Calibration-TDS X-RAY Calibration Standards

Calibration-TDS X-RAY Calibration Standards
1 / 40 PagesView full catalog

Calibration-TDS X-RAY Calibration Standards

Product catalog summary
Overview: The document provides comprehensive information on calibration standards and procedures by FISCHER for coating thickness and material analysis using Energy-dispersive X-ray fluorescence analysis (EDXRF). It emphasizes the significance of quality assurance in production and the necessity of reliable measurements for compliance with ISO-certified quality management systems.
Accreditation and Certification: FISCHER's Calibration Laboratory is accredited by DIN EN ISO/IEC 17025, enabling it to issue DAkkS calibration certificates for 'Mass per unit area' calibration standards, essential for calibrating X-ray fluorescence instruments.
Calibration Standards: The document details a portfolio of over 300 calibration standards for EDXRF analysis, applicable in industries like electronics, jewelry, and semiconductors. It also offers customer-specific calibration standards certified under DAkkS procedures.
Benefits of DAkkS Accreditation: DAkkS accreditation ensures regular verification of the calibration laboratory's expertise and provides traceability of measured results to internationally recognized base units, crucial for process and quality control.
Traceability and Measurement Procedures: Traceability of FISCHER calibration standards is achieved through accredited procedures and comparison measurements with other quantified standards. The document highlights the importance of using actual values for calibration data.
Calibration Certificate Details: The DAkkS calibration certificate includes details on the calibration object, measurement procedures, environmental conditions, and results, emphasizing the use of actual values over nominal values.
Conclusion: FISCHER offers a unique service of certifying customer products as calibration standards, enhancing production process security and compliance with quality management systems.
Ordering Process for Customer-specific Calibration Standards: The process involves sending an enquiry and material to FISCHER, testing for suitability, providing a cost estimate, and producing and dispatching the standards upon order confirmation.
Certification Options: Calibration standards come with a manufacturer's certificate, with some offering a DAkkS calibration certificate, indicated by a green check mark in the catalog.
Differences Between Certificates: Manufacturer's certificates do not include traceability or international recognition, while DAkkS certified standards do and are accepted by ISO 9000/9001 certification bodies worldwide.
Calibration Standards Details: The document lists various standards, including single coatings, double coatings, alloy coatings, pure elements, and foil standards, with DAkkS certification options.
X-RAY Accessories and Calibration Standards Catalog: The document also serves as a catalog for X-RAY accessories and calibration standards by Helmut Fischer GmbH, detailing positioning aids, measuring supports, and other accessories used in X-RAY measurement and analysis.
Positioning Aids and Measuring Supports: Includes tools like trap and support tables, measuring support tables, XUL positioning aids, and universal jaw vices for precise specimen placement.
Material Analysis Tools: Features tools like material analysis measuring cell sets and pressing tools for sample preparation.
Calibration Standards: Lists pure element plates, monitoring sets, and test samples for RoHS function checks.
Scattering and Reference Samples: Includes ABS and Al scattering samples and Ni/Cu reference samples for monitoring measurement devices.
Adjustment and Calibration Tools: Details adjustment plates and kits for specific XDV models and calibration holders for inline measurement systems.
Additional Accessories: Includes drift correction samples, solution analysis options, and measuring table inserts.
Global Presence: Helmut Fischer GmbH operates globally with subsidiaries in multiple countries, ensuring widespread availability of their products and services.
See more

Catalog excerpts

Calibration-TDS X-RAY Calibration Standards-1

Certified FISCHER X-RAY Calibration Standards and Accessories DIN ISO/IEC 17025 accredited March 2018/2 Q Coating Thickness [jjj] Material Analysis [5l Microhardness [§] Material Testing

 Open the catalog to page 1
Calibration-TDS X-RAY Calibration Standards-2

Quality Assurance with Energy-dispersive X-Ray Fluorescence Analysis (EDXRF Analysis) Quality assurance and optimum processing procedures play an ever greater role in production. From this stems the necessity to perform reliable measurements, which also comply with the demanding requirements of an ISO certified quality management system. Energy-dispersive X-ray fluorescence analysis (EDXRF analysis) is a particularly universal, fast and nondestructive measuring method. By using suitable reference materials, it enables coating thicknesses and material composition for a large number of applications...

