Product catalog summary
Overview: The document provides technical specifications and applications for the KACO Relay RY, a mains isolation relay used in various fields including medical equipment, communication engineering, and general applications.
Applications:
  • Medical Equipment: Suitable for dialysis machines, X-ray devices, defibrillators, and heating control for therapeutic water beds.
  • Communication Engineering: Used in remote signalling equipment.
  • General Applications: Applicable for current supplies, chargers, measuring equipment, and laboratory equipment.
Specifications:
  • The relay provides two-pole mains isolation with a contact gap greater than 3 mm.
  • It features double or reinforced insulation suitable for overvoltage category III and pollution degree 2, with a switching voltage of AC 230 V.
  • Various connection and mounting options are available, including PCB mounting, Faston connectors, and DIN rail or screw attachments.
  • The relay is made with self-extinguishing insulation material according to UL 94 standards.
Technical Details:
  • Contact materials include silver cadmium oxide, with tungsten available for special applications.
  • Standard versions come with two NO contacts, with options for one NO and one NC contact.
  • Approvals are available for the standard version with 2 NO contacts.
Performance:
  • Switching operations are detailed as functions of load, with specific data on switching current and number of operations.
  • Service life is determined for 20 switching operations per minute at an ohmic load of 230 V-AC.
  • Mechanical service life is approximately 3 x 107 cycles.
Electrical Characteristics:
  • Test voltages for coil/contacts and contact/contact are specified.
  • Ambient temperature and weight details are provided.
Mounting Options:
  • PCB mounting with solder pins, screw mount with Faston connectors, and standard rail fitting options are available.
See more

Catalog excerpts

relay RY en-2

The KACO relay RY is mainly used in deviceswhich require a 2 pole mains isolation accord-ing to DIN VDE 0106 part 101.As stipulated in e.g. DIN VDE 0631, 14 the RYrelay has an air gap and a creepage distance of>8 mm and an operational surge voltage of >4 kV. The contact gap is >3 mm. The standard version contains silver cadmium oxide contacts. Tungsten is also available as a possible contact material for special applications. The relay isproduced with two NO contacts or with one NOand one NC contact. On request, we offer a ver- sion with two NO contacts but without mains isolation. Because of...

 Open the catalog to page 2
relay RY en-4

Tab Screen MountingFaston KOO-.4RY_/1 Standard Rail FittingFaston > 3,7 For self-tapping screw according to DIN 7971 Nominal diameter 2.9 mm Dimensions without tolerance: 0.2 mm > 42,3 Switching operations as functions of the load Switching current Number of switching operations Service life at ohmic load (230 V-AC) determined for 20 switching operations/min. with contacts AgCdO >

 Open the catalog to page 4
relay RY en-5

Contact materialApparent power (50 Hz) Switching voltage max. Switching current max. Contact load max. Mechanical service life Pull-in/drop-out time (DC) Insulation group according to DIN VDE 0110Test voltage coil/contactsTest voltage contact/contact Ambient temperature T mW VA V-AC A W/VA Cycles ms ֖ V-AC V-AC ְC g AgCdO 2,5120/ 240 230/ 400 16 1.000 AgCdO2,5120/ 240 ח 230/ 400 10 1.000 > 1) / 3.6803 x 10 > 1) / 2.3003 x 10 > 7 7 ca. 15/3 C 250 B380 (400) 4.000 2.500ca. 50 ca. 15/3 C 250 B380 (400) 4.000 2.500ca. 50 > amb Weight > 1) sufficiently measured spark extinguishing. > Part No. RY_-4814480848015669...