1. Catalogs
  2. HIOKI E.E. CORPORATION
  3. HIOKI FT3470 Series Magnetic Field HiTESTERs
video corpo

HIOKI FT3470 Series Magnetic Field HiTESTERs

HIOKI FT3470 Series Magnetic Field HiTESTERs

HIOKI FT3470 Series Magnetic Field HiTESTERs

Product catalog summary
Overview: The FT3470-51/-52 Magnetic Field HiTester is designed for measuring magnetic flux density in various environments, including household appliances and electrical power equipment. It complies with ICNIRP 2010 guidelines and other international standards, ensuring accurate and reliable measurements.
Specifications: The device measures magnetic fields across three axes (X, Y, Z) and calculates a composite value. It supports a frequency range from 10 Hz to 400 kHz and offers different measurement modes for public and occupational exposure levels. The device is equipped with a 100cm2 sensor and an optional 3cm2 sensor for detailed analysis.
Features: Key features include a memory function for storing up to 99 data points, waveform output for oscilloscope connection, and compatibility with the MEMORY HiCORDER MR8880-20 for extended data observation. The device also supports batch data export and is compatible with Windows XP, Vista, and 7.
Measurement Methods: The FT3470-50 series adheres to IEC 62110/IEEE 644 and IEC 62233 standards for magnetic field measurement. It provides true RMS measurement with auto/manual range switching and a display update rate adjustable via a slow function.
Output and Accuracy: The device offers both waveform and RMS value level outputs, with an accuracy of ±3.5% rdg. ±0.5% f.s. for most ranges. It includes a USB interface for data transfer and operates within a temperature range of 0 to 40°C.
Applications: Suitable for compliance testing of household appliances, environmental magnetic field measurement, and monitoring magnetic fields near electrical power equipment. The device is also useful for long-term measurement and waveform observation.
Additional Information: The FT3470-51/-52 comes with necessary accessories, including an AC adapter, USB cable, and application software. It is powered by four LR6 alkaline batteries, providing approximately 10 hours of continuous use.
See more

Catalog excerpts

HIOKI FT3470 Series Magnetic Field HiTESTERs-1

MAGNETIC FIELD HiTESTER FT3470-51/-52 Environmental Measuring Instruments Complies with ICNIRP 2010 Providing robust support for 3-axis magnetic ux density measurement Measurement of environmental magnetic elds Measurement of magnetic elds in the vicinity of electrical power equipment Compliance testing of household appliances

 Open the catalog to page 1
HIOKI FT3470 Series Magnetic Field HiTESTERs-2

2 Your one-stop solution for magnetic eld measurement The FT3470-50 Series complies with the ICNIRP 2010 guidelines as well as other relevant standards for evaluation testing. 1. International guidelines ICNIRP 2010 compliant. The guideline value has been changed to 200 µT (for public exposure) at 50/60 Hz. The FT3470-50 Series completely supports related measurements. 2. Magnetic eld measurement methods The FT3470-50 Series complies with IEC 62110/IEEE 644 as well as IEC 62233. 3. Magnetic eld measuring instrument requirements The FT3470-50 Series complies with IEC 61786. Measurement underneath...

 Open the catalog to page 2
HIOKI FT3470 Series Magnetic Field HiTESTERs-3

1. Simple operation for easy measurement 2. User-selectable display units The FT347Q-50 series can use different units of magnetic flux density as required by the applicable Also consider: POWER QUALITY ANALYZER PW3198 Record and Analyze Power Supply Problems Simultaneously with a Single Unit The New World Standard for Power Quality Analysis •Assess power quality problems ■ < accorda™ ili international standards (IEC61000-4-30 Class A] Hiijh prraiikvi. ijiiplcss rr.'üü'di'i;] (V: +0 ' rdc.. A ant W + 0.? rdr:. - 0.1 ".->'.s J •CATIV 600V ■ Safe enough tor incoming power lines • High-order harmonics...

 Open the catalog to page 3
HIOKI FT3470 Series Magnetic Field HiTESTERs-4

Measurement accuracy will be maintained when the tester and sensor are used in an environment where the temperature is 23°C ±5°C and humidity is 80% RH or less with no condensation ■ Magnetic flux density accuracy specifications FT3470-51/52 (with lOOcm* Sensor) ■ Exposure level (General Public/ Occupational) 'Smoothed edge: Exposure level Is here defined as the time doman evalutlon Introduced In IEC/ EN 62233 applied to the magnetic flux density Indicated In the ICNIRP 2010 Guidelines.) Ordering Information MAGNETIC FIELD HiTESTER FT3470-51 Packing contents: Magnetic Field HITester FT3470-50,...

 Open the catalog to page 4

All HIOKI E.E. CORPORATION catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.