Polilyte Plus H MS 120

Polilyte Plus H MS 120

Polilyte Plus H MS 120

Product catalog summary
Product Overview
The Polilyte Plus Memosens pH sensors are designed for high accuracy and longevity in industrial applications, offering an upgrade for Memosens pH sensor installations.
Product Specifications
  • Sensor Family: Polilyte Plus
  • Parameter: pH
  • Sensor Output: Memosens
  • Length: 120 mm
  • Electrical Connector: Memosens
  • Measurement Principle: Combination electrode; pH potential measured against reference
  • Measuring Range: pH 0 to 14
  • Sensitivity: 57 to 59 mV/pH at 25 °C
  • Zero Point: 0 ± 20 mV
  • pH Glass: Hamilton Type H Glass
  • pH Glass Resistance: Moderate
  • Membrane Shape: Spherical
  • Diaphragm: Single Pore
  • Number of Diaphragms: 2
  • Electrolyte: Polisolve Plus
  • Reference System: Hamilton Everef-L
  • Liquid Earth: No
  • Temperature Sensor: NTC 30 kOhm
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Glass, FKM (Fluorocarbon Elastomer)
  • Serial Number: Yes
  • Recommended Storage: Hamilton Storage Solution or 3 M KCl
  • Certificate: Yes, with measured values in Buffer pH 4.01 and pH 7.00
  • ATEX Approval: II 1 G Ex ia IIC T3/T4/T6 Ga
  • IECEx Approval: II 1 G Ex ia IIC T3/T4/T6 Ga
  • Autoclavable: Yes, max. Temperature 130 °C
  • CIP: No
  • Steam Sterilizable: Yes, max. Temperature 130 °C
Operating Conditions
  • Operating Temperature Range: 0 to 130 °C
  • Sample Minimum Conductivity: 2 µS/cm
  • Pressure Range: 0 to 16 bar (100 °C), 0 to 10 bar (130 °C)
Additional Information
Specifications are subject to change without notice. For more details, visit the Hamilton Company website.
See more

Catalog excerpts

Polilyte Plus H MS 120-1

Polilyte Plus H MS 120 Specification Sheet (Part/REF # 242428-1413) Looking to upgrade your Memosens pH sensor installation? Polilyte Plus Memosens pH sensors offer high accuracy and longevity in industrial applications. Product Specifications Sensor Family Polilyte Plus Sensor Output Electrical Connector Measurement Principle Combination electrode; pH potential measured against reference Measuring Range Zero Point Hamilton Type H Glass Membrane Shape Single Pore Polisolve Plus Reference System Hamilton Everef-L Liquid Earth Temperature Sensor Process Connection Wetted Parts Glass FKM (Fluorocarbon Elastomer) Serial Number Recommended Storage Hamilton Storage Solution or 3 M KCl Yes, with measured values in Buffer pH 4.01 and pH 7.00 ATEX Approval IECEx Approval Steam Sterilizable

 Open the catalog to page 1
Polilyte Plus H MS 120-2

Operating Temperature Range Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/242428-1413 Spec. Version

 Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold G 120

    2  Pages

  2. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.