1. Catalogs
  2. HF scientific
  3. MicroTOL online turbidimeter_2014
video corpo

MicroTOL online turbidimeter_2014

MicroTOL online turbidimeter_2014

MicroTOL online turbidimeter_2014

Product catalog summary
Standard Features
  • Compliance with USEPA method 180.1 and ISO 7027 standards.
  • Available in models with ranges of 0-10 NTU, 0-100 NTU, or 0-1000 NTU.
  • One-piece design for easy installation and reduced module requirements.
  • Fast response and low-cost calibration with a 30 ml sample volume.
  • Modular design for cost efficiency.
  • Removable sample cuvettes for easy cleaning and calibration.
  • Optics not in contact with samples to prevent false readings.
  • Reusable primary calibration standards.
Instrument Overview

The Online Turbidimeter is designed for continuous turbidity measurement in various water applications, including filtered, raw, and wastewater. Optional HF Online software provides graphical trending, alarms, and filter analysis. An ultrasonic autoclean system is available for optical chamber maintenance.

Specifications
  • Optical design ensures consistent readings with laboratory and portable turbidimeters.
  • Bubble rejection system prevents air interference and cleans the optical chamber.
  • Calibration uses sealed cuvettes for a fast, clean, and reusable process.
  • White light and infrared light options for different regulatory standards and applications.
  • Compliance with USEPA method 180.1 and ISO 7027.
  • Optional data network interface for data collection and reporting.
  • Certified to CE, UL, and CSA standards.
Sample Specifications
  • Range: 0-10 NTU, 0-100 NTU, or 0-1000 NTU.
  • Measurement principle: Nephelometry (90 degrees).
  • Accuracy: 2% of reading or ±0.02 below 40 NTU; 5% above 40 NTU.
  • Resolution: 0.0001 NTU (selectable).
  • Response time: Adjustable (5 to 500 seconds).
  • Input pressure: 1-200 psi with built-in regulator set at 15 psi.
  • Standard outputs: 4-20 mA galvanic isolated or RS-485.
  • Light source: White light (5-year life) or infrared light (7-year life).
  • Operating temperature: 34°F to 122°F (1°C to 50°C).
  • Enclosure: NEMA 4X, IP66 for outdoor installation.
Dimensions and Installation

Dimensions are provided in millimeters and inches, with specific instructions for installation space and hardware requirements. The system includes a backpressure valve, drain tubing connection, and intake tubing connection for optimal operation.

Ordering Information

Models are available with various ranges and features, including ultrasonic autoclean and compliance with USEPA and ISO standards. Accessories and calibration kits are also available.

See more

Catalog excerpts

MicroTOL online turbidimeter_2014-1

A Watts Water Technologies Company Online Turbidimeter Specifications Instrument Overview The Online Turbidimeter is specifically designed for measuring turbidity continuously in filtered water, raw water, waste water final effluent and industrial applications. The optional HF Online software allows graphical trending, alarms and filter analysis for multiple networked turbidimeters. The optional ultrasonic autoclean system automatically cleans the optical chamber for finished or raw water applications. Standard Features • • • • • • • • Meets USEPA method 180.1 and ISO 7027 design and performance criteria Range of 0 - 10 NTU, 0 - 100 NTU or 0 - 1000 NTU (depending upon model) One-piece design eliminates the need to mount more than one module per turbidimeter Fast response time and inexpensive calibration due to low ( 30 ml ) sample volume Modular design reduces overall costs Removeable sample cuvettes allow for easy cleaning and calibrating Optics are not in contact with the sample which reduces the chance of false low readings Convenient reusable primary calibration standards Optical design New optical design allows consistant readings with laboratory and portable turbidimeters. Bubble rejection system The optical chamber of the has been designed to eliminate air in the sample while simultaneously creating a vortex cleaning action throughout the optical chamber. Calibration with primary standards is completed using sealed cuvettes, similar to laboratory procedures. This dry method of calibration is fast, clean and reusable. On-screen menu items guide you through the calibration procedure quickly and easily. One-piece mounted design allows for simple mounting and minimal use of space. Increased range of 0-1000 NTU allows for use of low NTU filtered water or raw water. New optical design increases accuracy and provides more consistent readings with online, laboratory and portable turbidimeters. Simple calibration Light source White light is recommended for use in turbidimeters reporting results under US EPA (US standard) jurisdiction. HF scientific has developed NEW krypton filled white light technology with lamp life expectancy up to 7 years. Infrared light is recommended for use in turbidimeters reporting results under ISO 7027 (European standard) jurisdiction. Infrared light is also recommended for waste water final effluent and industrial applications where color is present in the sample stream. Optional data network interface acquisition system The data acquisition system is designed to sequentially collect data from a series of interfaced HF scientific turbidimeters. The software system stores data, prints reports, graphs and alarms on each individual turbidimeter. In addition it can compare filters and monitor individual or multiple filter efficiency. Listed or Certified to CE, UL, CSA (ETL,ETLc) 3170 Metro Parkway, Ft. Myers FL 33916 USA • Phone (239) 337-2116 • Fax: (239) 332-7643 • Toll free: 888-203-7248 • [email protected] • www.hfscientific.com

 Open the catalog to page 1
MicroTOL online turbidimeter_2014-2

Sample specifications The continuous monitoring system shall include a single modular unit with power supply, display and sensor as one single unit. The turbidimeter shall meet all requirements specified by the USEPA Method 180.1 (White Light Model), ISO 7027 (Infrared Model) and Standard Methods 2130B. The turbidimeter shall have a similar optical design to a laboratory turbidimeter, for accuracy. The turbidimeter shall have consistent readings with laboratory and portable turbidimeters. The turbidimeter shall be Modbus compatible. The turbidimeter shall have the option of using an ultrasonic...

 Open the catalog to page 2
MicroTOL online turbidimeter_2014-3

Dimensions All Dimensions are in millimeters (inches). Provide at least 200 mm (8 inches) of free space above the sensor for easy removal of the flow head and insertion of the calibration standards. These 2 holes will accept up to 4.5 mm (3/16 inch) hardware. These 4 holes will accept up to 7 mm (1/4 inch) hardware (Holes are inside case) Shutoff clamp allows for shutoff of the intake flow during cuvette cleanings and replacements. Backpressure valve allows adjustment of the amount of back pressure, which will help to control flow rate and eliminate small bubbles. Inline Pressure Regulator Max....

 Open the catalog to page 3
MicroTOL online turbidimeter_2014-4

Ordering information* Catalog No. Ultrasonic Autoclean *All models are delivered fully calibrated and include 4-20mA, backlight display, RS-485/Modbus, inline pressure regulator, desiccant, universal power supply (100-240 VAC) & operator's manual. Model 2 also includes spare measuring cuvette w/light shield. HF Online Windows Software for data collection and reporting Remote display for an additional digital readout primary calibration kit, low range, .02, 1 and 10 NTU for TOL 5 or 6 primary calibration kit, .02, 10 and 100 NTU for TOL 3 primary calibration kit, full range, .02, 10 and 1000 NTU...

 Open the catalog to page 4

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.