ch, Logic 8ch Since...

Open the catalog to page 4
MEMORY HiCORDER MR8740, MR8741 - 5

The MR8741 is a bench-top instrument that delivers affordable measurement performance. It features area judgment functionality and external control terminals. Accommodates up to 8 measurement units. Standard support for 16 logic channels Area judgment function and external control terminals [Front panel] External control terminals LOGIC terminais [Rear] A vent (fan), power inlet, and power switch are located on the rear of the instrument. DVI-D Connector 100BASE-TX Connector USB Connector (Type A, for USB memory stick or mouse) Use as a multi-channel WAVE COMPARATOR. High-speed waveform...

Open the catalog to page 5
MEMORY HiCORDER MR8740, MR8741 - 6

Unit-based architecture accommodates a variety of measurement applications. High precision and high resolution DIGITAL VOLTAGE METER New unit designed exclusively for the MR8740/MR8741 The MR8990 DVM UNIT is a two-channel DC voltage measurement unit designed exclusively for use with the MR8740/MR8741. It can measure minute fluctuations in output from sensors in automobiles and other equipment and voltage fluctuations in devices such as batteries at high levels of precision and resolution. High resolution: 24bit, 6.5-digit display Thanks to a resolution of 0.1 \N, the MR8990 can measure even...

Open the catalog to page 6
MEMORY HiCORDER MR8740, MR8741 - 7

Choose from nine input units according to your measurement application. The MR8740/MR8741 uses the same measurement units as the MR8847 MEMORY HiCORDER. Distortion measurement HIGH RESOLUTION UNIT 8968 Voltage measurement Voltage measurement Voltage measurement (DC/RMS selectable) Distortion measurement 2ch 2ch 2ch 2ch Isolated BNC connector Isolated BNC connector Isolated BNC connector Weidmuller SL 3.5/7/90G Input connectors (input impedance 1 MΩ, input capacitance 30 pF), (input impedance 1 MΩ, input capacitance 30 pF), (input impedance 1 MΩ, input capacitance 30 pF), (via Conversion...

Open the catalog to page 7
MEMORY HiCORDER MR8740, MR8741 - 8

Analyze data on a computer LAN terminal USB receptacle A (For USB memory device) Easy recording of measurement data Compatible to USB memory sticks Measurement data can be saved on any generic USB memory device. Measurement data can be easily recorded, and a USB flash drive can be used to easily copy data to a computer. LAN communications capability HTTP/FTP server function A 100BASE-TX LAN port is built in as standard equipment. Access the unit via a web browser running on a computer, for waveform observation and remote operation. Waveform data of the MR8740/MR8741 series can also be...

Open the catalog to page 8
MEMORY HiCORDER MR8740, MR8741 - 9

Convenient functions Connect a display and mouse to enable standalone use. Display and mouse connectivity Measure without using a PC. By connecting a display and mouse to the MR8740/ MR8741, you can display waveforms and operate the instrument with a mouse. The monitor display screen uses the same layout as the MR8847 Memory HiCorder series display. A mouse can be used to operate and congure the instrument, providing a user experience that approximates use of a keyboard. (Display and mouse not included.) Value monitor (DMM display) DMM display (example: measurement using the MR8990)...

Open the catalog to page 9
MEMORY HiCORDER MR8740, MR8741 - 10

FFT function Frequency area data analysis (FFT function) Specify an analysis point Electrical distortion analysis/mechanical vibration analysis FFT analysis function This function comprises single-signal FFT for tasks such as frequency component analysis, dual-signal FFT for transfer function analysis, and octave analysis for acoustic measurements. The signal source for analysis are selectable from 1,000 to 10,000 data points. FFT analysis from captured time domain data (used with Memory function) To use measurement data captured with the Memory function, the mouse serves to specify...

Open the catalog to page 10

All HIOKI E.E. CORPORATION catalogs and technical brochures