1. Catalogs
  2. HIOKI E.E. CORPORATION
  3. HIOKI LR5000 Data Logger Series
video corpo

HIOKI LR5000 Data Logger Series

HIOKI LR5000 Data Logger Series

HIOKI LR5000 Data Logger Series

Product catalog summary
Overview: The HIOKI LR5000 Series comprises compact data loggers for long-term recording of temperature, voltage, current, and instrumentation signals. These loggers feature expanded memory, user-friendly software, and the ability to continue recording during battery replacement.
Key Features:
  • Increased recording capacity, seven times that of previous models.
  • Capability for data import during recording and continuous measurement.
  • Intuitive PC software for efficient data management and analysis.
  • Flexible installation options, including magnetic straps and wall-mounted holders.
Applications:
  • Voltage Logger: Suitable for recording pyranometer output for insulation evaluation.
  • Clamp Logger: Monitors current consumption for energy-saving initiatives.
  • Instrumentation Logger: Records outputs from fluid flow and pressure sensors.
  • Temperature/Humidity Logger: Manages environmental conditions in offices and factories.
Operation:
  1. Install the data logger and configure the recording interval.
  2. Transfer data to a PC via USB or infrared communication.
  3. Utilize bundled software for data viewing and management.
Recording Capabilities:
  • Recording intervals range from 1 second to 60 minutes.
  • Statistical value recording for capturing detailed fluctuations.
  • Large internal memory capable of storing up to 60,000 data points per channel.
Advanced Features:
  • IP54 splash-proof rating for use in humid environments.
  • Battery life extends up to 2 years, with backup to preserve data during battery changes.
  • Compatibility with data from previous HIOKI models.
Software and Data Management:
  • LR5000 Utility program for data import, graph creation, and analysis.
  • Compatible with Windows XP, Vista, and 7 via USB interface.
  • Data can be combined and managed on a PC.
Specifications:
  • Humidity Logger LR5001 and Temperature Logger LR5011 models available.
  • Measurement ranges: Temperature (-40°C to 180°C), Humidity (0% to 100%rh).
  • Various sensor options for different applications.
Accessories:
  • Includes Magnetic Strap, Wall-mounted Holder, SD Memory Card, and various sensors.
Product Overview: The document details specifications for various HIOKI loggers and sensors, including the Clamp Logger LR5051, Instrumentation Logger LR5031, and Voltage Loggers LR5041, LR5042, and LR5043. It also covers clamp-on sensors and leak sensors.
Specifications:
  • Clamp Logger LR5051: Designed for recording load and leak currents at 50Hz/60Hz, with multiple measurement ranges and an accuracy of ±0.5%rdg. ±5dgt. Not waterproof, operates between 0˚C to 50˚C, with a battery life of up to 1 year under specific conditions.
  • Instrumentation Logger LR5031: Records 4-20 mA instrumentation signals, with a measurement range of DC -30.00 to 30.00mA, IP54 splash-proof construction, and operates between -20˚C to 70˚C.
  • Voltage Loggers LR5041, LR5042, LR5043: Measure DC voltage with ranges from -50.00mV to 50.00V, sharing similar waterproof and operating conditions as the LR5031.
Sensor Specifications:
  • Clamp on Sensors: Models 9669, CT6500, and 9695-02 detailed with conductor diameters, primary current ratings, and accuracy, rated for CAT III 600V or 300V depending on the model.
  • Leak Sensors: Models 9675 and 9657-10 specified for measuring leak currents with a primary current rating of 10A AC and accuracy of ±1.0%rdg, rated for CAT III 300V.
Accessories and Options: Each logger and sensor comes with specific accessories such as connection cables and instruction manuals. Optional communication adapters and data collectors are available for data analysis on a PC.
General Notes: Specifications are subject to change without notice, and the document includes contact information for HIOKI offices in Shanghai, Singapore, and India.
See more

Catalog excerpts

HIOKI LR5000 Data Logger Series-1

HIOKI DATA LOGGER LR5000 Sries Data Loggers 111 I I 80© EUEUEU:EU:EU.EU;EU © © =h ^ 5DmV ^ 5V ^ 50V rh C t mA Complte Line of Easy-to-Use Compact Loggers with Expanded Memory The new HIOKI compact data logger s騩ries easily records temprature, voltage, current. and instrumentation signais over long periods. Carried over from its highly reputed predecessor. this sries includcs featurcs and fu ne lions such as 7 times the recording capacity of former modcls. data import during record in g. continuons measuremeut even during battery replacement, and intuitive PC soft驭ware. Flexible and easy-io-use at single and multiple locations, the new HIOKI compact data logger sries is idcal for any application that requtes simple set-up but long-tcrmT rcliablc recording canabilif ics. www-hiokJ.com ; ISO 9001 ISO14001

 Open the catalog to page 1
HIOKI LR5000 Data Logger Series-2

Meet a Wide Variety of Data Logging Applications Instrumentation Logger/ voltage Logger Record fldd flow such as for water. gas and DlL Measure tlow meter outpul signais to monrtor flow trends Temprature Logger/Humidity Logger Manage the temprature and humkjfty in offices and tactories.\ Visually monitor the data to save on air conditionna ard neathg costs. Use as a voltage Logger to record pyranometer output for evaiuating insulatba ^4 Use as a Clamp Logger and leakage sensor to record and monitor leakage trends. Use as a temp驩rature Lcgger to record warehouse temperatues (or rtsualty monitoring...

