VisiTrace mA 120 L1

VisiTrace mA 120 L1

VisiTrace mA 120 L1

Product catalog summary
Product Overview
The VisiTrace mA optical oxygen sensor is designed for measuring low levels of dissolved oxygen down to parts per billion (ppb) levels. It is suitable for applications requiring precise oxygen measurement.

Specifications
  • Sensor Family: VisiTrace mA
  • Parameter: Dissolved Oxygen (DO) - Optical
  • Sensor Output: 4 to 20 mA / HART
  • Length: 120 mm
  • Electrical Connector: M12
  • Measurement Principle: Oxygen-dependent luminescence quenching with temperature compensation
  • Measuring Range: 0 to 2 ppm (0 to 50 mbar)
  • Accuracy at 25°C: ±0.5 ppb or 2%, whichever is greater
  • Drift at Room Temperature: <1% per week @ 100 ppb
  • Sensor Cap: ODO Cap L1
  • Electrolyte: None
  • Oxygen Consumption: None
  • Temperature Sensor: NTC 22 kOhm
  • Configurable Values: DO in various units; Temperature in °C, °F, K
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Stainless Steel 1.4435, EPDM, Silicone (FDA compliant), VMQ
  • Surface Quality of Steel: Ra < 0.4 µm (N5)
  • 4 to 20 mA Current Range: 3.5 to 22 mA
  • Analog Interface: Two-wire sink 4 to 20 mA for DO, programmable
  • 4 to 20 mA Accuracy: <0.3% + 0.05 mA
  • Transmitting Power: Max. 4 dBm
  • Digital Interface: Bluetooth 5.0, HART Version 7.0
  • Operating Voltage: 18 to 30 VDC
  • Serial Number: Yes
  • Response Time: t90% < 20 s in gas; < 90 s in water
  • Certificate: Yes, with measured current in oxygen

Compliance and Approvals
  • ATEX Approval: II 1 G Ex ia IIC T3/T4/T5/T6 Ga; II 1 D Ex ia IIIC T135°C Da
  • IECEx Approval: II 1 G Ex ia IIC T3/T4/T5/T6 Ga; II 1 D Ex ia IIIC T135°C Da

Operational Features
  • Autoclavable: Yes
  • CIP: Yes
  • Steam Sterilizable: Yes, max. Temperature 140°C
  • Operating Temperature Range: -10 to 140°C; no DO reading above 85°C
  • Pressure Range: -1 to 12 bar
  • Protection Rating: IP 68
  • Required Flow: None

Note: Specifications are subject to change without notice. For more details, visit the Hamilton Company website.
See more

Catalog excerpts

VisiTrace mA 120 L1-1

VisiTrace mA 120 L1 Specification Sheet (Part/REF # 10068709-1111) The VisiTrace optical oxygen sensor is specifically designed for low level dissolved oxygen down to ppb levels. Product Specifications Sensor Family Sensor Output Electrical Connector Measurement Principle Oxygen dependent luminescence quenching, temperature compensation Measuring Range Drift at Room Temperature Oxygen Consumption Temperature Sensor Configurable Values DO: mbar; %-sat; %-vol; µg/l; mg/l; ppb/ppm (gas); ppb/ppm (dissolved oxygen); Temperature: °C, °F, K Process Connection Wetted Parts Stainless Steel 1.4435 EPDM (Ethylene propylene elastomer) Silicone - FDA 21 CFR 177.2600 VMQ (Silicone elastomer) See compliance details in Material Specification document Surface Quality of Steel Two wire sink 4 to 20 mA for DO, programmable Transmitting Power Digital Interface Bluetooth 5.0 HART Version 7.0 Operating Voltage Serial Number Response Time Yes, with measured current in oxy

 Open the catalog to page 1
VisiTrace mA 120 L1-2

ATEX Approval IECEx Approval Steam Sterilizable Operating Temperature Range -10 to 140 °C; the sensor provides no DO reading above 85 °C Protection Rating Required Flow Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/10068709-1111 Spec. Version

 Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold G 120

    2  Pages

  2. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.