 Open the catalog to page 2
Calibration-TDS X-RAY Calibration Standards-3

Traceability of Calibration Standards The traceability of FISCHER calibration standards is effected on the basis of the following procedures: The DAkkS Calibration Certificate . . . . . . . . . . . . . 4 In accordance with a procedure that is accredited by the German Accreditation Service, Deutscher Akkreditierungsservice (DKD/DAkkS), see the accreditation certificate. This procedure can be applied for many pure-element coatings. The FISCHER Calibration Laboratory ensures traceability to internationally recognized material measures. You will find the accreditation document on the FISCHER web...

 Open the catalog to page 3
Calibration-TDS X-RAY Calibration Standards-4

The DAkkS Calibration Certificate The Front Page of the DAkkS Calibration Certificate The insignia designates the DAkkS calibration certificate as an official document. Strict requirements apply to calibration laboratories, and these requirements are regularly checked by the accreditation body of the German Calibration Service, Deutscher Kalibrierdienst (DKD). akkreditiert durch die / accredited by the Deutsche Akkreditierungsstelle GmbH als Kalibrierlaboratorium im / as calibration laboratory in the Calibration certificate Calibration mark Gegenstand Object Hersteller Manufacturer Flächenmassenormal...

 Open the catalog to page 4
Calibration-TDS X-RAY Calibration Standards-5

The Back Page of the DAkkS Calibration Certificate Description of the calibration object The traceability of the calibration standards is effected via a comparison measurement by means of Xray fluorescence analysis against gravimetrically measured reference standards. Identification of the calibration standard (as it is also printed on the standard), material and the mass per unit area that is determined. The mass per unit area is the size, which an X-ray fluorescence instrument determines. The coating thickness is derived from the mass per unit area via the material density. In the case of coatings:...

 Open the catalog to page 5
Calibration-TDS X-RAY Calibration Standards-6

Customer-specific Calibration Standards FISCHER offers you the unique service of being able to certify your own product as a calibration standard. In many cases it is even possible to issue an internationally recognized DAkkS calibration certificate. You can show this to your customers and populate it to an existing quality management system in accordance with ISO 9000/1. You also gain security with this in your production process, if a calibration standard corresponds to your products. The production of traceable calibration standards is very time-consuming and also dependent on the quality...

 Open the catalog to page 6
Calibration-TDS X-RAY Calibration Standards-7

DAkkS Column All calibration standards will be delivered with a manufacturer‘s certificate. If the table starting on Page 8 displays a green check mark  , then you do have the option to order a DAkkS calibration certificate for this standard. If the table row doesn‘t show  , then the standard is available with manufacturer‘s certificate only. Differences between Standards with a Manufacturer‘s Certificate and DAkkS Certified Standards Produced by a DAkkS certified calibration laboratory Standard with manufacturer‘s certificate DAkkS certified standard Produced in accordance with DAkkS certified...

 Open the catalog to page 7
Calibration-TDS X-RAY Calibration Standards-8

X-RAY Calibration Standards Catalog Solid Standards Coatings KAL-N 1: Solid Standards Single Coatings Coating structure Base material

 Open the catalog to page 8
Calibration-TDS X-RAY Calibration Standards-10

Part No. Designation DAkkS 604-635 KAL-N dd Ag/Ni/Alloy42 1.0/1.0 |jm 604-636 KAL-N dd Ag/Ni/Alloy42 1.0/3.0 jm 604-637 KAL-N dd Ag/Ni/Alloy42 5.0/1.0 jm 604-638 KAL-N dd Ag/Ni/Alloy42 5.0/3.0 jm 604-619 KAL-N dd Ag/Ni/Cu 1.0/1.0 jm 604-620 KAL-N dd Ag/Ni/Cu 1.0/3.0 jm 604-116 KAL-N dd Ag/Ni/Cu 1.0/15 jm 604-621 KAL-N dd Ag/Ni/Cu 5.0/1.0 jm 604-622 KAL-N dd Ag/Ni/Cu 5.0/3.0 jm 604-651 KAL-N dd Ag/Ni/CuZn30 1.0/1.0 jm 604-652 KAL-N dd Ag/Ni/CuZn30 1.0/3.0 jm 604-653 KAL-N dd Ag/Ni/CuZn30 5.0/1.0 jm 604-654 KAL-N dd Ag/Ni/CuZn30 5.0/3.0 jm 604-629 KAL-N dd Au/Ni/Alloy42 0.25/1.0 jm Part No....

