Product Specifications:
  • Sensor Family: Polilyte Plus
  • Parameter: pH
  • Sensor Output: Traditional mV
  • Length: 120 mm
  • Electrical Connector: S8
  • Measurement Principle: Combination electrode; pH potential measured against reference
  • Measuring Range: pH 0 to 14
  • Sensitivity: 57 to 59 mV/pH at 25 °C
  • Zero Point: 0 ± 20 mV
  • pH Glass: Hamilton Type HF Glass, suitable for specific HF concentrations
  • pH Glass Resistance: Moderate
  • Membrane Shape: Spherical
  • Diaphragm: Single Pore, 2 diaphragms
  • Electrolyte: Polisolve Plus
  • Reference System: Hamilton Everef-L
  • Liquid Earth: No
  • Temperature Sensor: No
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Glass, FKM (Fluorocarbon Elastomer)
  • Serial Number: Yes
  • Recommended Storage: Hamilton Storage Solution or 3 M KCl
  • Certificate: Yes, with measured values in Buffer pH 4.01 and pH 7.00
  • ATEX Approval: II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
  • IECEx Approval: II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
  • Autoclavable: No
  • CIP: No
  • Steam Sterilizable: No
Operating Conditions:
  • Operating Temperature Range: -10 to 100 °C
  • Sample Minimum Conductivity: 2 µS/cm
  • Pressure Range: 0 to 16 bar (at 100 °C)
Additional Information: Specifications are subject to change without notice. For more details, visit the Hamilton Company website.
See more

Catalog excerpts

Polilyte Plus HF S8 120-1

Polilyte Plus HF S8 120 Specification Sheet (Part/REF # 242428-3213) Polilyte Plus Traditional pH sensor offer multiple electrical connectors and length options along with a standard mV output compatible with most transmitters and controllers. Product Specifications Sensor Family Polilyte Plus Sensor Output Electrical Connector Measurement Principle Combination electrode; pH potential measured against reference Measuring Range Zero Point Hamilton Type HF Glass: For < 0.01M / 200 mg/l HF at 20 °C or 0.05M / 1000mg/l HF at 50 °C Membrane Shape Single Pore Polisolve Plus Reference System Hamilton Everef-L Liquid Earth Temperature Sensor Process Connection Wetted Parts Glass FKM (Fluorocarbon Elastomer) Serial Number Recommended Storage Hamilton Storage Solution or 3 M KCl Yes, with measured values in Buffer pH 4.01 and pH 7.00 ATEX Approval II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db IECEx Approval II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db Steam Sterilizable

 Open the catalog to page 1
Polilyte Plus HF S8 120-2

Operating Temperature Range Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/242428-3213

 Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold G 120

    2  Pages

  2. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.