Product Overview
OneFerm pH sensors are designed for compatibility with various single-use bags and bioreactor installations, offering multiple electrical connector options and sensor lengths.

Specifications
  • Sensor Family: OneFerm
  • Parameter: pH
  • Length: 120 mm
  • Electrical Connector: VP6
  • Sensor Output: Traditional mV
  • Measurement Principle: Combination electrode; pH potential measured against reference
  • Measuring Range: pH 3 to 10
  • Sensitivity: 57 to 60 mV/pH at 25 °C before gamma irradiation; shift of ± 1.5 mV/pH after irradiation
  • Zero Point: ± 20 mV before gamma irradiation; shift of ± 10 mV after irradiation
  • Calibration Data: Provided on sensor label
  • pH Accuracy: ± 0.15 pH after gamma irradiation and dry storage; ± 0.1 pH after product calibration
  • Drift at Room Temperature: < 0.1 pH per week after wet-in time, in buffer pH 7
  • Membrane Shape: Spherical
  • Diaphragm: HP-Coatramic
  • Temperature Sensor: Pt1000
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Glass, VMQ (Silicone elastomer)
  • O-ring Material: FDA 21 CFR 177.2600; EG 1935/2004; USP <87>; USP <88> Class VI (121 °C)
  • Biocompatibility: Tested according to ISO 10993-5 and USP <87>
  • Wet-in Time: 30 min after dry storage
  • Shelf Life: 12 months prior to gamma irradiation; 18 months after gamma irradiation
  • Autoclavable: No
  • CIP: No
  • Steam Sterilizable: No
  • Gamma Irradiatable: Yes, up to 45 kGy
  • Operating Temperature Range: 4 to 50 °C
  • Sample Minimum Conductivity: 100 µS/cm
  • Pressure Range: 0 to 2 bar

Additional Information
Specifications are subject to change without notice. For more details, visit the Hamilton Company website.
See more

Catalog excerpts

OneFerm pH VP 120 Pt1000-1

OneFerm pH VP 120 Pt1000 Specification Sheet (Part/REF # 243267) OneFerm pH sensors are offer multiple electrical connector options and sensor lengths to work with most single use bags and bioreactor installations. Product Specifications Sensor Family Electrical Connector Sensor Output Measurement Principle Combination electrode; pH potential measured against reference Measuring Range Before gamma irradiation: 57 to 60 mV/pH at 25 °C (97 % -101 %); Shift after gamma irradiation: ± 1.5 mV/pH Zero Point Before gamma irradiation: ± 20 mV; Shift after gamma irradiation: ± 10 mV Calibration Data Yes, provided on sensor label After gamma irradiation and dry storage: ± 0.15 pH with provided calibration values. ± 0.1 pH after product calibration at the measurement temperature and within 2 pH units from the product calibration point Drift at Room Temperature < 0.1 pH per week after wet-in time, in buffer pH 7 Membrane Shape Temperature Sensor Process Connection Wetted Parts Glass VMQ (Silicone elastomer) O-ring Material Wet-in Time Shelf Life 12 months prior gamma irradiation. 18 months after gamma irradiation Steam Sterilizable Gamma Irradiatable Operating Temperature Range 4 to 50 °C Sample min Conductivity

 Open the catalog to page 1
OneFerm pH VP 120 Pt1000-2

Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/243267

 Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold G 120

    2  Pages

  2. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.