 Open the catalog to page 10
Calibration-TDS X-RAY Calibration Standards-11

Part No. Designation DAkkS 601-895 KAL-N dd Au/Ni/Cu 0.8/2.5 jm 604-615 KAL-N dd Au/Ni/Cu 1.0/1.0 jm 604-616 KAL-N dd Au/Ni/Cu 1.0/3.0 jm 602-087 KAL-N dd Au/Ni/Cu 1.5/10 jm 604-617 KAL-N dd Au/Ni/Cu 2.5/1.0 jm 604-618 KAL-N dd Au/Ni/Cu 2.5/3.0 jm 602-086 KAL-N dd Au/Ni/Cu 2.5/5.0 jm 604-645 KAL-N dd Au/Ni/CuZn30 0.25/1.0 jm 604-646 KAL-N dd Au/Ni/CuZn30 0.25/3.0 jm 604-647 KAL-N dd Au/Ni/CuZn30 1.0/1.0 jm 604-648 KAL-N dd Au/Ni/CuZn30 1.0/3.0 jm 604-649 KAL-N dd Au/Ni/CuZn30 2.5/1.0 jm 604-650 KAL-N dd Au/Ni/CuZn30 2.5/3.0 jm 602-996 KAL-N dd Au/Pd/Ni 0.5/0.8 jm Part No. Designation DAkkS...

 Open the catalog to page 11
Calibration-TDS X-RAY Calibration Standards-12

2.5 pm 602-070 KAL-N dc Sn60Pb40/Ni 11 pm 602-071 KAL-N dc Sn60Pb40/Ni 40 pm 602-072 KAL-N dc Sn90Pb10/Ni 2.0 pm 602-073 KAL-N dc Sn90Pb10/Ni 10.5 pm 12 X-RAY Calibration Standards

 Open the catalog to page 12
Calibration-TDS X-RAY Calibration Standards-13

KAL-N: Solid Standards Pure Elements Part No. Designation DAkkS Bulk KAL-N: Solid Standards Ultra-High Purity Elements1 Delivery with certificate of analysis

 Open the catalog to page 13
Calibration-TDS X-RAY Calibration Standards-14

Solid Standards Bulk Part No. Designation DAkkS Part No. Designation DAkkS Continued on the next page The standard measurement uncertainty is determined for each calibration standard individually. For gold and silver standards the standard measurement uncertainty of the calibration standards is always below 0.25 %. 14 X-RAY Calibration Standards

 Open the catalog to page 14
Calibration-TDS X-RAY Calibration Standards-15

KAL-N: Solid Standards Alloys (Bulk) - Continuation Part No. Designation DAkkS 601-849 KAL-N c Kovar2 604-036 KAL-N c Pt950Cu50 604-037 KAL-N c Pt960Ru40 603-896 KAL-N c RoHS PE Low: Br 500 ppm, Hg 200 ppm, Cr 398 ppm, Pb 400 ppm, Cd 100 ppm 603-895 KAL-N c RoHS PE High: Br 1099 ppm, Hg 1100 ppm, Cr 1000 ppm, Pb 1201 ppm, Cd 300 ppm 603-988 KAL-N c RoHS PVC Blank 603-987 KAL-N c RoHS PVC-Low: Br 499 ppm, Hg 200 ppm, Cr 400 ppm, Pb 399 ppm, Cd 100 ppm 603-989 KAL-N c RoHS PVC High: Br 1099 ppm, Hg 1100 ppm, Cr 1000 ppm, Pb 1199 ppm, Cd 301 ppm 604-268 KAL-N c Sn62Pb36Ag2 The standard measurement...

 Open the catalog to page 15

All HELMUT FISCHER GMBH - INSTITUT FÜR ELEKTRONIK UND catalogs and technical brochures

  1. MMS-PC2

    16  Pages

  2. XAN250-252

    4  Pages

  3. XAN220-222

    4  Pages

  4. XAN Series

    36  Pages

  5. PMP10 Duplex

    4  Pages

  6. FMP100/150

    8  Pages

  7. Probe catalogue

    12  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.