OxyGold G 120

OxyGold G 120

OxyGold G 120

Product catalog summary
Product Overview: The OxyGold G is a traditional polarographic dissolved oxygen (DO) sensor designed for trace oxygen measurement in parts-per-billion with a nanoampere (nA) output signal. It is compatible with most transmitters and controllers available in the market.
Specifications:
  • Sensor Family: OxyGold G
  • Parameter: DO (Electrochemical)
  • Sensor Output: Traditional: nA
  • Length: 120 mm
  • Electrical Connector: VP6
  • Measurement Principle: Electrochemical reduction of oxygen
  • Measuring Range: 1 ppb to 40 ppm (DO) or 0.02 to 1000 mbar (pO₂)
  • Drift at Room Temperature: < 1% per week
  • Membrane Cap: Optiflow
  • Electrolyte: Oxylyte G
  • Oxygen Consumption: Approximately 100 ng/h in air at 25 °C
  • Temperature Sensor: NTC 22 kOhm
  • Current in Air at 25 °C: 180 to 500 nA
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Stainless Steel 1.4435, EPDM, VMQ, Silicone
  • Surface Quality of Steel: Ra < 0.4 µm (N5)
  • Electrode System: Silver platinum combination
  • Polarization Voltage and Time: -670 ± 50 mV, ≥ 2 hours
  • Serial Number: Yes
  • Response Time t90%: < 60 s at 25 °C
  • Certificate: Material and Performance Test Certificate
  • ATEX Approval: II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
  • IECEx Approval: II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
  • Autoclavable: Yes
  • CIP: Yes
  • Steam Sterilizable: Yes
  • Operating Temperature Range: 0 to 130 °C
  • Pressure Range: 0 to 12 bar
  • Max. CO₂ Partial Pressure: 0.01 bar
  • Required Flow: ≥ 0.1 m/s
Compliance and Certifications: The sensor complies with ATEX and IECEx standards, ensuring safety in explosive environments. It is also autoclavable, CIP, and steam sterilizable, making it suitable for various industrial applications.
Additional Information: Specifications are subject to change without notice. For more details, visit the Hamilton Company website.
See more

Catalog excerpts

OxyGold G 120-1

OxyGold G 120 Specification Sheet (Part/REF # 237395) OxyGold G traditional polarographic DO sensors offer parts-per-billion trace oxygen measurement with a nA output signal compatible with most transmitters and controllers on the market. Product Specifications Sensor Family Sensor Output Electrical Connector Measurement Principle Electrochemical reduction of oxygen Measuring Range Drift at Room Temperature Oxygen Consumption Temperature Sensor Process Connection Wetted Parts Stainless Steel 1.4435 EPDM (Ethylene propylene elastomer) VMQ (Silicone elastomer) Silicone See compliance details in Material Specification document Surface Quality of Steel Electrode system Silver platinum combination Polarization Voltage and Time Serial Number Material and Performance Test Certificate ATEX Approval II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db IECEx Approval II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db Steam Steriliza

 Open the catalog to page 1
OxyGold G 120-2

Operating Temperature Range Required Flow Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/237395 Spec. Version

 Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.