1. Catalogs
  2. Hamilton Process Analytics
  3. Polilyte Plus ORP Arc 120

Polilyte Plus ORP Arc 120

Polilyte Plus ORP Arc 120

Polilyte Plus ORP Arc 120

Product catalog summary
Overview: The Polilyte Plus ORP Arc 120 sensor combines the benefits of the Polilyte Plus family with polymer reference technology to prevent clogging and poisoning. It features Arc technology for direct communication with Process Control Systems.
Product Specifications:
  • Sensor Family: Polilyte Plus
  • Parameter: ORP
  • Sensor Output: Arc: Modbus, 4 to 20 mA
  • Length: 120 mm
  • Electrical Connector: VP8
  • Measurement Principle: Combination electrode; ORP potential measured against reference
  • Measuring Range: -1500 mV to +1500 mV
  • ORP Element: Pt wire
  • Membrane Shape: Spherical
  • Diaphragm: Single Pore
  • Number of Diaphragms: 2
  • Electrolyte: Polisolve Plus
  • Reference System: Hamilton Everef-L
  • Liquid Earth: Yes
  • Temperature Sensor: NTC 22 kOhm
  • Configurable Values: ORP: mV; Temperature: °C, K, °F
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Platinum, Glass, FKM (Fluorocarbon Elastomer)
  • Analog Interface 1: 4 to 20 mA, programmable
  • Analog Interface 2: 4 to 20 mA for Temp., programmable
  • Analog Interface 1 and 2: Galvanically not isolated; pulse width modulation 3.5 kHz
  • Digital Interface: RS485 Modbus (max. 31 addresses)
  • Baud Rate: 4800, 9600, 19200, 38400, 57600, 115200 Bd
  • Operating Voltage: 7 to 30 VDC max. 150 mW
  • Serial Number: Yes
  • Recommended Storage: Hamilton Storage Solution or 3 M KCl
  • Certificate: Yes, with measured value in ORP buffer +475 mV
Additional Specifications:
  • ATEX Approval: No
  • IECEx Approval: No
  • Autoclavable: Yes, max. Temperature 130 °C
  • CIP: No
  • Steam Sterilizable: Yes, max. Temperature 140 °C
  • Operating Temperature Range: Analog interface: 0 to 110 °C; Digital interface: 0 to 140 °C
  • Sample min Conductivity: 2 µS/cm
  • Pressure Range: 0 to 16 bar (100 °C), 0 to 10 bar (130 °C), 0 to 3 bar (140 °C)
See more

Catalog excerpts

Polilyte Plus ORP Arc 120-1

Polilyte Plus ORP Arc 120 Specification Sheet (Part/REF # 243060) All the advantages of the Polilyte Plus family with polymer reference to eliminate clogging and poisoning are included. Arc technology allows direct communication with the Process Control System. Product Specifications Sensor Family Polilyte Plus Sensor Output Electrical Connector Measurement Principle Combination electrode; ORP potential measured against reference Measuring Range Membrane Shape Single Pore Polisolve Plus Reference System Hamilton Everef-L Liquid Earth Temperature Sensor Configurable Values Process Connection Wetted Parts Platinum Glass FKM (Fluorocarbon Elastomer) Galvanically not isolated; pulse width modulation 3.5 kHz Digital Interface Baud Rate Operating Voltage Serial Number Recommended Storage Hamilton Storage Solution or 3 M KCl Yes, with measured value in ORP buffer +475 mV

 Open the catalog to page 1
Polilyte Plus ORP Arc 120-2

ATEX Approval IECEx Approval Steam Sterilizable Operating Temperature Range Analog interface: 0 to 110 °C; Digital interface: 0 to 140 °C Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/243060 Spec. Vers

 Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold G 120

    2  Pages

  2. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.