VisiFerm mA 120 H4

VisiFerm mA 120 H4

VisiFerm mA 120 H4

Product catalog summary
Product Overview
The VisiFerm mA 120 H4 is designed for GMP production environments and explosive-rated areas, offering less frequent calibration and longer lifetime. It features a 2-wire 4-20mA/HART interface.

Specifications
  • Sensor Family: VisiFerm mA
  • Length: 120 mm
  • Electrical Connector: M12
  • Measurement Principle: Oxygen-dependent luminescence quenching with temperature compensation
  • Measuring Range: 4 ppb to 25 ppm (DO), 0 to 62.85 %-vol, or 0 to 300 %-sat
  • Accuracy at 25°C: 1 ± 0.05 %-vol; 21 ± 0.2 %-vol; 50 ± 0.5 %-vol
  • Drift at Room Temperature: < 1 % per week
  • Sensor Cap: ODO Cap H4
  • Electrolyte: None
  • Oxygen Consumption: None
  • Temperature Sensor: NTC 22 kOhm
  • Configurable Values: DO in various units; Temperature in °C
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Stainless Steel 1.4435, PTFE, EPDM
  • Surface Quality of Steel: Ra < 0.4 µm (N5)
  • 4 to 20 mA Current Range: 3.5 to 22 mA
  • Analog Interface: Two-wire sink 4 to 20 mA for DO, programmable
  • 4 to 20 mA Accuracy: < 0.3 % current value + 0.05 mA
  • Transmitting Power: Max. 4 dBm
  • Digital Interface: Bluetooth 5.0, HART Version 7.0
  • Operating Voltage: 18 to 30 VDC
  • Serial Number: Yes
  • Response Time: t98% < 60 s at 25°C
  • Certificate: Yes, with measured current in air
  • ATEX Approval: II 1 G Ex ia IIC T3/T4/T5/T6 Ga; II 1 D Ex ia IIIC T135°C Da
  • IECEx Approval: II 1 G Ex ia IIC T3/T4/T5/T6 Ga; II 1 D Ex ia IIIC T135°C Da
  • Autoclavable: Yes
  • CIP: Yes
  • Steam Sterilizable: Yes, max. Temperature 140°C
  • Operating Temperature Range: -20 to 140°C; no DO reading above 85°C
  • Pressure Range: -1 to 12 bar
  • Protection Rating: IP 68
  • Required Flow: None

Additional Information
Specifications are subject to change without notice. For more details, visit the Hamilton Company website.
See more

Catalog excerpts

VisiFerm mA 120 H4-1

VisiFerm mA 120 H4 Specification Sheet (Part/REF # 10070760-1121) Less frequent calibration and longer lifetime. 2 wire 4-20mA / HART for GMP production environments. Explosive rated environments. Product Specifications Sensor Family Electrical Connector Measurement Principle Oxygen dependent luminescence quenching, temperature compensation Measuring Range Drift at Room Temperature Oxygen Consumption Temperature Sensor Configurable Values DO: mbar; %-sat; %-vol; µg/l; mg/l; ppb/ppm (gas); ppb/ppm (dissolved oxygen); Temperature: °C Process Connection Wetted Parts Stainless Steel 1.4435 PTFE (Polytetrafluoroethylene) EPDM (Ethylene propylene elastomer) See compliance details in Material Specification document Surface Quality of Steel Two wire sink 4 to 20 mA for DO, programmable Transmitting Power Digital Interface Bluetooth 5.0 HART Version 7.0 Operating Voltage Serial Number Response Time Yes, with measured current in air ATEX Approval IECEx Approval

 Open the catalog to page 1
VisiFerm mA 120 H4-2

Steam Sterilizable Operating Temperature Range -20 to 140 °C; the sensor provides no DO reading above 85 °C Protection Rating Required Flow Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/10070760-1121

 Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold G 120

    2  Pages

  2. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.