HIOKI 8826 MEMORY HiCORDER - 1

32-channel recorder with large display and wide printer installed For use with the 8826. HIOKI pro\ ides a H ido assortment of input units suitable :■" i ' i ■■ i V - ■■■!' -. : IVI: h. 8H2G I M- i hiçdi ii ait form sampling rate of 1M sample/s which is carried out for all .'O channels simultaneously. In addition, the 8826 supports limitici m m id record i ¡i LI « ill I U idc 10.4-iiicli recording paper, a memory capacity of M nu^awonls for .ill y.', eh: m nol - (when expansion memory is

Open the catalog to page 1
HIOKI 8826 MEMORY HiCORDER - 2

2 Level Recorders Play a Vital Role in Many Areas of Modern Industry. - Features ● 10.4 inch color TFT display The large, bright display shows waveforms for all channels (32 analog + 32 logic channels) in 12 colors, allowing easy visual identification. The display also greatly facilitates operation of the unit. ● Simultaneous measurement in up to 32 analog channels Featuring the highest number of input channels in this class, the 8826 lets you simultaneously record in 32 analog + 32 logic channels. All analog inputs are isolated. ● Built-in PC card slot A PC card type III slot is provided...

Open the catalog to page 2
HIOKI 8826 MEMORY HiCORDER - 3

3 - Application Load shutdown tests in power generation plants For load shutdown tests in power generation plants, the pretrigger function can be used to measure and record waveforms before and after the test, enabling accurate analysis. The vernier function, which can be used to perform fine adjustment of amplitudes, is also a useful tool. Plant maintenance In plants, memory recorders are used to measure and record the operation of solenoid and control valves. Since the MEMORY HiCORDER 8826 can simultaneously create X-Y plots, the relationship between flow and valve lift can also be...

Open the catalog to page 3
HIOKI 8826 MEMORY HiCORDER - 4

4 High-Speed Response for Capturing Transient Events - Function Details Large memory capacity allows long-term recording of high- speed data A/D conversion Function outline of memory recorder Write on memory In the standard configuration, the 8826 can store a total of 16 mega-words, and with memory expansion 64 mega-words, using internal solid-state memory. This provides ample capacity to store data for all 32 channels. The table at right shows possible recording times, according to the time axis setting and the number of channels in use. A reduction in the number of channels prolongs the...

Open the catalog to page 4
HIOKI 8826 MEMORY HiCORDER - 5

5 - Function Details - Real time save function *supported in version 2.50 or later The real time save function enables data to be saved to the PC card while waveform is being measured. The compressed waveform is displayed on the screen at real time. This function is valid at the recorder and memory function. Write to the PC card can be up to 1kS/s (=100ms/DIV) for 16 channels, or up to 500S/s (200ms/ DIV) for 32 channels. With the real time save function, data measured at the pre-determined sampling below the maximum sampling can be saved to the PC card at any time. In addition, since the...

Open the catalog to page 5
HIOKI 8826 MEMORY HiCORDER - 6

6 - Function Details - Support for connection to PCs via Ethernet ■ Connect HiCORDER to departmental LAN (using TCP/IP communication protocol) The 8826 can be connected to Ethernet, a standard network protocol in the Internet age. (using the optional LAN CARD and LAN COMMUNICATOR 9333. Caution: The optional LAN card has been discontinued so new LAN applications cannot be supported.) For those who frequently analyze measurement data on PCs, this function offers a good match. Ethernet Note: Because LAN card, GP-IB card, and RS-232C card all use the same PC card slot of the 8826, when one of...

Open the catalog to page 6
HIOKI 8826 MEMORY HiCORDER - 7

7 - Function Details - Time domain FFT analysis functions FFT capability includes single-signal FFT for analyzing frequency components, two-signal FFT for transfer function analysis, and octave analysis for acoustic analysis. The source signal can be selected from waveform data captured by the memory recorder, and isolating required sections is also possible. (Number of data points: 1,000 to 10,000) Frequency domain Function outline of recorder (REC) The input signal is conver ted to digital for m and displayed and printed in real time. The maximum chart speed is 20 mm/s (in the 500 ms/DIV...

Open the catalog to page 7
HIOKI 8826 MEMORY HiCORDER - 8

8 Example Printouts & Screen (WIDE mode, actual size) B4 size paper (264 mm width) allows full-size mixed printing or division into up to 16 sections Example of B4 report output Screen image output at high resolution, B4 size. Example of X-Y plotting Time-axis waveforms of each input channel can be defined on vertical and horizontal axes to create X-Y plots. Example of Logging output This prints the instantaneous numerical value for each sample. ■ Online help and error indication Simply pressing the help key will bring up relevant information on functions and operation steps. If an error...

Open the catalog to page 8
HIOKI 8826 MEMORY HiCORDER - 9

Waveform monitor (color TFT LCD) Dial knob with analog sensibility nput module slots (for 16 modules; Main unit Note: The optional MO drive unit is discontinued External control connector _oglc probe connection (for S units) PC card slot Thernal printer

Open the catalog to page 9
HIOKI 8826 MEMORY HiCORDER - 10

10 ■ Main unit specifications REC & MEM functions (version 2.00 or later. Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm, approx. 300 g (10.6 oz) Accessories: None Real-time save function: version 2.20 or later) Time axis (REC) 20ms to 1hour/DIV; 16 settings, 1 DIV = 100 samples, time axis compression 1/2 to 1/500, 8 settings Time axis (MEM) 100µs to 5 minutes/DIV; 20 settings, 1 DIV = 100 samples, time axis zoom ×2 to ×10; 3 settings, compression 1/2 to 1/10,000, 12 settings Recording length Note: Sampling period 1/100 of time axis range at memory...

Open the catalog to page 10
HIOKI 8826 MEMORY HiCORDER - 11

Dimensions and mass: approx. 170 (6.69in) W x 20 (0.79in)H x 148.5 (5.85in) D mmj approx. 310 g (10.9 oz) Accessories: None Cable length and mass: Main unit cable 1.3 m (4.27 ft), input section cable 46 cm Cable length and mass: Main unit cable 1.5 m (4.92 ft), input section cable 30 cm Note: The unit-side plug of the 9320 is different from the 9320-01 Cable length and mass: Main unit cable 1.5 m (4.92 ft), input section cable 1 m Note: The unit-side plug of the MR9321 is different from the MR9321-01 Distribution media Display functions File loading Data conversion Print functions Running...