 Open the catalog to page 2
HIOKI LR5000 Data Logger Series-3

Easy opration injustJ 3 steps ! Instal a Data Loggor, set an interval. and start measurtng, -I Easy to start recording Unlimited installation capabilities Set up&Record Set your recwding rrtervaL HoW the REC Ixjttcn tkxn one aaxrd (o 60 rrinutas) fcc two seccods to start recordng. frciritoicmpt Oab the Data Logger from me worksite and connect to a PC, 1 h ■ s Communication Adapter LR5091 Transfer datafrom Data Logger BtoPC■ Kiirii mia tan Data Lora* la the Data collector s internai memory, Impcrt data from up to 16 Data Loggers Instafed cr» site.' 1*0 A- rOnequires optimal Communication Adapter...

 Open the catalog to page 3
HIOKI LR5000 Data Logger Series-4

table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:11.00pt "Arial", sans-serif; } .font1 { font:18.00pt "Arial", sans-serif; } .font2 { font:3.50pt "Arial Black", sans-serif; } .font3 { font:7.00pt "Lucida Sans Unicode", sans-serif; } .font4 { font:8.00pt "Lucida Sans Unicode", sans-serif; } .font5 { font:11.00pt "Lucida Sans Unicode", sans-serif; } .font6 { font:12.00pt "Lucida Sans Unicode", sans-serif; } .font7 { font:15.00pt "Lucida Sans Unicode", sans-serif; } .font8 { font:17.00pt "Lucida Sans Unicode", sans-serif; } .font9 { font:20.00pt "Lucida Sans Unicode",...

 Open the catalog to page 4
HIOKI LR5000 Data Logger Series-5

Bundled Software Ensures Smooth and Easy Data Analysis Import data to a PC and create graphs Use the LR5000 Utilrty program to rnport Data Logger data to a PC to maks graphs and analyze data turther. Easity pmt results usina your PC. Show spcifie values using the cursor function Use the A/B cursors to seiect any location on a graph and dlspiay its value. The PC software can aLso cafculate maximum, minimum, and average values between A and B cursors. Simple file aggregation and management Transferred data can be combined wlth data prevrousty transterred (from the same Data Logger mit) into one...

 Open the catalog to page 5
HIOKI LR5000 Data Logger Series-6

6 Physical appearance Model Communication Adapter LR5091 Data Collector LR5092-20 Features Transfer data from a Data logger to a PC ՕTransfer Data Logger configurations or clock settings from a PC to the Data Logger Collect recorded data from the Data Logger to internai memory or SD card ՕView collected data in a graph Transfer Data Logger configurations or clock settings from internai memory or SD card to the Data Logger ՕTransfer data from a Data Logger to a PC Transfer Data Logger configurations or clock settings from a PC to the Data Logger Interface with Data Logger Infrared optical communications...

 Open the catalog to page 6
HIOKI LR5000 Data Logger Series-7

table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:3.50pt "Arial Black", sans-serif; } .font1 { font:7.00pt "Lucida Sans Unicode", sans-serif; } .font2 { font:8.00pt "Lucida Sans Unicode", sans-serif; } .font3 { font:11.00pt "Lucida Sans Unicode", sans-serif; } .font4 { font:5.00pt "Times New Roman", serif; } .font5 { font:6.00pt "Times New Roman", serif; } .font6 { font:6.30pt "Times New Roman", serif; } .font7 { font:7.00pt "Times New Roman", serif; } .font8 { font:11.00pt "Times New Roman", serif; } 7 Physical appearance *Sensor is sold separately. Model Humidity...

 Open the catalog to page 7
HIOKI LR5000 Data Logger Series-8

table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:3.50pt "Arial Black", sans-serif; } .font1 { font:5.00pt "Lucida Sans Unicode", sans-serif; } .font2 { font:7.00pt "Lucida Sans Unicode", sans-serif; } .font3 { font:8.00pt "Lucida Sans Unicode", sans-serif; } .font4 { font:10.00pt "Lucida Sans Unicode", sans-serif; } .font5 { font:12.00pt "Lucida Sans Unicode", sans-serif; } .font6 { font:28.00pt "Lucida Sans Unicode", sans-serif; } .font7 { font:6.00pt "Times New Roman", serif; } .font8 { font:6.30pt "Times New Roman", serif; } .font9 { font:7.00pt "Times New Roman",...

 Open the catalog to page 8

All HIOKI E.E. CORPORATION catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.