Product catalog summary
Overview: The LGT-106 is a flameproof bypass type tunable laser gas analyzer designed for industrial on-line analysis and environmental monitoring. It is particularly suitable for challenging conditions such as high temperature, high humidity, high dust, high pressure, and small pipes where in-situ analyzers are not applicable.
System Composition: The system consists of a transmitter, receiver, and gas cell. The transmitter emits laser wavelengths that pass through the gas cell to the receiver, which processes the signal to determine gas concentration. The system includes an auto purge function to reduce maintenance costs.
Features:
  • Suitable for harsh conditions.
  • Integrated structure for high reliability and stability.
  • Utilizes "Single line spectrum" technology to avoid interference from background gases.
  • Innovative Hall button design for easy operation.
  • Automatic calibration and backblow functions.
  • Uses imported TO encapsulate semiconductor lasers with antireflective coating to reduce optical noise.
Technical Specifications:
  • Linearity Error: ≤±1%F.S.
  • Repeatability: ≤1%
  • Span Drift: ≤±1%F.S./half year
  • Zero Drift: ≤±1%F.S./half year
  • Calibration Cycle: ≤2 times/year
  • Protection Type: ExdIICT6 Gb
  • Enclosure Rating: IP66
  • Preheating Time: ≤15 minutes
  • Response Time (T90): ≤30s
Interface and Operating Conditions:
  • Analog Output: 2× 4-20mA output
  • Analog Input: 2× 4-20mA input for temperature and pressure compensation
  • Digital Output: RS485/RS232/GPRS
  • Relay Output: 3 (24V, 1A)
  • Power Supply: 24VDC, <20W
  • Purge Gas: 0.3MPa~0.8MPa industrial nitrogen or purified air
  • Ambient Temperature: -20℃~+60℃
Installation: The analyzer is installed using a by-pass type installation. It supports explosion-proof probes for zone 1 and zone 2, and automatic backblow for dirty sample gases. The system can be controlled internally without additional PLCs.
See more

Catalog excerpts

zetian/ By-pass laser gas analyzer/TDLAS/extracive gas analyzer/industrial process/LGT-106/iron making, steel making, coking, refining, petrochemical, flue gas, cement, thermal power-1

LGT-106 Laser Gas Analyzer LGT-106 is a flameproof bypass type tunable laser gas analyzer for industrial on-line analysis and environmental on-line monitoring, especially suitable for harsh working conditions such as high temperature, high humidity, high dust, high pressure, and for small pipes where in-situ analyzer is not applicable. The transmitter and receiver are installed on a back board. The analyzer contains auto purge function which greatly reduces maintenance cost. System Composition LGT-106 laser gas analyzer consists of transmitter, receiver and gas cell. The transmitter drives the tunable diode to emit laser of certain wavelengths, which passes through the gas cell, then reaches the receiver. The receiver performs signal processing to obtain second harmonic signal, then calculates the concentration according to the relationship between the second harmonic signal and the gas concentration. Receive unit Transmit unit Signal input and output 24V input Measuring gas chamber Gas Path Extracted by a jet pump, the sample gas is filtered(get rid of dust), heated(avoid condensation), cooled(for high temperature condition), then reaches analyze module to be measured. Shutting off ball valve and analyze module , the analyzer carries out a calibration to ensure accuracy. Shutting off ball valve and opening back-flush valve ensures gas path unblocked. Configurations of LGT-106 is different in different situations. For example, for high pressure situation, a relief valve is added and air pump is removed. For high temperature situation, the heat tracing temperature is higher. Counter blowing Probe Exhaust outlet Jet Trace tube, 1 - 2meter long Ball valve

 Open the catalog to page 1
zetian/ By-pass laser gas analyzer/TDLAS/extracive gas analyzer/industrial process/LGT-106/iron making, steel making, coking, refining, petrochemical, flue gas, cement, thermal power-2

• Suitable for harsh working conditions such as high temperature, high humidity, high dust, high pressure, and for small pipes where are not suitable for in situ measurement occasions. • Integrated structure design, high reliability and stability. • "Single line spectrum" technology, free from interference of background gas. • Innovative Hall button design, high sensitivity, easy operation. • Automatic calibration, automatic backblow. • Adopts imported TO encapsulate semiconductor lasers, optical noises are restrained with antireflective coating end cap. Technical Indicators Linearity Error S...

 Open the catalog to page 2

All Hangzhou Chunlai Technology Co., Ltd. catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.