Product Specifications
  • Sensor Family: VisiTrace RS485
  • Parameter: Dissolved Oxygen (Optical)
  • Output: Modbus, 4-20 mA (via Arc Wi 2G Adapter)
  • Length: 120 mm
  • Electrical Connector: VP8
  • Measurement Principle: Oxygen-dependent luminescence quenching, temperature compensation
  • Measuring Range: 0 to 2 ppm (0 to 50 mbar)
  • Accuracy at 25°C: ±0.5 ppb or 2%, whichever is greater
  • Drift at Room Temperature: <1% per week @ 100 ppb
  • Sensor Cap: ODO Cap L1
  • Electrolyte: None
  • Oxygen Consumption: None
  • Temperature Sensor: NTC 22 kOhm
  • Configurable Values: DO: mbar; %-sat; %-vol; µg/l; mg/l; ppb/ppm; Temperature: °C, °F, K
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Stainless Steel 1.4435, EPDM, Silicone (FDA 21 CFR 177.2600), VMQ
  • Surface Quality of Steel: Ra < 0.4 µm (N5)
  • Digital Interface: RS485 Modbus (max. 32 addresses)
  • Baud Rate: 4800, 9600, 19200, 38400, 57600, 115200 Bd
  • Operating Voltage: 10 to 27 VDC max. 1.5 W
  • Serial Number: Yes
  • Response Time: t90% < 20 s in gas; < 90 s in water
  • Certificate: Yes
  • ATEX Approval: No
  • IECEx Approval: No
  • Autoclavable: Yes
  • CIP: Yes
  • Steam Sterilizable: Yes, max. Temperature 140 °C
  • Operating Temperature Range: -20 to 140 °C; no DO reading above 85 °C
  • Pressure Range: -1 to 12 bar
  • Protection Rating: IP 68
  • Required Flow: None

Specifications are subject to change without notice. For more details, visit the Hamilton Company website.

See more

Catalog excerpts

VisiTrace RS485 120 L1-1

VisiTrace RS485 120 L1 Specification Sheet (Part/REF # 10140043-1111) TEST SHORT Product Specifications Sensor Family Sensor Output Arc: Modbus, 4-20 mA (only via Arc Wi 2G Adapater) Electrical Connector Measurement Principle Oxygen dependent luminescence quenching, temperature compensation Measuring Range Drift at Room Temperature Oxygen Consumption Temperature Sensor Configurable Values DO: mbar; %-sat; %-vol; µg/l; mg/l; ppb/ppm (dissolved oxygen); Temperature: °C, °F, K Process Connection Wetted Parts Stainless Steel 1.4435 EPDM (Ethylene propylene elastomer) Silicone - FDA 21 CFR 177.2600 VMQ (Silicone elastomer) See compliance details in Material Specification document Surface Quality of Steel Digital Interface Baud Rate Operating Voltage Serial Number Response Time ATEX Approval IECEx Approval

 Open the catalog to page 1
VisiTrace RS485 120 L1-2

Steam Sterilizable Operating Temperature Range -20 to 140 °C; the sensor provides no DO reading above 85 °C Protection Rating Required Flow Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/243666-111-2

 Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold G 120

    2  Pages

  2. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.