  • Sensor Family: VisiFerm RS485
  • Parameter: Dissolved Oxygen (DO) - Optical
  • Sensor Output: Arc: Modbus, 4-20 mA (via Arc Wi 2G Adapter), ECS
  • Length: 120 mm
  • Electrical Connector: VP8
  • Measurement Principle: Oxygen-dependent luminescence quenching, temperature compensation
  • Measuring Range: 4 ppb to 25 ppm (DO), 0 to 62.85 %-vol, or 0 to 300 %-sat
  • Accuracy at 25°C: 1 ± 0.05 %-vol; 21 ± 0.2 %-vol; 50 ± 0.5 %-vol
  • Drift at Room Temperature: < 1 % per week
  • Sensor Cap: ODO Cap H2
  • Electrolyte: None
  • Oxygen Consumption: None
  • Temperature Sensor: NTC 22 kOhm
  • Configurable Values: DO: mbar, %-sat, %-vol, µg/l, mg/l, ppb/ppm; Temperature: °C, °F, K
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Stainless Steel 1.4435, PTFE, EPDM
  • Surface Quality of Steel: Ra < 0.4 µm (N5)
  • Analog Interface: ECS
  • Digital Interface: RS485 Modbus (max. 32 addresses)
  • Baud Rate: 4800, 9600, 19200, 38400, 57600, 115200 Bd
  • Operating Voltage: 10 to 27 VDC max. 1.5 W
  • Serial Number: Yes
  • Response Time: t98% < 60 s at 25°C
  • Certificate: Yes
  • ATEX Approval: No
  • IECEx Approval: No
  • Autoclavable: Yes
  • CIP: Yes
  • Steam Sterilizable: Yes, max. Temperature 140°C
  • Operating Temperature Range: -20 to 140°C; no DO reading above 85°C
  • Pressure Range: -1 to 12 bar
  • Protection Rating: IP 68
  • Required Flow: None

Specifications are subject to change without notice. For more details, visit the Hamilton Company website.

See more

Catalog excerpts

VisiFerm RS485-ECS 120 H2-1

VisiFerm RS485-ECS 120 H2 Specification Sheet (Part/REF # 10118255-1121) TEST SHORT Product Specifications Sensor Family Sensor Output Arc: Modbus, 4-20 mA (only via Arc Wi 2G Adapter), ECS Electrical Connector Measurement Principle Oxygen dependent luminescence quenching, temperature compensation Measuring Range Drift at Room Temperature Oxygen Consumption Temperature Sensor Configurable Values DO: mbar; %-sat; %-vol; µg/l; mg/l; ppb/ppm (dissolved oxygen); Temperature: °C, °F, K Process Connection Wetted Parts Stainless Steel 1.4435 PTFE (Polytetrafluoroethylene) EPDM (Ethylene propylene elastomer) See compliance details in Material Specification document Surface Quality of Steel Analog Interface Digital Interface Baud Rate Operating Voltage Serial Number Response Time ATEX Approval IECEx Approval

 Open the catalog to page 1
VisiFerm RS485-ECS 120 H2-2

Steam Sterilizable Operating Temperature Range -20 to 140 °C; the sensor provides no DO reading above 85 °C Protection Rating Required Flow Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/243666-111-3-3

 Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold G 120

    2  Pages

  2. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.