Catalog excerpts
Radioisotope Fume Hoods UniFlow Radioisotope fume hoods are available in 48”, 60”, & 72” widths. Interior is constructed of welded type 304 stainless steel to prevent absorption of radioactive and corrosive materials. Stainless steel bafe is removable for ease of cleaning. Work surface is welded integral to the fume chamber and reinforced to support heavy isotope shielding materials. UniFlow Superstructure exclusive unitized dual wall construction for superior chemical resistance, strength,and durability. Access Panel removable to access ducting connections and electrical services from a single point electrical box,115V/60Hz AC operation. Full 30” Viewing Height ease of access and viewing tall apparatus setup. Sash is 3/16 thick tempered safety glass, vertical moving for efcient air ow and ease of movement Angled Picture Frame Opening the aerodynamic face opening with stainless steel air foil provides uniform air ow into the fume chamber. Vertical Sash to be counterbalanced connected with coated stainless steel cables, 3/16” tempered safety glass with stainless steel framing track, pull, and aerodynamic sash lift. Air Foil & Worksurface are constructed of type 304 stainless steel. Sash Stop constructed of stainless steel and located at 1/2 open sash height or to suit your specic requirements. Optional equipment Cat. No. 51651 Air Flow Monitor continuously monitors face velocity airflow, analog or digital available. Meets ANSI and OSHA requirements. Optional equipment. Fume Chamber surface is type 304 stainless steel, # 4 satin nish. VaraFlow Bafe System maintains uniform airow thru the fume chamber to exhaust collar outlet. UniFlow Radioisotope Fume Hood Cat. No. 14601, shown with optional accessories. Plumbing Fixtures are color coded to specic service. RADIOISOTOPE EXHAUST BLOWERS & FILTERS Radioisotope Exhaust Blowers HEMCO offers specialty exhaust blowers to safely and efficiently vent radioisotope fume hoods. Blowers are sized based on the hood size and length of ducting. Fume Hood Face Velocity cfm Sash in full open position should be for setup of apparatus & maintenance service only. If design opening is at 1/2 open at 100 fpm (feet per minute), face velocity at full open would be approximately 50 fpm. The recommended face velocity for efciency & safety should be 80-100 fpm. Lower face velocity may compromise user safety. Radioisotope Filters Bag-In/Bag-Out filtration systems to meet specific exhaust requirements. carbon filters are used to absorb fumes and HEPA filters are used to collect particulate. HEMCO Corporation 711 S. Powell Road Independence, MO 64056 Radioisotope Fume Hood Sash Management & Design Size hood
Open the catalog to page 1Radioisotope Fume Hoods UNIFLOW RADIOISOTOPE FUME HOOD DESCRIPTIONS 1. Uniow Radioisotope Fume Hood with Stainless Steel Interior Liner. Seamless type 304 stainless steel fume chamber with integral work surface all coved corners with bafe and exhaust collar. Picture frame sash opening with counterbalanced clear tempered safety glass sash with chemical resistant stainless steel framing and track and aerodynamic sash lift. Vapor-proof light xture and switch pre-wired to single point junction box, 115V/60Hz AC, All electrical components UL listed. Hood Depth FUME HOOD WIDTH A 48” 60” Cat.No....
Open the catalog to page 2All HEMCO Corporation catalogs and technical brochures
-
Universal Fume Hood
1 Pages
-
Casework
4 Pages
-
Canopy Hoods
2 Pages
-
Lab Planning Solutions 2015
18 Pages
-
UniFlow Fume Hoods
28 Pages
-
VSE Balance Enclosure
1 Pages
-
Vented Demo Hood
1 Pages
-
24910 Table Top
1 Pages
-
CMM Rooms
1 Pages
-
Modular Labs
2 Pages
-
24000 Table Top Hood
1 Pages
-
Trace Metals Fume Hood
2 Pages
-
SE Dual Entry Fume Hood
2 Pages
-
Perchloric Acid Fume Hood
2 Pages
-
UniFlow SE Fume Hood
2 Pages
-
Auxiliary Aire Fume Hood
2 Pages
-
UniFlow CE Fume Hood
2 Pages
-
UniMax Brochure 2013
4 Pages
-
Clean Rooms
7 Pages