OxyFerm FDA Arc 120

OxyFerm FDA Arc 120

OxyFerm FDA Arc 120

Product catalog summary
Overview
The OxyFerm FDA Arc 120 is a sensor that combines hygienic design with Arc technology for both analog and digital communication. It is designed for measuring dissolved oxygen (DO) using electrochemical methods.

Product Specifications
  • Sensor Family: OxyFerm FDA
  • Parameter: Dissolved Oxygen (DO)
  • Sensor Output: Arc: Modbus, 4 to 20 mA
  • Length: 120 mm
  • Electrical Connector: VP8
  • Measurement Principle: Electrochemical reduction of oxygen
  • Measuring Range: 10 ppb to 40 ppm (DO) or 0.2 to 1000 mbar (pO₂)
  • Accuracy at 25°C: ± 2%
  • Drift at Room Temperature: < 1% per week
  • Sensor Cap: Delivered with Optiflow FDA
  • Electrolyte: Oxylyte
  • Oxygen Consumption: Approximately 20 ng/h in air at 25°C
  • Temperature Sensor: NTC 22 kOhm
  • Configurable Values: DO: mbar; %-sat; %-vol; µg/l; mg/l; ppb/ppm (gas); ppb/ppm (dissolved oxygen); Temperature: °C, °F, K
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Stainless Steel 1.4435, FKM, Silicone - FDA 21 CFR 177.2600, VMQ
  • Surface Quality of Steel: Ra < 0.4 µm (N5)
  • Electrode System: Silver platinum combination
  • Analog Interface 1: 4 to 20 mA for DO, programmable
  • Analog Interface 2: 4 to 20 mA for Temperature, programmable
  • Analog Interface 1 and 2: Galvanically not isolated; pulse width modulation 3.5 kHz
  • Baud Rate: 4800, 9600, 19200, 38400, 57600, 115200 Bd
  • Polarization Voltage and Time: -670 ± 50 mV, ≥ 2 hours
  • Operating Voltage: 7 to 30 VDC max. 150 mW
  • Serial Number: Yes
  • Response Time: t98% < 60 s at 25°C
  • Certificate: Yes, with parameter settings and materials used

Compliance and Approvals
  • ATEX Approval: No
  • IECEx Approval: No
  • Autoclavable: Yes
  • CIP: Yes
  • Steam Sterilizable: Yes

Operating Conditions
  • Operating Temperature Range: Analog interface: 0 to 110°C; Digital interface: 0 to 130°C
  • Pressure Range: 0 to 4 bar
  • Max. CO₂ Partial Pressure: 0.01 bar
  • Required Flow: ≥ 0.03 m/s
  • Automatic Polarization: Yes, max. 2 hours stabilization time

Note: Specifications are subject to change without notice.
See more

Catalog excerpts

OxyFerm FDA Arc 120-1

OxyFerm FDA Arc 120 Specification Sheet (Part/REF # 243100) A combination of the hygienic design of the OxyFerm with the advantages of Arc technology for analog and digital communication. Product Specifications Sensor Family Sensor Output Electrical Connector Measurement Principle Electrochemical reduction of oxygen Measuring Range Drift at Room Temperature Delivered with Optiflow FDA Oxygen Consumption Temperature Sensor Configurable Values DO: mbar; %-sat; %-vol; µg/l; mg/l; ppb/ppm (gas); ppb/ppm (dissolved oxygen); Temperature: °C, °F, K Process Connection Wetted Parts Stainless Steel 1.4435 FKM (Fluorocarbon Elastomer) Silicone - FDA 21 CFR 177.2600 VMQ (Silicone elastomer) See compliance details in Material Specification document Surface Quality of Steel Electrode system Silver platinum combination Galvanically not isolated; pulse width modulation 3.5 kHz Baud Rate Polarization Voltage and Time Operating Voltage Serial Number Response Time Yes, with parameter settings and mate

 Open the catalog to page 1
OxyFerm FDA Arc 120-2

ATEX Approval IECEx Approval Steam Sterilizable Operating Temperature Range Analog interface: 0 to 110 °C; Digital interface: 0 to 130 °C Required Flow Automatic Polarization Yes, max. 2 hours stabilisation time Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/243100 Spec. Versio

 Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold G 120

    2  Pages

  2. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.