video corpo

CLX-Ex

CLX-Ex

Product catalog summary
Overview
The HF scientific -Ex OnLine Residual Chlorine Monitor is designed for hazardous environments, particularly for shipboard ballast water treatment systems. It provides continuous monitoring and control of residual Total Chlorine levels, featuring user-selectable cycle times and outputs for controlling feed pumps and alarms.

Specifications
  • Measurement Range: 0 - 10 mg/l
  • Cycle Time: User selectable from 110 seconds to 10 minutes, with an optional range of 60 seconds to 10 minutes
  • Accuracy: 0 - 6 mg/l at ±5% or 0.03 mg/l of Cl2, whichever is greater; 6 - 10 mg/l at ±10%
  • Resolution: 0.01 mg/l
  • Method: USEPA accepted DPD method for Total Residual Chlorine
  • Outputs: 4-20 mA and RS-485 with Modbus
  • Operating Temperature: 0° - 55°C (32° to 131°F)
  • Sample Pressure: 0.3 bar - 10.3 bar (5 - 150 psi)
  • Compressed Air Pressure: 5.5 - 7 bar (80 - 101.5 psi)
  • Enclosure: IP66
  • Power: 240 VAC (130 to 240VAC), 47-63 Hz, 250VA
  • Certifications: CE, ATEX
  • Flow Rate to Waste: 200-400 ml/min

Features
  • Optimized for marine applications and suitable for environments with combustible fuel vapors
  • ATEX certified for Zone 1 and Zone 2 environments
  • Low operating costs due to minimal reagent use
  • Designed for unattended operation up to 90 days
  • Colorimetric DPD chemistry
  • User-selectable alarm for sample concentration
  • Applicable for ballast water treatment, drinking water, wastewater, and more

Accessories
  • Replacement tubing/cuvette kit (Cat. No. 09950)
  • CLX-Ex OnLine Chlorine Residual Monitor 240VAC (Cat. No. 28031)
  • Total Chlorine reagent for CLX-EX 12-month supply (Cat. No. 09991)

Dimensions and Weight
  • Shipping Dimensions: 558mm x 662mm x 400mm (22” x 24 1/2” x 15 3/4”)
  • Shipping Weight: 38 kg (83 lbs.)
  • Actual Weight: 26.3 kg (58 lbs.)
  • Installed Dimensions: 19.7” x 18.7” x 9.4” (50cm x 48cm x 23.8cm)
See more

Catalog excerpts

CLX-Ex-1

Water Quality Measurement Solutions A Watts Water Technologies Company CC^C-Ex OnLine Residual Chlorine Monitor for hazardous environments II The HF scientific cizx:-Ex OnLine Residual Chlorine Monitor is a rugged, safe, and low maintenance instrument for shipboard ballast water treatments systems that require continuous monitoring and control of residual Total Chlorine levels in hazardous environments. User selectable cycle times, 4-20mAand RS-485 with Modbus output that can be used to control feed pumps and a user selectable alarm allows complete manual or automatic control of chlorine dosing. Standard Features Optimized for marine applications Use in atmospheres where combustible fuel vapors may be Housing certified for ATEX, and Zone 1 and Zone 2 Total chlorine Can be used for Total Chlorine measurements Able to operate in ambient temperatures up to 55°C Low reagent use translates into low operating cost Simple operation Designed for unattended operation up to 90 days Cycle times Cycle times can vary between 110 seconds to 10 minutes (optional 60 User selectable alarm for sample concentration Ballast water treatment, Drinking water, Wastewater and more ATEX certified Suitable for Zone 1 and Zone 2 environments USEPA accepted method Ordering Information CLX-Ex OnLine Chlorine Residual Monitor 240VAC for Total Chlorine complete with operating manual. Reagents must be purchased separately. Specifications subject to change without notice. Reagent Kits Catalog number Total Chlorine reagent for CLX-EX 12 month supply (based on estimated BWTS operational cycles) HF scientific 3170 Metro Parkway, Ft. Myers FL 33916 e-mail: [email protected] ph: (239)-337-2116 • fax: (239) 332-7643 • toll free: 888-203-7248 www.hfscientific.com

 Open the catalog to page 1
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.